+
Camera for transmission He ion microscopy
Karen L. Kavanagh, Christoph Herrmann, and John A. Notte
Citation: Journal of Vacuum Science & Technology B, Nanotechnology and Microelectronics: Materials,
Processing, Measurement, and Phenomena 35, 06G902 (2017); doi: 10.1116/1.4991898
View online: http://dx.doi.org/10.1116/1.4991898
View Table of Contents: http://avs.scitation.org/toc/jvb/35/6
Published by the American Vacuum Society
Camera for transmission He1 ion microscopy
Karen L. Kavanagha) and Christoph Herrmann
Physics Department, Simon Fraser University, Burnaby, British Columbia V5A 1S6, Canada
John A. Notte
Carl Zeiss Microscopy, Peabody, Massachusetts 01690
(Received 23 June 2017; accepted 11 September 2017; published 26 September 2017)
The intensity of transmitted Heþ ions and atoms from a focused Heþ beam was measured using a
direct impact onto a camera located 20 cm below the sample. The camera consisted of a 256 256
array of Si p-i-n diodes (pixels) each 55 55 lm2 in area. Static, focused beam intensity profiles
show a uniform distribution, as expected from the small de Broglie wavelength (80 fm) and coherent
source. From the size of the beam spot, the half-angle beam convergence angle can be directly
measured and compared with theoretical predictions based on column geometries. The detector
count rate was consistent with an efficiency of 75%, when compared to the current measured by
electrostatic beam blanking. The intensity profile of a neutral atom component was measured to
have a 17% peak or 1.3% integrated intensity at a beam energy of 25 keV, compared to the ion
beam. The intensity of transmitted particles through 50 nm amorphous carbon was less than that of
graphite flakes, likely affected by thickness and channeling. Transmission images as a function of
C 2017 American Vacuum Society.
time to form 20 nm holes in graphite were obtained. V
[http://dx.doi.org/10.1116/1.4991898]
I. INTRODUCTION
One of the first applications of helium (He) ions was to
probe the nature of the atomic nucleus.1 The detection of
backscattered 5 MeV Heþ2 ions confirmed the existence of a
dense atomic nucleus and changed our understanding of the
atom. The same approach was developed into a very powerful analysis technique generally known as Rutherford backscattering spectroscopy.2 Backscattered ion energy profiling,
channeling, and blocking are used to study the atomic
composition, crystallinity, atomic positions, and fraction of
defects and interstitial atoms in materials. Subsequent
experiments with lower energy Heþ2 ions and neutral atom
beams have extended the beam solid interactions to include
atomic surfaces and the detection of He atom diffraction.3,4
The development of field ion microscopy and liquid metal
ion sources has enabled the construction of focused Heþ ion
beams (20–40 kV) for milling, lithography, and scanning secondary electron (SE) imaging with a nanometer resolution.5,6
Most recently, time-of-flight detection of backscattered and
secondary ions has been added to such techniques.7 The spatial coherence of these Heþ beams is thought to be significant
due to the atomic size of the emission sources. Channeling is
evident from the reduction in SE emission at the surface as a
function of sample orientation;5,6,8 scanning transmission
images with an atomic resolution are theoretically predicted,9
but such transmission images are yet to be obtained.
We define three classes of transmitted Heþ detectors,
depending on the information collected. The simplest detectors (class 1) detect all Heþ ions without any information on
angular distribution. The detection of secondary electrons on
a conversion plate, excited below thin samples, has been
used as an endpoint or thickness detector,10 which is also
a)
Electronic mail: kavanagh@sfu.ca
06G902-1 J. Vac. Sci. Technol. B 35(6), Nov/Dec 2017
possible with a scintillator, a channeltron, or a microchannel
plate. Class 2 detectors detect a specific range of transmitted
angles via an aperture, such as annular dark field detectors in
transmission electron microscopy.11,12 In this way, the influence of the sample thickness and density can be probed using
a scanning focused beam.
In this work, we detect the intensity and scattering angles
of forward-scattered Heþ ions and atoms (class 3) using a
camera, consisting of an array of Si diodes located below the
sample stage of a Heþ ion microscope (HIM). We characterize transmitted neutral He atoms and Heþ ion beam intensity
profiles as a function of microscope parameters and apply
this to milling nanometer holes in graphite.
II. EXPERIMENT
The HIM (Zeiss Orion Nanofab) was operated at 25–35 keV
with beam currents below 2 pA. Scanning SE images were
obtained using an Everhart–Thornley electron detector. We
added a camera below the sample stage, consisting of a
commercially available square array of Si p-i-n diodes.13
Designed primarily for the detection of x-rays, it has
256 256 pixels each 55 55 lm2 square, for a total detector area of 1.4 1.4 cm2. The n-type Si side of the array
(1 lm thick) was exposed directly to the Heþ beam without
optical filters or scintillation layers, other than a Si native
oxide. The camera was operated with a reverse bias voltage
of 100 V, with a minimum signal discrimination of 4 kV.
There were no counts detected (1 s exposure) if the valve
between the ion source and the chamber is closed (preventing all He from reaching the detector). Thus, the dark current was zero save for infrequent cosmic rays (1/min) that
were identified by their multiple pixel tracks. The camera is
supported on a narrow Al plate (1 mm thick) on the floor of
the HIM vacuum chamber, placed over an exit flange
2166-2746/2017/35(6)/06G902/5/$30.00
C 2017 American Vacuum Society
V
06G902-1
06G902-2 Kavanagh, Herrmann, and Notte: Camera for transmission He1 ion microscopy
06G902-2
(turbopump below the chamber). It is powered via a universal serial bus cable to an external personal computer,
through an added vacuum-compatible socket on the chamber left-hand side door. Images are collected using a software (Pixelite) interfaced to the camera controller.
A diagram of the ion column is shown in Fig. 1, indicating the beam limiting aperture, the blanking plates, the
objective lens, sample, and geometry of the camera at the
bottom of the chamber.6 The distance between the surface of
the camera (thickness 1.15 cm) and the bottom of the objective lens pole-piece is 21 cm. Like scanning electron microscopy, the beam current in a HIM can be controlled by the
size of a beam-limiting aperture inserted before the final
objective lens, and the spot size on that aperture is determined by the cross-over point of the upper lens. In addition,
the current is also directly proportional to the He gas pressure in the gun, which determines the rate of adsorption and
emission from the source tip. Beam currents are measured
by deflecting the beam into a large fixed aperture below the
blanking plates. Values as low as 10 fA were measureable,
limited by the sensitivity of the current meter.
The HIM has a standard sample holder that is moved
onto the HIM stage through a vacuum load-lock. The stage
provides the displacement of the sample holder in the (x, y)
plane (625 mm) and out of the plane, z (8 mm), and rotation, R, about z (360 ). It is also attached to a tilting
“cradle” that tilts the entire stage and the sample holder
about the x-axis, T, (þ52 to 5 ). But there is no hole
through the stage center. To allow beam transmission, it
was necessary to add a slit (3.0 4.5 cm2) in the left side
panel of the tilt cradle. With the cradle tilted þ45 and the
stage displaced to an extreme corner (x, y) ¼ (25, 25) mm
with respect to the centre, the beam avoids hitting the stage
or the cradle.
Samples were 3 mm diameter TEM-compatible grids or
substrates that were mounted onto a special arm on the sample holder that extended (x, y) ¼ (–25, –25) mm over its
edge, positioning thin samples under the beam. To provide
normal incidence, this arm also had a –45 fixed tilt.
III. RESULTS AND DISCUSSION
Figure 2(a) shows a scanning SE, HIM image from a
polycrystalline-Si film (90 nm thick) suspended in a Cu folding grid. The SE silicon intensity is the greatest where the
sample is transparent, in between some of the grid openings.
This might be due to charging but more likely due to the
generation of SE from both sides of the sample. The projected range for Heþ ions (30 keV) into Si is 280 6 100 nm,
and so, 99% of He transmitted the sample. Figure 2(b) shows
the corresponding scanning transmission HIM image
obtained after one focused beam scan (total exposure time
1 s). Detection of Heþ ions or atoms occurs over the entire
area of the camera, which subtends a solid angle of 5 msr.
Figure 2(c) shows a profile of the average intensity per pixel
(counts/s) as a function of the position inside the rectangle in
(b), showing the periodicity of the Cu grid openings. The
maximum current detected was 7 cps/pixel through the
J. Vac. Sci. Technol. B, Vol. 35, No. 6, Nov/Dec 2017
FIG. 1. (Color online) Schematic diagram of the Heþ ion column including
the beam limiting aperture, beam blanker location, bottom objective lens,
sample, and He camera added to the bottom of the chamber. [Figure 1.4
adapted with permission from Helium Ion Microscopy, edited by Hlawacek
and G€
olzh€auser (Cham : Springer International Publishing : Imprint:
Springer, 2016). Copyright 2016 by Springer Nature.)
hexagonal Cu grid holes consistent with the fast focusedbeam scanning rate. No counts were detected within the
region of poly-Si.
The grid holes are visible as magnified circles spaced 35
pixels apart or 1.9 mm on the detector. Since their actual
spacing is 150 lm, this means that a magnification of 13
occurs due to the location of the raster pivot point. The blurriness of the image in Fig. 2(b) is attributed to the divergence
06G902-3 Kavanagh, Herrmann, and Notte: Camera for transmission He1 ion microscopy
FIG. 2. Poly-Si (90 nm) in between a Cu folding grid (3 mm diameter). (a)
Scanning Heþ ion microscopy image via secondary electron detection (25
fA, 5 ls scan rate). (b) Corresponding He camera image (25 fA, total exposure 1 s). (c) Profile of average within the exposure time of the scan intensity
(counts/s) collected from the pixels within the rectangle in (b). (d) Camera
image with the Heþ ion beam blanked (500 s exposure).
of the beam as it passes through the sample (a divergence
half angle of a few mrad). Transmission is blocked by the
sample holder on the periphery of the image.
The pixels indicated as dead were previously exposed to a
Heþ ion dose that exceeded a threshold for permanent failure. Dead pixels developed during our earliest investigations
of the detector. The maximum number of counts for each
pixel is 10 000, equivalent to only an ion current of 1.6 fA, if
1 count equals one He. Currents as high as 60 pA are readily
available with the instrument, and so, it was easy to saturate a
pixel. Too many ions into a biased pixel, apparently, permanently damage its sensitivity, but we have not investigated
this in any detail. We subsequently avoided pixel currents
above 1 pA and tried never to expose the direct beam to a
biased detector for more than 10 ms.
Figure 3 shows a camera image (0.1 s exposure) of the
focused beam spot with the sample removed. In this case,
the microscope column operated with an accelerating voltage of 25 kV, but with the same 20 lm limiting aperture and
spot size of 4, with low He pressure such that the blanked
beam current was 50 6 10 fA. The ion beam was static (spot
mode) displaced from the center and focused at the sample,
with a distance between the sample surface and the objective
lens (working distance) of 10 mm. The larger spot is the ion
beam, while the smaller one is due to ions that neutralized
within the gun or elsewhere along the column. This neutral
He fraction is not focused, and therefore, these accelerated
atoms are detected predominantly directly below the gun
position. The plot shows the average intensity (cps) through
the beam spot center within the rectangular area. As
expected, we see a flat distribution in the ion beam intensity
06G902-3
FIG. 3. (Color online) Camera image of the static beam (spot mode) with a
profile plot of the average intensity per pixel (counts/s) as a function of the
pixel position within the rectangle. The beam was focused at the sample
height, and a limiting 20 lm aperture, an exposure of 0.1 s, and a current of
50 fA were used to obtain this image. The image has a grayscale contrast
with white which is the highest range.
to within a 10% statistical error. Furthermore, the focused ion
beam has a diameter of 20 pixels for a total number of pixels
in this spot of 314. Hence, given that the total beam current
was 50 6 10 fA, the current per pixel was 0.16 6 03 fA or
1000 6 200 Heþ ions. Compared to the average number of
counts detected (750 6 50/s), the efficiency of the detector is
75%. Thus, either 25% of the beam is lost between the blanking plates and the detector (unlikely) or the camera misses a
statistical fraction of the He beam (some will not penetrate
into Si to the sufficient depth to reach the depletion width of
the reverse-biased diodes).
When the ion beam is blanked, an accelerated, neutral He
atom fraction continues to the detector. From the intensity
profile in Fig. 3, the neutral-to-ion peak intensity ratio was
17%, while the integrated count from neutrals (2.5 kcps)
compared to the ion beam (192 kcps) is 1.3 6 0.1%. The
analysis of SE intensity ratios from images produced using
higher beam currents (5 pA) found a smaller neutral fraction
of 0.02%.5,6
The full width half maximum of the beam is 16 pixels,
while the neutral was smaller, 3 pixels. The beam focused at
the sample expanded to a circle on the camera of radius
440 lm (8 pixel radius 55 lm/pixel), giving a half-angle
convergence, b, of 2.2 mrad (¼ 440 lm/20 cm). Measurements
of the ion beam radius, as a function of the increasing spot
size, and limiting aperture size, gave the expected reciprocal, parabolic behavior (larger spot size ¼ smaller beam
area). Theoretical calculations of the beam convergence
angle, b, give similar values.6 Most of the neutral beam
intensity must originate close to the gun and the ion beam
path since its footprint on the detector is a projection. Thus,
its exact position and width are primarily determined by the
mechanical position of the gun and the effect of the gun tilt
JVST B - Nanotechnology and Microelectronics: Materials, Processing, Measurement, and Phenomena
06G902-4 Kavanagh, Herrmann, and Notte: Camera for transmission He1 ion microscopy
and shift controls on the alignment condition. These change
each time the trimer source is adjusted, and so, both the neutral atom spot and zero-deflection ion beam spot positions
are alignment dependent.
Figure 2(d) shows a camera image for a 500 s exposure
with the ion beam blanked. The neutral atoms are most
intense within the 3 mm diameter Cu grid area, but there is a
small flux throughout the detector area from scattering or
less frequent neutralization collisions. The connected pixel
events are due to cosmic ray particles that impact our detector at a rate of 1/min. This means that there will always be a
low level of unfocussed He atoms reaching the sample.
However, their largest intensity is limited to a small number
of pixels, and thus, they will have a negligible effect on SE
images.
Figure 4(a) shows a typical SE image obtained from a
commercial graphite flake TEM support grid. It consisted of
exfoliated graphite flakes supported by a lacey-carbon mesh
on a Cu grid. The brightest regions are the amorphous laceycarbon (60 nm thick) which has a distinctive structure consisting of long narrow strands. The material in between is a
film with abrupt steps in its SE intensity but continuous,
covering 99% of the grid openings, often with faceted
edges and filamentary white lines consistent with ripples
common to the thinnest layers of graphene. STEM imaging
and diffraction (not shown) concluded that the material is a
multilayered, hexagonal matrix, but with random rotation
about the perpendicular axis consistent with multiple
graphite sheets. Electron energy loss spectroscopy finds the
thickness to vary from 7 to 60 nm based on the ratio of the
inelastic to zero-loss peak areas.
Hole milling investigations confirmed that the brighter
regions in the SE image of Fig. 4 were associated with the
thickest graphite films. Figure 4(b) shows SE images after a
hole was milled into a brighter region of the graphite. Next
to it is a failed milling attempt where only a white donut
shape is visible. Figure 4(c) shows the transmitted peak
intensity as a function of time for a static, focused beam spot
in darker (black squares) and brighter regions (red triangles)
of the SE image. After a delay of seconds, the intensity
begins to rapidly increase. The corresponding transmission
images showed that a hole forms just when the intensity
begins to rapidly increase, with the delay longer for the
brighter regions compared to the darker graphite. The rate of
increase is similar, but the maximum intensity for the
brighter regions is greater than that for the darker since the
hole size is larger. Both subsequently begin to decrease with
time. The brighter regions took longer to mill compared to
the darker regions, indicating that the thickness was indeed
correlated with a brighter SE intensity. In many cases, a hole
formed that was 20 nm in diameter, as visible via the SE
image. In other cases, the same exposure resulted in a buildup of the material, not a hole, confirmed by a reduction in
the camera signal. Carbon and associated organic residues
are susceptible to beam-induced reactions by the He beam,14
resulting in surface diffusion and accumulation near the
beam spot. Previous exposures at a given location made
J. Vac. Sci. Technol. B, Vol. 35, No. 6, Nov/Dec 2017
06G902-4
FIG. 4. (Color online) SE images of a graphite on the lacey-carbon TEM support
grid at a beam current of 1 pA: (a) typical area and (b) high magnification of the
indicated square region in (a). The inset shows a higher magnification image of
the milled holes. (c) Peak transmission intensity (counts/s) as a function of time
for darker (black squares) and brighter (red triangles) SE regions.
subsequent milling more difficult, probably due to the
increased thickness due to the accumulation of carbonaceous
deposits.
Figure 5(a) shows a camera image from transmission
through a darker graphite region in the SE image and (b)
from the lacey carbon. The intensity profiles from these
06G902-5 Kavanagh, Herrmann, and Notte: Camera for transmission He1 ion microscopy
06G902-5
transmission Heþ ion microscopy. The camera counts single
Heþ ions and neutral atoms (energy 20–40 keV), providing
information on the beam position, shape, and angular intensity profile, as well as forward scattering patterns through
thin samples. Based on the size of the beam at the detector,
the half-angle divergence angle of the focused beam was 2
mrad for a limiting aperture of 20 lm and a beam energy of
30 keV. Forward-scattered, transmission images as a function of milling time of graphite were obtained. Nanometer
scale holes (20 nm diameters) were obtained when surface
carbon accumulation was avoided.
An improved forward-scattering resolution would be feasible with a smaller camera pixel size or longer distance
from the sample. Detection of channeling and diffraction
from 30 keV He (de Broglie wavelength 80 fm; diffraction
angle 0.2 mrad) should be feasible given suitably thin and
uniformly thick membranes.
ACKNOWLEDGMENTS
The authors are grateful for funding from NSERC, CFI/
BCKDF, and facilities managed by 4DLabs.
1
E. Rutherford, Philos. Mag. 21, 669 (1911).
L. C. Feldman, J. W. Mayer, and S. T. Picraux, Materials Analysis by Ion
Channeling (Academic, New York, 1982).
3
Helium Atom Scattering from Surfaces, Springer Series in Surface
Sciences Vol. 27, edited by E. Hulpke (Springer-Verlag, Berlin, 1992).
4
J. Seifert, J. Lienemann, A. Sch€
uller, and H. Winter, Nucl. Instrum.
Methods Phys. Res., B 350, 99 (2015).
5
D. C. Joy, He Ion Microscopy Principles and Applications (Springer, New
York, 2013).
6
olzh€auser
Helium Ion Microscopy, edited by G. Hlawacek and A. G€
(Cham : Springer International Publishing : Imprint: Springer, 2016).
7
L. Pillatsch, N. Vanhove, D. Dowsett, S. Sijbrandijb, J. A. Notte, and T.
Wirtz, Appl. Surf. Sci. 282, 908 (2013).
8
L. A. Giannuzzi and J. R. Michael, Micros. Microanal. 19, 344 (2013).
9
A. J. D’Alfonso, B. D. Forbes, and L. J. Allen, Ultramicoscopy 134, 18
(2013).
10
A. R. Hall, Microsc, Microanal. 19, 740 (2013).
11
J. A. Notte, R. Hill, S. M. McVey, R. Ramachandra, B. J. Griffin, and D.
C. Joy, Microsc. Microanal. 16, 599 (2010).
12
T. Woehl, Microsc. Microanal. 22, 544 (2016).
13
“Modupix camera,” Institute of Experimental and Applied Physics, Czech
Technical University, Prague, http://www.widepix.cz/products.
14
S. Tan, K. Klein, D. Shima, R. Livengood, E. Mutunga, and A. Vladar,
J. Vac. Sci. Technol., B 32, 06FA01 (2014).
2
FIG. 5. (Color online) Camera images of the transmission of a static,
focussed He beam through the graphite-coated TEM support grid of Fig. 4
(1 s exposure, 1 pA) located on (a) a thinner (darker) graphite area or (b)
amorphous lacey carbon. (b) Intensity profiles from the rectangular area in
(a) graphite (black squares) lacey carbon (red triangles).
images show that lacey carbon had a much lower transmission intensity compared to the graphite region, primarily due
to a difference in the thickness. The degree of channeling
through the graphite films is under investigation.
IV. CONCLUSIONS
We have investigated the properties of a commercial Si
p-i-n diode array (256 256 pixels) as a camera for
JVST B - Nanotechnology and Microelectronics: Materials, Processing, Measurement, and Phenomena