Journal of Electrostatics 88 (2017) 127e133
Contents lists available at ScienceDirect
Journal of Electrostatics
journal homepage: www.elsevier.com/locate/elstat
A practical comparison of surface resistance test electrodes
Jeremy Smallwood
Electrostatic Solutions Ltd, 13 Redhill Crescent, Bassett, Southampton, SO16 7BQ, UK
a r t i c l e i n f o
a b s t r a c t
Article history:
Received 19 September 2016
Received in revised form
22 December 2016
Accepted 27 December 2016
Available online 19 January 2017
Four surface resistance test electrodes are compared using a selection of materials under similar test
conditions. The results vary considerably with some materials due to variation in surface resistivity.
Using a relatively uniform material two concentric ring electrodes compliant with the same standard
differed in results by a factor of 1.8. Silver stripe and copper tape electrodes gave results a factor 0.4 and
0.7 compared to the reference electrode. A 2-pin electrode gave results a factor 4.7 greater. The 2 pin
probe cannot be expected to give similar results to the other electrodes for materials that have variable
resistivity.
© 2017 Elsevier B.V. All rights reserved.
Keywords:
Surface resistance
Standard measurement electrodes
1. Introduction
Surface resistance test methods are used for a variety of purposes in industry including evaluation of resistive materials for
electrostatic ignition risk, qualification of equipment intended for
use in potentially explosive atmospheres under European Directive
2014/34/EU (ATEX) [1] and characterization of packaging materials
for use in Electrostatic Discharge (ESD) protection in the electronics
industry. Several standard test methods exist for these purposes.
The general objective of the measurement in each case is to evaluate the electrostatic charge dissipative ability of the material
surface, for avoidance of charge build up and subsequent electrostatic discharges. When the resistance between the electrodes is
measured, it is typically reported without compensation for the
electrode form, as “surface resistance” rather than surface resistivity. This greatly increases convenience, especially for routine
and frequent measurements. Related standards (e.g. IEC 61340-5-3
[9] or IEC 60079-0 [8]) then specify materials in terms of surface
resistance for material classification and evaluation purposes.
Conductive rubber faced concentric ring electrodes (IEC 613402-3 [5,6] and ESD STM 11.11 [2,3]) were developed for compliance
verification evaluation of electrostatic discharge (ESD) control
materials used in packaging for the electronics industry. IEC 613402-3 [5,6] is also specified for general measurements of properties of
materials with resistance or resistivity in the range 104e1012 U use
in electrostatic control. The concentric ring electrodes typically
have an outer ring outer diameter of 63 mm and mass of 2.5 kg.
E-mail address: jeremys@static-sol.com.
http://dx.doi.org/10.1016/j.elstat.2016.12.019
0304-3886/© 2017 Elsevier B.V. All rights reserved.
They are suitable for moderate sized flat samples of material, but
are unsuited for measurement of small, curved or other non-planar
surfaces.
With the development of miniature ESD control packaging
products there evolved a need for a miniature surface resistance
test method that could also be used on non-planar surfaces or
within depressions in moulded products. A miniature point-topoint 2-pin probe electrode (IEC 61340-2-3 [5,6] and ESD STM
11.13 [4]) was developed for this purpose. This consists of two
sprung loaded and conductive rubber faced pins 3.2 mm in diameter and separated by 3.2 mm.
In evaluation of materials for electrostatic hazard control materials in processes and ATEX in Europe, a different electrode system has historically been used for many years, e.g. in IEC 600790 [8]. This consists of two stripes of conductive paint (e.g. silver
paint) 1 mm wide and 10 mm long, separated by a gap of 10 mm.
This electrode system has recently been published again in IEC
60079-32-2 [7]. This standard also allows the electrodes to be made
from conductive rubber or foam strips mounted on insulating
supports.
The standards that refer to these test methods typically specify
classifications or requirements in terms of surface resistance
measured according to these methods. It is typically the high
resistance limits that are of most interest. For example, in IEC
61340-5-3 [9] a material is classified as insulative if the surface
resistance measured according to IEC 61340-2-3 [2,3] (2 pin or CR
probe) is 1011 U. An insulative material would, where possible, be
excluded from use in an ESD Protected Area. In contrast in IEC
60079-0 [8] requires the surface resistance of enclosures for
equipment used within a flammable atmosphere area to be 109 U
128
J. Smallwood / Journal of Electrostatics 88 (2017) 127e133
calculated for surface resistivity given the same surface resistance
measurement result. In order to avoid this problem in calculating
KCR, the formula given in IEC 61340-2-3:2000 [5] is used to compare
electrode systems in Table 1. Calculation of KCR according to IEC
61340-2-3:2016 [6] gives only a small difference compared to the
IEC 61340-2-3:2000 [5] formula (0.100 compared to 0.096). As the
surface resistance values measured with each electrode are
compared directly by experiment, differences in results due to
different formulae for conversion to surface resistivity are avoided.
For a parallel stripe electrode such as IEC 60079-0 [8] (which is
the same as IEC 60079-32-2 [7]) the electrode constant is
Kst ¼
g
l
Fig. 1. Concentric ring (CR) electrode.
(@ 50 ± 5% r.h.) or 1011 U (@ 30 ± 5%r.h.) measured according to the
conductive paint stripe electrode method given in the standard. So,
it is particularly important that the measurement results are
repeatable and reproducible in these resistance ranges.
In theory the surface resistance Rs measured by an electrode
system across a surface of uniform surface resistivity rs is of the
form
Rs ¼ K rs
where K is a constant dependent on the electrode surface contact
geometry. Any electrode systems having the same K might be expected to give the same results when measuring a material of
uniform surface resistivity. For the concentric ring electrode system
the value of K is defined differently in different standards. In STM
11.11e2015 [3] surface resistivity is simply quoted as a factor of 10
times the surface resistance, making KCR ¼ 0.1.
In IEC 61340-2-3:2000 [5], the relationship between surface
resistivity and surface resistance was given as
rs ¼
Rs ðd1 þ gÞp
g
which leads to
KCR ¼
g
pðd1 þ gÞ
where d1 is the diameter of the inner contact electrode and g is the
gap between the inner electrode and inside of the outer electrode
(IEC 61340-2-3 [5,6]). In practice it is the inner diameter of the
outer electrode that is specified instead of the electrode gap. All
dimensions are specified in the standards with a tolerance (see
Fig. 1).
In STM 11.11e2006 [2] and 61340-2-3:2016 [6] the formula
relating Rs and rs is defined as
rs ¼
2pRs
loge dd2
1
This leads to
KCR ¼
loge d2=d
1
2p
As the objective of this paper is to directly compare surface
resistance measured using different electrodes, conversion to surface resistivity is not required. Clearly, different conversion factors
defined in different standards could lead to different values
where l is the length of the stripe and g is the gap between the
electrodes, and any fringe effects at the electrode ends are
neglected. The dimensions may be in meters or mm (see Fig. 2).
When the electrode tolerances are taken into account, each
electrode system gives a range for K that would be expected for
electrodes built according to the standards. For the CR electrodes,
the minimum K is given for minimum gap, and vice versa. For the
stripe electrode the minimum K is given for minimum gap and
maximum length, and vice versa. The calculated minimum, nominal and maximum values are given in Table 1. It can be seen that the
range of constants for the various standard cells is very similar,
ranging from about 0.091 to about 0.106. The approximate variation
in measurement result due to this, for a uniform resistivity sample,
should be about ±5% for the CR electrode and ±6% for the stripe
electrode system.
The electrode configuration for the 2 pin electrode system is
shown in Fig. 3. So far the author has not found an equation
deriving a constant K for this configuration.
Typical materials and products under test can have variable
resistance characteristics across the surface that can lead to differences in results according to the type and scale of the electrode
and direction or location of measurement. The concentric ring
electrode has infinite rotational symmetry and can be expected to
give the same result with all orientations even on an anisotropic
material. In contrast a parallel stripe electrode can be expected to
give results variable with orientation on an anisotropic material.
The concentric ring electrodes are designed for flat surfaces, and
typical products measured may have curved or textured surfaces.
Some products may have features that are too small to apply large
electrode systems. The 2 pin electrode structure allows measurements on small areas or curved surfaces. Conductive rubber faced
electrodes are easy to apply, do not affect the surface material and
are easily removed leaving no trace after measurement. The IEC
60079-0 [8] painted stripe electrode can conform to a surface
curvature or profile but may affect the surface material and could
be difficult or impossible to remove after measurement. For this
reason, IEC 60079-32-2 [7] allows alternative electrodes of the
same geometry. The standard electrode systems therefore have
advantages and disadvantages according to the form of the sample
under test.
This paper gives a comparison of results and experience of using
standard test electrodes on a variety of materials with resistance in
the range GU to TU. In addition, the use of self-adhesive conductive
(Cu) tape electrodes is explored. These electrodes have been used in
practice by the author for many years. These can conform to
moderate surface contours or textures and are often conveniently
applied. They normally do not physically affect the surface and can
be removed, leaving only a residue of adhesive that can be easily
cleaned off if necessary (see Fig. 4).
The objective is to directly compare the electrodes and surface
J. Smallwood / Journal of Electrostatics 88 (2017) 127e133
129
Table 1
Standard electrode parameters (CR electrodes calculated according to IEC 61340-2-3:2000) [5].
Electrode dimensions
IEC 61340-2-3 [5,6]
STM 11.11 [2,3]
IEC 60079-32-2 [7]
IEC 60079-0 [8]
Type
CR
CR
paint stripe
Electrode parameters
d1
d2
g
30.5 ± 1
30.5 ± 0.64
57 ± 1
57.15 ± 0.64
12.5e14.0
12.7e14.0
10 ± 0.5
Fig. 2. Paint stripe electrode.
Fig. 3. 2 pin electrode system (IEC 61340-2-3 [5,6] and ESD STM 11.13 [4]).
l
K min
K nominal
K max
100 ± 1
0.091
0.092
0.094
0.096
0.097
0.100
0.101
0.101
0.106
dimensions of these electrodes are compared in Table 2.
The electrodes are shown in Fig. 5. The Trek 152P-CR-1 probe
(TCR) is an active probe that was used with a compatible Trek 152-1
resistance meter. The Warmbier 880 CR probe (WCR) was used with
a Metriso 3000 resistance meter. Both probes had mass of 2.5 kg.
Resistance measurements above 1 MU were made at 100 V test
voltage. The Trek 152-1 m had a lower resistance measurement
limit of 1 kU and upper limit of 10 TU. The Metriso had an upper
limit of 1 TU.
Before measurements started, the contact resistance of the
commercial CR and 2 pin electrodes and Cu strip electrodes was
measured using a stainless steel test plate. In all cases the resistance
was found to be ≪ 1 kU at 10 V test voltage.
The measurements made with CR electrodes were used first to
establish a comparison of the CR electrodes and afterwards as a
basis for comparison with the 2 point probe electrode system. After
this, comparison was made with silver (Ag) paint stripe electrodes
and copper (Cu) tape electrodes. The Ag paint electrodes were
fabricated by drawing lines with a silver paint pen (ITW Circuitworks® Flex Conductive Pen CW2900) by hand using a ruler (Fig. 6).
In practice it was found to be difficult to maintain a uniform electrode line width and inter electrode gap and avoid smudging of the
paint.
Copper (Cu) tape electrodes were fabricated from conductive
Table 2
Commercial electrode parameters.
Fig. 4. Copper (Cu) strip electrodes.
resistance results obtained with them, and so the electrodes are
compared under the same humidity and test voltage conditions
rather than those specified in the standard test methods. For all of
the electrodes except the 2 pin electrodes, the value of K is calculated to be 0.1. For a homogenous test material and identical test
conditions, the measured surface resistance result should, in the
absence of other influential factors, be the same. The 2 pin electrodes are intended by the standards to be used in situations where
the CR electrodes are unsuitable due to curvature or size of the
specimen under test, and so also would be expected to give similar
results. This paper provides experimental evidence of variations
that occur in practical measurements and discusses the influence of
material non-homogeneity on the results.
Electrode
Type
Electrode parameters
d1
d2
g
K
Warmbier 880 (WCR)
Trek 152P-CR-1 (TCR)
CR
CR
30.2
30.7
57.6
57.7
13.68
13.52
0.099
0.097
2. Experimental
Two types of commercial CR electrodes available for the experiments. These were measured with a vernier gauge to determine
their constant K and establish compliance with the standards
before for use in experiments. As the electrode contacts were
flexible polymeric materials and could easily be deformed, it was
difficult to do this precisely. The electrodes were within the requirements of IEC 61340-2-3 [5,6] and ESD STM 11.11 [2,3]. The
Fig. 5. Commercial electrodes.
130
J. Smallwood / Journal of Electrostatics 88 (2017) 127e133
Fig. 7. Results of CR and 2 pin electrode measurements on insulative material.
Fig. 6. Fabricated electrodes.
self adhesive copper “EMC” tape (Advance AD528) 10 mm wide and
100 mm in length applied by hand to the surface. A 10 mm wide
tape is used as is easily handled, was readily available in the author's toolkit and from suppliers. The resistance result was not
expected to vary greatly with the electrode width. The interelectrode gap was 10 mm. The calculated constant for this arrangement
was 0.10 (see Fig. 6).
In order to compare measurements as nearly as possible on the
same position, a 100 mm 100 mm grid was marked on the material. Electrodes could then be set by eye on the center of each grid
square and measurement of the marked material was avoided. The
material was supported on a highly insulating (>1013 U) PVC slab.
All samples were conditioned for at least 24 h and tested at
50% ± 3% humidity rather than the humidity levels specified by the
standards. This is because the objective was to compare the
different electrodes under identical conditions, and not to evaluate
the materials under standard conditions. In practice, varying humidity can have significant effect on resistance measurement results, typically giving lower resistance results at higher humidity.
For the same reason, the fabricated electrodes were also used at the
same applied voltage (10 V and 100 V) as the standard electrodes.
Typical materials can often give lower resistance results at higher
applied voltage.
3. Results
Initial measurements were made on a flexible insulative material measured using the CR probes to have resistance of the order
1011e1012 U. A flexible material was used to avoid problems with
obtaining good electrode contact over the large CR electrode surface. Measurements were also made on a flexible “conductive” mat
material used in ESD control. Measurements were then made using
the 2 pin probe. In each case the resistance result was plotted on a
graph against the measurement location, so that measurements at
each position could be directly compared (Fig. 7).
The CR electrode measurements on the insulative material gave
reasonably consistent results. In order reduce the effect of material
resistance on the comparison of the electrodes, the results at each
position were “normalised” arbitrarily taking the WCR electrode as
the reference at each measurement location. The ratio of the result
for each electrode to the WCR electrode result was calculated for
each location. The average and standard deviation of these ratios
could then be calculated (Table 3). Using this method it was
calculated that the TCR electrode gave on average a factor of 2.7
times higher resistance result than the WCR electrode.
When 2 pin electrode measurements were added, these were
found to be considerably higher than the CR electrodes and highly
variable. In many cases 2 pin probe results were an order of
magnitude or more higher than the CR probe results. 75% of these
results exceeded the 1013 U upper measureable limit of the
equipment although none of the CR probe results exceeded 1012 U.
Those that did so were recorded as 1013 U to allow some analysis to
be made. The 2 pin electrode gave on average a factor of nearly 85
times higher resistance than the Warmbier CR electrode (see Fig. 7).
Measurements on the “conductive” ESD control material using
the CR electrodes showed the material to be highly variable in
resistance in different measurement positions. Measurements with
the 2 pin electrodes showed extraordinarily variable results, on
average a factor of about 1.5 104 times the WCR electrode (See
Fig. 8).
It was suspected from these results that the materials used in
initial tests had highly variable surface resistance that would make
comparison of electrodes difficult. Other materials were then tested
to try to find a more uniform material for electrode comparison
purposes that would have a resistance approximately in the desired
range. The most promising material found was an art paper. After
conditioning and test with the CR and 2 pin electrodes this
appeared to give reasonably uniform results (Fig. 9). The 2 pin
electrode was used to check for anisotropy by comparing measurements in 12 positions with the electrodes oriented in x and y
directions. No significant anisotropy was found.
Once again the TCR electrodes gave higher resistance than the
WCR electrodes (Table 4). The 2 pin electrodes gave higher resistance results than the CR electrodes. It can be seen that there
remain differences in the surface resistance results at different
positions on the material although it appears much more uniform
than the materials initially used.
Two types of fabricated electrodes were then tested on a selection of positions on the art paper. A smaller number of measurements of this type were made partly because once applied,
there would be no possibility of removing them without damage to
the material. Further measurements on those locations with
different electrodes would not be possible. The silver paint electrodes were found to be most difficult and time consuming to
fabricate accurately.
This analysis showed that for art paper TCR electrode gave results an average of 1.8 times the WCR electrode (Table 4). The 2 pin
electrode gave results an average of 4.7 times higher. In contrast,
the Ag paint and Cu tape electrodes gave results an average of 0.4
J. Smallwood / Journal of Electrostatics 88 (2017) 127e133
131
Table 3
Normalised ratios of measurements with each electrode at each location on insulative and conductive materials.
Electrode
WCR
TCR
2 pin
Insulative material
Conductive material
No. of measurements
Average
Standard deviation
No. of measurements
Average
Standard deviation
38
38
38
1.0
2.7
84.9
0.0
0.53
67.7
16
13
15
1.0
1.6
14800
0.0
1.0
36400
and 0.7 times the WCR electrode respectively. The standard deviations of results (TCR 0.38, 2 pin 0.79, Cu 0.19 and Ag 0.11) in each
case showed reasonable consistency.
A similar analysis was done on measurements made on a more
variable paper (Tyvek®). The basic results are given in Fig. 10. In this
case TCR electrode gave results an average of 1.2 times the WCR
electrode whereas the 2 pin electrode gave results an average of 5.5
times higher. The Cu tape electrodes gave results an average of 1.8
times the WCR electrode. The Ag stripe electrode was not tested.
The variability of the 2 pin electrode results for Tyvek® can be
seen in the higher standard deviation of 3.51 compared to 0.79 for
the art paper. In contrast the standard deviation of results for the
TCR electrode was 0.38 for art paper and 0.56 for Tyvek®. The
standard deviation for the Cu tape electrode for art and Tyvek papers were 0.19 and 0.92 respectively.
4. Discussion
Fig. 8. Results of measurements on „conductive“ ESD control material.
It is clear from these experiments that practical measurement
results can be much more variable that would be expected from
theoretical calculations assuming materials of uniform surface resistivity. Part of this variability is undoubtedly due to inherent
variability in the surface resistivity of the material under test. There
are, however, other unidentified factors. Based on electrode
Fig. 9. Measurement results on art paper.
Fig. 10. Measurement results on Tyvek® paper.
Table 4
Normalised ratios of measurements with each electrode at each location for art and Tyvek® paper.
Tyvek® paper
Electrode
Art paper
No. of measurements
Avg.
Std. Dev.
No. of measurements
Avg.
Std. Dev.
WCR (reference)
Trek CR
2 pin
Cu tape
Ag stripe
24
24
24
8
4
1.0
1.8
4.7
0.7
0.4
0.0
0.38
0.79
0.19
0.11
20
20
20
4
0
1.0
1.2
5.5
1.8
0.0
0.56
3.51
0.92
132
J. Smallwood / Journal of Electrostatics 88 (2017) 127e133
dimensions the ratio of the electrode constants of the Warmbier CR
and Trek CR electrodes would predict that the Trek CR would give
results a factor of 0.98 times the Warmbier CR electrode. In practice
a factor of 1.8 was found. Similarly the Ag stripe and Cu tape
electrodes would be expected to give 1.1 times the resistance of the
Warmbier CR electrode but in practice 0.4 and 0.7 were found
respectively. Whether these differences are due to resistance of the
contact between the electrode material with the test sample, or
other additional factors, is unclear.
It seems clear that the 2 pin electrode responds to variation in
the surface resistance of the material under test in a very different
way to the other electrodes. Examination of the geometry of the
electrodes suggests an explanation for this. The CR electrodes
measure the material surface resistance over a distance of at least
the perimeter of the inner electrode, which is around 96 mm. The
Ag stripe and Cu tape electrodes measure the material over a distance of 100 mm. Ignoring the possibility of conduction from the
rear of the electrode the 2 pin probe measures the material over a
much smaller distance that is around half the periphery of each
electrode, or about 5 mm. It is instructive to consider the response
of the electrodes to small regions of more conductive or more
insulative material that may occur in the measurement location
(Fig. 11 and Fig. 12). For the CR and stripe or tape electrodes, a small
region of high resistance material of the order of the dimension of
the electrode gap and crossing the gap is likely to have only a small
effect on the measured resistance because most of the material
between the electrodes remains unaffected. It acts like a high
resistance in parallel with the average resistance of the material
measured. In contrast a small region of low resistance material of
similar dimensions between the electrodes will have a relatively
large effect in reducing the measured result. The result tends to be
highly influenced by the minimum resistance region found between the electrodes. A small region of high or low resistance that
does not bridge the electrode gap has relatively little effect.
For the 2 pin electrode system, a similar size region of high or
low resistance bridging the gap may have a significant effect on the
measured result as the majority of the conduction path between
the electrodes is affected (Fig. 12). If one of the electrodes happens
to sit on a region of high resistance, a high resistance result will be
obtained. Thus, small regions of high or low resistance can be expected to significantly affect the result.
This can easily be demonstrated in practice by placing material
of another resistance under the electrodes. In a simple experiment,
10 mm strips of office paper and polymer foil were used to give
more conducting and more insulating regions on the art paper
surface. The art paper substrate was measured first with each
electrode in one location. The surface resistance of the strips was
measured with the strip laid on the art paper and both pins of the 2
pin electrode on the strip. The resistance was then measured with
Fig. 11. The effect of regions of more or less conductive material on the CR and stripe
or tape electrodes.
Fig. 12. The effect of regions of more or less conductive material on the 2 pin
electrodes.
each strip laid on the art paper across the diameter of the TCR
electrode, and under one pin of the 2 pin electrode (Table 5).
It can be seen that the presence of an insulating polymer strip
laid under the TCR electrode raised the result by a factor of
approximately 2 from 7.4 109 U to 1.6 1010 U. The same strip
under one pin of the 2 pin electrode raised the result from
2.2 1010 U to 1.2 1013 U, an increase of 550 times. With the strip
under both of the 2 pin electrodes the resistance was >2 1013 U. A
more conducting paper strip under the CR electrode reduced the
result by approximately a factor of 0.2 from 7.4 109 U to
1.3 109 U. The same strip under one pin of the 2 pin electrode
reduced the result from 2.2 1010 U to 1.1 1010 U, a factor of 0.5.
With the strip bridging both 2 pin electrodes the resistance was
1.0 109 U. Similar effects can be expected to occur with high and
low resistance regions of dimensions down to the inter electrode
gaps and diameter of the electrode pins.
A statistical number of measurements made using the 2 pin
electrode arguably would better represent the range of resistivity of
the material than similar measurements made using the other
electrodes. For evaluation of materials for electrostatic control and
risk avoidance all these methods may be equally useful. The CR, Cu
tape and Ag stripe electrodes give comparable results even with
variation in resistance of the material. The 2 pin electrode cannot,
however, be expected to give similar results to the other electrodes
when there is appreciable variation in material resistance.
5. Conclusions
Standard concentric ring electrodes according to IEC 61340-2-3
[5,6] and ESD STM 11.11 [2,3] have been compared with a standard
2 pin electrode according to IEC 61340-2-3 [5,6] and ESD STM 11.13
[4] and a silver paint stripe electrode according to IEC 60079-0 [8]
and IEC 60079-32-2 [7]. These standard electrodes have been
compared with an experimental conductive self adhesive copper
tape electrode system. All except the 2 pin electrode configuration
had a theoretical K of around 0.1.
All these methods were found to be susceptible to considerable
variation in results due to variability of local resistivity of the material under test and other unknown factors. When compared using
the most uniform material available (art paper), the TCR electrodes
gave results on average a factor of 1.8 higher, the 2 pin electrode
system 4.7 higher, copper tape 0.7 and silver paint stripe 0.4 times
the WCR electrode arbitrarily used as reference.
When test materials had significant variability, the WCR and
TCR electrodes remained reasonably comparable although the 2 pin
electrodes gave highly variable results. A small region of high
resistance under one electrode can lead to a high resistance result
whereas a small region of low resistance can bridge the electrodes
and give a low resistance result. The WCR, TCR, Cu tape and Ag
stripe electrodes are only moderately affected by small regions of
high resistance. All the electrodes are highly influenced by regions
J. Smallwood / Journal of Electrostatics 88 (2017) 127e133
133
Table 5
Effect of high and low resistance strips placed under electrodes on the measurement result.
Material strips used
Measurement configuration
Under both pins of 2 pin electrode
No strip
polymer strip
office paper strip
Under one pin of 2 pin electrode
Across diameter of TCR electrode
1.2 1013
1.1 1010
7.4 109
1.6 1010
1.3 109
10
2.2 10
>2 1013
1.0 109
of lower resistance bridging the electrodes.
It seems likely that with common ESD control materials poor
reproducibility and variation of results is likely due to material local
resistance variations. In particular, the 2 pin electrode is likely to
give a wide variation of results reflecting high and low resistance
regions in the material. The other electrodes are likely to give results representing mainly the lower resistance regions in the material. A statistical number of measurements made using the 2 pin
electrode could better represent the range of resistivity of the
material than similar measurements made using the other
electrodes.
With a relatively uniform test material (art paper) there
remained differences between the resistance measured with each
electrode that are unexplained by differences in electrode geometry. These indicate that contact resistance or other as yet unidentified factors have considerable effect on the measurement results.
The concentric ring, Cu tape and Ag stripe electrodes can be
considered to give reasonably similar surface resistance results
within an order of magnitude. The 2 pin probe cannot be expected
to give similar results to the other electrodes for materials that may
have be non-uniform resistivity.
References
[1] Directive 2014/34/EU of the European Parliament and of the Council of 26
February 2014 on the Harmonisation of the Laws of the Member States Relating
to Equipment and Protective Systems Intended for Use in Potentially Explosive
Atmospheres (Recast), 2014, p. 309. OJEU L 96, 29.3.2014.
[2] ESD Association, Surface Resistance Measurement of Static Dissipative Planar
Materials, ANSI/ESD STM 11.11e2006, 2006, ISBN 1-58537-124-6.
[3] ESD Association, Surface Resistance Measurement of Static Dissipative Planar
Materials, ANSI/ESD STM 11.11e2015, 2015, ISBN 1-58537-280-3.
[4] ESD Association, Two-Point Resistance Measurement, ANSI/ESD STM
11.13e2015, 2015. ISBN: 1-58537-279-x.
[5] International Electrotechnical Commission, Electrostatics e Part 2e3: Methods
of Test for Determining the Resistance and Resistivity of Solid Planar Materials
Used to Avoid Electrostatic Charge Accumulation, IEC 61340-2-3:2000, 2000,
ISBN 978-2-8322-3475-4.
[6] International Electrotechnical Commission, Electrostatics e Part 2e3: Methods
of Test for Determining the Resistance and Resistivity of Solid Planar Materials
Used to Avoid Electrostatic Charge Accumulation, IEC 61340-2-3:2016, 2016,
ISBN 978-2-8322-3475-4.
[7] International Electrotechnical Commission, Explosive Atmospheres Part 32-2.
Electrostatic Hazards Tests. IEC 60079-32-2, ISBN 978 2 8322 2276 8.
[8] International Electrotechnical Commission, Explosive Atmospheres - Part 0:
Equipment - General Requirements, IEC 60079e0:2011, 2011, ISBN 978-288912-519-7.
[9] International Electrotechnical Commission, Electrostatics e Part 5-3: Protection
of Electronic Devices from Electrostatic Phenomena - Properties and Requirements Classifications for Packaging Intended for Electrostatic Discharge
Sensitive Devices, IEC 61340-5-3:2015, 2015, ISBN 978-2-8322-2787-9.