Academia.eduAcademia.edu

Measuring the linewidth of a stabilized diode laser

2015, arXiv: Atomic Physics

We demonstrate a straight-forward technique to measure the linewidth of a grating-stabilized diode laser system---known as an external cavity diode laser (ECDL)---by beating the output of two independent ECDLs in a Michelson interferometer, and then taking the Fourier transform of the beat signal. The measured linewidth is the sum of the linewidths of the two laser systems. Assuming that the two are equal, we find that the linewidth of each ECDL measured over a time period of 2 \textmu s is about 0.3 MHz. This narrow linewidth shows the advantage of using such systems for high-resolution spectroscopy and other experiments in atomic physics.

RESEARCH COMMUNICATIONS Measuring the linewidth of a stabilized diode laser Lal Muanzuala, Harish Ravi, Karthik Sylvan and Vasant Natarajan* Department of Physics, Indian Institute of Science, Bengaluru 560 012, India We demonstrate a straightforward technique to measure the linewidth of a grating-stabilized diode laser system – known as an external cavity diode laser (ECDL) – by beating the output of two independent ECDLs in a Michelson interferometer, and then taking the Fourier transform of the beat signal. The measured linewidth is the sum of the linewidths of the two laser systems. Assuming that the two are equal, we find that the linewidth of each ECDL measured over a time period of 2 s is about 0.3 MHz. This narrow linewidth shows the advantage of using such systems for high-resolution spectroscopy and other experiments in atomic physics. Keywords: Diode laser, grating stabilization, interferometer, linewidth, Littrow configuration. T HE advent of diode lasers in the last couple of decades has revolutionized laser spectroscopy in atoms, and made possible several experimental studies in fields such as precision measurements1–3, laser cooling and trapping of atoms4, atomic clocks5, quantum optics6–8, and so on. This is because most experiments are done using the D lines of alkali atoms, which are in the near-infrared (IR) and hence accessible with diode lasers. In addition, alkali atoms have high vapour pressure at room temperature, so that vapour cells with sufficiently high atomic density can be used. However, in order to be useful for high-resolution atomic spectroscopy (where transitions have linewidths of a few MHz)9, the laser linewidth should be below 1 MHz. Since the linewidth of a commercial diode laser (of the kind that is used in CD players, for example) is of the order of a few gigahertz, it is necessary to reduce this linewidth. The required reduction is typically achieved using optical feedback from a diffraction grating, in what is called the Littrow configuration. This also serves the purpose of making the frequency of the laser tunable by changing the angle of the grating. The grating is usually mounted on a piezoelectric transducer so that the angle can be changed electronically. The configuration, shown schematically in Figure 1, is arranged so that the –1th diffraction order is fed back to the laser, while the specular reflection is the output. From the grating equation, we have *For correspondence. (e-mail: vasant@physics.iisc.ernet.in) CURRENT SCIENCE, VOL. 109, NO. 4, 25 AUGUST 2015 2d sin  = m, where d is the spacing between the successive lines of the grating, and  is the angle of the mth-order diffraction. Since the specular reflection is the output beam, it is convenient to have  close to 45. This is achieved by choosing d appropriately, e.g. the grating for accessing the D lines of K, Rb and Cs (770–900 nm) has 1800 lines/mm. The power available after optical feedback is usually about 70% of the open-loop power. Thus, although the linewidth reduction of the diode laser is by more than a factor of 1000, the loss in power is only 30%, showing that this is not wavelength selection (as for a grating used to disperse the light from a white-light source), but actual reduction in wavelength uncertainty. In effect, the grating along with the back facet of the diode forms a second lasing cavity – which is why this configuration is called an external cavity diode laser (ECDL) – with the longer cavity resulting in a smaller linewidth. The linewidth of the laser can be measured in a Michelson interferometer, with the requirement that the phase difference between the two arms be larger than the phase coherence of the laser 10. If the linewidth of the laser is about 1 MHz, the corresponding coherence length is 300 m. This means that the two arm-lengths of the interferometer have to differ by a kilometer or more. This is not easy to implement in the laboratory unless one uses a coiled optical fibre of that length. An alternate (and easier) way would be to use two identical laser systems and interfere/beat them in the interferometer. Since the phase of the two lasers is independent, the two arm lengths can be nominally equal, with the understanding that the beat signal will represent the convolution of the two laser linewidths. If we assume that the two ECDLs have Lorentzian distribution with centre frequencies 1 and 2 , and linewidths (full-width-at-half-maximum, FWHM) Figure 1. Diode laser stabilization in Littrow configuration. Optical feedback from a grating is used to reduce the linewidth of the laser. The grating is mounted on a piezoelectric transducer (PZT) to enable electronic tuning of the wavelength. This configuration is called an external cavity diode laser (ECDL). 765 RESEARCH COMMUNICATIONS 1 and 2 respectively, then the normalized distribution in frequency-space is Li ( )  i / 2 1 ,  (   i )2  (i / 2)2 (1) where i is 1 or 2 for the two laser systems. In this study, we present the results of such a linewidth measurement on two home-built ECDLs beat over a time period of 2 s. As expected, the linewidth of each laser is below 0.5 MHz. To observe if there is an effect of locking the frequency of a laser, we have done three studies – (i) both lasers free-running, (ii) one laser locked and the other free-running, and (iii) both lasers locked. The results indicate that locking lasers has no effect on the linewidth, at least over this timescale. The diode laser system consists of a Sharp laser diode (GH0781JA2C) operating with a free-running wavelength of 784 nm and power of 120 mW. The laser is stabilized using feedback from an angle-tuned grating with 1800 lines/mm (Figure 1). The system is mounted on a thermo-electric cooler for temperature stabilization. Using a combination of operating temperature and operating current, the laser system is brought near the Rb D2 line (5S1/2  5P3/2 transition) at 780 nm. Part of the laser output is fed into a saturated absorption spectrometer (SAS)10, so that the laser can be locked to a hyperfine transition, if necessary. The locking is achieved by modulating the laser current at 20 kHz, and demodulating the SAS signal using a lock-in amplifier. Figure 2 shows the experimental set-up for obtaining the beat signal. The two lasers have a fixed frequency difference of about 16 MHz. This is so that the beat signal is centred around a non-negative value, and the variation around this value can be measured unambiguously. By contrast, if the frequency difference is zero, the Figure 2. Experimental schematic for beat signal measurement.  /2, Halfwave retardation plate; PBS, Polarizing beam splitter cube; BS, Beam splitter; PD, Photodiode. 766 lineshape would be a half-Lorentzian function, because only positive frequencies would appear in the spectrum. The output of the two ECDLs is mixed on a 50–50 nonpolarizing beam splitter, as shown in Figure 2. The beat signal is measured on a fast photodiode, with response time sufficiently fast in order to measure the 16 MHz signal. The signal is measured at a sampling rate of 1 GHz for a total time of 2 s, corresponding to 2000 points. A fast Fourier transform (FFT) of the signal gives the frequency spectrum, with sufficient zero padding to make the spectrum smooth. Before turning to the experimental results, we see what is the expected lineshape for the beat signal. A Lorentzian centred at non-zero frequency can be simulated using a function of the form f (t) = e–2  t/2 cos(2 f0t), (2) where  is the linewidth and f0 is the centre frequency. We take typical values of  = 0.6 MHz and f0 = 16 MHz. Using experimental values of 2000 samples at a sampling rate of 1 GHz and total time of 2 s, the FFT of this function (magnitude squared with zero padding of 100,000 points) is shown in Figure 3. The lineshape is essentially the convolution of a Lorentzian with a sync function (because of the finite time duration of the function) as evidenced by the zeros of the spectrum. The line-shape near the peak is mainly Lorentzian as seen from the nearperfect overlap with the Lorentzian fit, and the linewidth obtained from the fit is 0.63 MHz – close to the chosen value of 0.6 MHz. Therefore, in the following, the experimentally measured spectrum is fit to a Lorentzian, and the linewidth determined from the fit. Figure 4 shows a typical experimental FFT spectrum obtained with two free-running lasers. The data are taken Figure 3. Calculated power spectrum for a function given by eq. (2) with centre frequency 16 MHz, linewidth 0.6 MHz and lasting for a time of 2 s (shown with a dotted line). The solid line is a Lorentzian fit to the central peak, which matches the spectrum almost perfectly and yields a linewidth of 0.63 MHz. CURRENT SCIENCE, VOL. 109, NO. 4, 25 AUGUST 2015 RESEARCH COMMUNICATIONS Figure 4. Experimental power spectrum obtained by beating two ECDLs. A Lorentzian fit (not shown) yields a linewidth of 0.64 MHz. Table 1. Measured linewidths of the beat signal under different conditions of locking of the two external cavity diode lasers (ECDLs). Listed is the average value from three measurements Condition of ECDLs Both free-running One locked and one free-running Both locked Average linewidth (MHz) 0.57 0.60 0.58 at a sampling rate of 1 GHz and for a total time of 2 s – exactly the conditions used for the theoretical results presented in Figure 3. It has a similar lineshape with zero points due to the finite signal duration. A Lorentzian fit to the central peak yields a linewidth of 0.61 MHz. If we assume that the two lasers are identical, then eq. (1) shows that the linewidth of each laser is 0.3 MHz. Since this is the linewidth obtained after 2 s, it can be regarded as an average linewidth over this period. The instantaneous linewidth is expected to be lower. In order to study the effect of locking the laser to a hyperfine transition, we have repeated the above experiment with one of the lasers locked and also with both the lasers locked. Table 1 lists the results for all three sets. Each measurement was repeated three times and what is listed is the average value. There is not much difference in the values, which indicates that the linewidth does not change because of locking, at least over the 2 s timescale of the measurement. We expect that the effect of locking will be to prevent long-term drift of the laser frequency, which can be important in experiments like laser cooling and quantum optics. Noting that there is not much change in the average value over the three sets, and in order to have sufficient points to get a meaningful standard deviation, all nine measurements were combined into one set and the standard deviation calculated for the entire set – this value is CURRENT SCIENCE, VOL. 109, NO. 4, 25 AUGUST 2015 0.059 MHz, and can be regarded as an error bar on the linewidth measurement. In conclusion, we have presented a technique where the linewidth of a grating-stabilized diode laser can be measured using a Michelson interferometer. Instead of the usual technique of having a kilometre long fibre in one of the arms of the interferometer to create the required phase delay, we use the simpler technique of having two independent diode lasers with nominally equal path lengths in the two arms. If we assume that the linewidths of the two laser systems are equal, we find that the linewidth averaged over 2 s is about 0.3 MHz. This shows the advantage of using such stabilized diode laser systems (ECDLs) for high-resolution spectroscopy and other experiments in atomic physics. 1. Banerjee, A., Das, D. and Natarajan, V., Precise frequency measurements of atomic transitions by use of a Rb-stabilized resonator. Opt. Lett., 2003, 28, 1579–1589. 2. Das, D. and Natarajan, V., Hyperfine spectroscopy on the 6P3/2 state of 133 Cs using coherent control. Europhys. Lett., 2005, 72, 740–746. 3. Ravishankar, H., Chanu, S. R. and Natarajan, V., Chopped nonlinear magneto-optic rotation: a technique for precision measurements. Europhys. Lett., 2011, 94, 53002. 4. Rapol, U. D., Wasan, A. and Natarajan, V., Loading of a Rb magneto-optic trap from a getter source. Phys. Rev. A, 2001, 64, 023402. 5. Brandt, S., Nagel, A., Wynands, R. and Meschede, D., Buffer-gasinduced linewidth reduction of coherent dark resonances to below 50 Hz. Phys. Rev. A, 1997, 56, R1063–R1066. 6. Rapol, U. D., Wasan, A. and Natarajan, V., Observation of subnatural linewidths for cold Rb atoms in a magneto-optic trap. Europhys. Lett., 2003, 61, 53–59. 7. Iftiquar, S. M., Karve, G. R. and Natarajan, V., Subnatural linewidth for probe absorption in an electromagnetically inducedtransparency medium due to Doppler averaging. Phys. Rev. A, 2008, 77, 063807. 8. Chanu, S. R., Pandey, K., Bharti, V., Wasan, A. and Natarajan, V., Polarization-rotation resonances with subnatural widths using a control laser. Europhys. Lett., 2014, 106, 43001. 9. Das, D. and Natarajan, V., High-precision measurement of hyperfine structure in the D lines of alkali atoms. J. Phys. B, 2008, 41, 035001. 10. Born, M. and Wolf, E., Principles of Optics: Electromagnetic Theory of Propagation, Interference and Diffraction of Light, Cambridge University Press, Cambridge, 1999, 7th edn. ACKNOWLEDGEMENT. This work was supported by the Department of Science and Technology, Government of India. Received 13 April 2015; revised accepted 16 May 2015 767