Journal of Archaeological Science 53 (2015) 1e13
Contents lists available at ScienceDirect
Journal of Archaeological Science
journal homepage: http://www.elsevier.com/locate/jas
Portable XRF analysis of archaeological sediments and ceramics
Alice M.W. Hunt*, Robert J. Speakman
Center for Applied Isotope Studies, University of Georgia, United States
a r t i c l e i n f o
a b s t r a c t
Article history:
Received 20 May 2014
Received in revised form
6 November 2014
Accepted 22 November 2014
Available online 2 December 2014
Recently, there has been significant interest in the use of portable X-ray fluorescence spectrometers
(pXRF) for cultural materials applications, especially ceramics and sediments. Although modern pXRF
spectrometers have lower detection limits and better resolution than those of decades past, portable
instruments remain subject to the same limitations as bench-top ED-XRF instruments, particularly with
respect to sample preparation, instrument calibration, and ability to accurately quantify low-Z elements.
In this paper, we evaluate the strengths and limitations of pXRF analysis for the quantitative compositional analysis of archaeological ceramics and sediments and propose an analytical protocol and calibration designed to optimize pXRF performance for these materials.
© 2014 Elsevier Ltd. All rights reserved.
Keywords:
pXRF
Sediments
Ceramics
Phosphorus
Calibration
Analytical protocol
1. Introduction
Bulk chemical characterization of archaeological materials, such
as ceramics and sediments, is important for understanding the
human past, from determining raw material provenance to understanding economic organization and trade networks to evaluating use of space and assigning activity areas. Compositional
analysis by neutron activation (INAA), inductively coupled plasma
mass spectrometry (ICP-MS) and X-ray fluorescence (XRF) can be
expensive, destructive and/or time consuming. It is not surprising,
therefore, that non-destructive compositional analyses by portable
XRF (pXRF) have become increasingly popular.
Manufacturers of pXRF spectrometers advertise the ability of
these instruments to accurately and precisely quantify the chemical
composition of a range of materials, from metals to sediments,
‘right out of the box’ using any one of a number of factory calibrations and analytical protocols. However, these factory calibrations are generally not appropriate for archaeological materials
analysis; the calibrations are not matrix matched for archaeological
materials and rarely contain all the elements of interest and/or an
adequate elemental dynamic range suitable for archaeological
materials characterization.
* Corresponding author. Tel.: þ1 706 542 2143; fax: þ1 706 542 6106.
E-mail address: ahunt@uga.edu (A.M.W. Hunt).
http://dx.doi.org/10.1016/j.jas.2014.11.031
0305-4403/© 2014 Elsevier Ltd. All rights reserved.
pXRF studies of archaeological materials using this ‘black box’
approach rarely generate high quality, accurate compositional data.
Partly, this results from the use of pXRF calibrations and analytical
protocols developed by manufacturers for non-archaeological
materials and partly from a failure by the end-user to recognize
or understand the limitations of the instrument and/or material of
interest. In this paper, we discuss the optimization of pXRF performance for the compositional characterization of archaeological
ceramics and sediment, including a matrix matched calibration and
analytical protocol developed by the Center for Applied Isotope
Studies (CAIS), University of Georgia.
2. Portable XRF of archaeological sediments and ceramics
2.1. Analytical limitations
ED-XRF analyses of archaeological sediments and ceramics are
typically designed to identify activity areas and determine raw
material provenance. As such, the elements of interest include lowZ elements, sodium (Na) to titanium (Ti), manganese (Mn) and iron
(Fe), often referred to as the major elements, typically reported as
oxide wt.%, and minor and trace elements reported in ppm. Not all
of the elements of interest can be excited and/or measured by EDXRF spectrometry; some elements, such as the low-Zs, can only be
measured imperfectly and semi-quantitatively on conventional
lab-based ED-XRF instruments. Elements typically analyzed by ED-
2
A.M.W. Hunt, R.J. Speakman / Journal of Archaeological Science 53 (2015) 1e13
XRF for archaeological sediments and ceramics are highlighted in
Fig. 1.
Portable ED-XRF spectrometers are smaller, more compact, and
oftentimes, less powerful than conventional lab-based ED-XRF
spectrometers. Despite the point-and-shoot nature of handheld
pXRF spectrometers, these instruments are subject to the same
analytical limitations as bench-top ED-XRF spectrometers. In
addition to possessing less powerful X-ray tubes in terms of current
and often voltage, decreasing the range of elements which can be
optimally excited, pXRF spectrometers cannot generate a full vacuum, further reducing their ability to detect and quantify low-Z
elements (see discussion in 2.3). Furthermore, non-destructive
pXRF analysis of unprepared archaeological ceramics and sediments not only goes against conventional XRF wisdom but, this
analytical short-cutting, introduces matrix effects and chemical
contamination, from a variety of processes, which seriously limit
pXRF spectrometry for unprepared archaeological specimens.
As discussed in Section 2.4, under the right conditions, a pXRF
spectrometer can perform as well as a bench-top ED-XRF for most
elements of interest for archaeological ceramic and sediment
characterization studies. The notable exceptions and pXRF specific
limitations are described below.
2.1.1. Low-Z elements
WD-XRF is the best XRF approach for measuring and quantifying low-Z elements. ED-XRF spectrometers are subject to limitations, such as Bremsstrahlung radiation and escape peaks, which
affect detection and quantification of low-Z elements in particular.
Although we will demonstrate in Section 2.3 that, under the right
conditions, pXRF spectrometers can simulate the performance of
bench-top ED-XRF instruments, there are two notable exceptions: P
and Na.
Phosphorus, typically reported as phosphorus pentoxide (P2O5),
is a low-Z element (Z ¼ 15) commonly reported in pXRF characterization studies of archaeological ceramics and sediments. In
ceramics and non-anthropogenic sediments, P is difficult to detect
and quantify even in full vacuum, on a prepared sample, using a
bench-top ED-XRF spectrometer (Fig. 2). Several factors contribute
to the poor detection of P.
First, archaeological ceramics and non-anthropogenic sediments typically contain relatively low concentrations of P (<1 wt.%).
This means that the number of characteristic X-ray generated is
correspondingly small. In addition, the characteristic X-rays of P are
low energy, with K lines at 2.013 and 2.142 keV, which means that
they are readily (re)absorbed into the sample matrix and/or the
detector and scattered as Bremsstrahlung radiation. Therefore,
fewer P X-rays reach the detector than are excited, reducing the
observed intensity of response for P. In combination, these two
factors contribute to a P response too low for most detectors to
accurately differentiate counts and background (Fig. 3).
A further complication is that the Ca escape peak at 1.950 keV
creates a significant shoulder to the left of the P Ka line. In many
pXRF spectrometers, a portion of the Ca escape peak X-rays are
measured as P due to the inability of the software to deconvolute
the two peaks. This inability results in false counts being reported
for the P peak where none exist and leads to the calculation of false
P concentrations (Fig. 4). In the example in Fig. 4, samples MS15BL1-19-IVSD and MS15SDO-1-IVSD have similar P concentrations,
0.114 and 0.138 wt.% respectively, but significantly different Ca
concentrations, ca. 19 and 0.7 wt.% respectively. Table 1 reports the
raw and net counts for these two samples measured using the same
pXRF spectrometer and analytical conditions. The Ca escape peak
adds almost 24 the number of processed counts recorded as P.
These results, in combination with the low concentrations of P in
archaeological ceramics and non-anthropogenic sediments, suggest that P cannot be accurately measured by pXRF and that concentrations for P < 1 wt.% reported by pXRF should be treated with
extreme caution.
As discussed, pXRF spectrometers cannot measure P at the
concentrations typically found in archaeological ceramics and nonanthropogenic sediments. The same is also true for Na. Sodium Xrays are extremely low energy, Ka line at 1.041 keV, and so, to an
even greater extent than other low-Z elements, are (re)absorbed
into the sample matrix and/or detector and scattered as
Fig. 1. Elements of interest for the analysis of archaeological ceramics and sediments typically analyzed by XRF.
A.M.W. Hunt, R.J. Speakman / Journal of Archaeological Science 53 (2015) 1e13
3
Fig. 2. Biplots of measured vs expected concentrations of phosphorus for 20 CRMs analyzed by WD-XRF (blue triangle), ED-XRF (green circle), and pXRF (red square). (For
interpretation of the references to color in this figure legend, the reader is referred to the web version of this article.)
Fig. 3. pXRF spectrum illustrating the lack of response (recorded counts) for sodium and minimal response for P even under optimal analytical conditions.
4
A.M.W. Hunt, R.J. Speakman / Journal of Archaeological Science 53 (2015) 1e13
Fig. 4. pXRF spectrum illustrating the spectral overlay interference created by the escape peak of Ca onto the Ka line of P.
Bremsstrahlung radiation. Even at relatively high concentrations,
ca. 3.5 wt.%, few Na X-rays reach and are counted by the detector; at
the low concentrations typical of archaeological ceramics and
sediments (3 wt.%), the response for Na is virtually invisible by
pXRF. For example, in Fig. 3, the pXRF spectrum for SARM 69, with a
recommended Na2O concentration of 0.79 wt.%, has no visible peak
at 1.041 keV. Therefore, like P, Na cannot be measured by pXRF at
the concentrations typically found in archaeological ceramics and
sediments.
All other low-Z elements have the potential to be measured and
quantified as well by pXRF as by a bench-top ED-XRF (Fig. 8).
2.1.2. Low/mid-Z trace elements
Of the low/mid-Z trace elements (Z ¼ 21e30), vanadium (V),
chromium (Cr), cobalt (Co) and nickel (Ni) can only be measured
semiquantitatively by pXRF due to spectral overlay interference
and the low concentrations of these elements typical in archaeological ceramics and sediments. As illustrated in Fig. 5, detection
and quantification of Ti, V, and Cr is complicated by spectral overlapping of the Kb line of the lower Z element and the Ka line of the
next higher Z element, for example V Kb line overlays the Ka line of
Cr. Additionally, in high iron (Fe) samples the escape peak of Fe
(4.660 keV) overlaps the Ka line of Ti (4.508 keV). However, Ti can
be present in archaeological ceramics and sediments at concentrations up to 3 wt.% TiO2, whereas V and Cr are typically present
only in the low 100 s ppm. The combined effect of the low
Table 1
Concentrations of P and Ca for the two samples in Fig. 6. Notice how the Ca escape
peak is measured as P counts.
Sample
MS15BL-1-19-IVSD
MS15SDO-1-IVSD
P2O5
concentration/small number of counts for V and Cr and interference
from spectral overlay radiation is that these two elements can only
be semiquantitatively measured by pXRF (Fig. 6), whereas Ti can be
detected as well by pXRF as by a bench-top ED-XRF spectrometer
(Fig. 8).
Detection and quantification of Co also is affected by spectral
overlap interference, in this case the Kb line of Fe, and typically low
elemental concentrations in the sample material. However, for Co,
it is the latter which most dramatically limits its detection and
quantification. Although Co can occur naturally in archaeological
ceramics and sediments in concentrations as high as
200e250 ppm, it more commonly is found at concentrations
<100 ppm. At these lower concentrations (<150 ppm), the detector
is unable to differentiate Co X-rays from background radiation
(limit of detection or LOD)1 and/or the Fe Kb line, and so the
element is not measured. As illustrated in Fig. 8, at concentrations
150 ppm pXRF instrument response to Co is both linear and accurate. However, these high Co concentrations are the exception
rather than the rule in archaeological ceramics and sediments.
Detection and quantification of Ni in archaeological ceramics
and sediments by pXRF is complicated. Nickel, typically present in
archaeological ceramics and sediments at concentrations <70 ppm
in North and South America and at concentrations as high as
150 ppm in other parts of the world, can be affected by spectral
overlap radiation of the Cu Ka line onto its Kb line. However, filters
also appear to be a factor in Ni detection and quantification. As we
discuss in Section 2.4, on our pXRF the best filter for measuring and
quantifying trace elements in a clay/sediment matrix is composed
of Al, Ti and Cu. Filters of other compositions, for example in Fig. 6
the bench-top ED-XRF uses a palladium (Pd) filter and the
CaO
Concentration
Raw counts
Net counts
Concentration
0.114 wt.%
0.138 wt.%
27,040
13,824
12,000
521
19.037 wt.%
0.709 wt.%
1
Limit of Detection (LOD) is defined by the International Union of Pure and
Applied Chemistry (IUPAC) as the elemental concentration at which an instrument
in no longer able to differentiate between signal and noise (i.e. the element and
background).
A.M.W. Hunt, R.J. Speakman / Journal of Archaeological Science 53 (2015) 1e13
5
Fig. 5. pXRF spectrum illustrating overlapping K line radiation.
‘mudrock’ pXRF protocol (discussed below) uses an Al:Ti filter, do
not appear to affect the quality of Ni measurement significantly.
Although we cannot explain this phenomenon to our satisfaction,
we believe it is a filter effect and, as a result, that low concentration
Ni measurements, like Co measurements, should be considered
semiquantitative in a ceramic or sediment matrix.
Most other low/mid-Z trace elements of interest for compositional analysis of archaeological ceramics and sediments and elements with L line energies measured in the low/mid-Z range (e.g.,
Cu, Zn, Rb, Sr, Y, Zr, Nb, Th, and Pb) can be measured as accurately
and precisely by pXRF as by bench-top ED-XRF instruments (Fig. 9),
under appropriate analytical conditions.
2.1.3. High-Z trace elements
There are many higher energy elements of potential interest for
compositional analysis of archaeological ceramics and sediments,
including barium and a few lanthanide group elements. However,
barium (Ba) is really the only higher energy element that can be
measured well by ED-XRF. The K lines of Ba are 32.065 and
36.553 keV, placing their optimal excitation energies >50 kV: out of
range for many pXRF spectrometers. Ba L lines, on the other hand,
at 4.467 and 4.828 keV, are easily excited by pXRF. However, Ba L
lines overlap with the K lines of Ti and V (Fig. 7). Likewise, for the
few lanthanide group elements that it might be possible to measure
by pXRF, assuming they are present at higher concentrations, their
L lines overlap the K lines generated by many of the first row
transition metals, and/or their K lines are barely visible above
background and cannot be measured or quantified with accuracy or
precision.
2.2. Calibration and matrix matching
Laboratory based XRF spectrometers typically do not come
calibrated. Scientists operating these instruments are responsible
for developing calibrations appropriate for the sample material of
interest. Conversely, most pXRF instruments are calibrated by the
manufacturer and it is oftentimes not possible for the user to create
a matrix matched empirical calibration. Some software packages
allow users to adjust preset elemental correction factors/calibration
coefficients based on the manufacturer's measured and expected
concentrations. While modifying correction factors might improve
the accuracy of reported concentrations within a specific dynamic
range, they may or may not improve the accuracy of reported
concentrations outside that range. Alternatively, users can correct
their data offline by developing regressions based on the measurement of multiple CRMs and applying these equations to their
unknown samples. However, we underscore that the adjustment of
correction factors and/or calibration coefficients is not in any way
comparable to calibrating an instrument and/or creating a true
empirical matrix-matched calibration.
Most, if not all, pXRF manufacturers offer a ‘soils’ or sediments
mode/calibration. These ‘soils’ calibrations tend to focus on heavy
metals related to environmental issues and often do not include
either the elements of interest for archaeological ceramics and
sediments and/or the dynamic range necessary for their accurate
quantification. In addition, the details of these calibrations, such
as analytical range, matrix of the standards/reference materials
used to develop it, and even the composition of the filters, are
oftentimes considered proprietary and are not readily available to
6
A.M.W. Hunt, R.J. Speakman / Journal of Archaeological Science 53 (2015) 1e13
Fig. 6. Biplots of measured vs expected concentrations of V, Cr, Co and Ni for 20 CRMs analyzed by WD-XRF (blue triangle), ED-XRF (green circle), pXRF clay/sediment calibration
(red square), and pXRF ‘mudrock’ calibration (black diamond). (For interpretation of the references to color in this figure legend, the reader is referred to the web version of this
article.)
the user (e.g., Frahm et al., 2014). Issues with this ‘black box’
approach are myriad. In a future paper, we address in detail issues of pXRF instrument response and the accuracy and precision
of these factory calibrations for compositional analysis of
archaeological materials (Hunt and Speakman, in preparation).
For the purposes of this exercise, however, we are concerned
with the optimization of pXRF performance for archaeological
ceramics and sediments.
The three critical aspects of a calibration are: (a) that it include
all the elements of interest in the sample material; (b) that is has a
dynamic range appropriate for the elemental concentrations
typical or expected in the sample material; and (c) that the certified
reference materials (CRMs) or standards used to build the calibration have a similar matrix to the samples. The importance of each of
these factors should be self-evident and are discussed here only
briefly.
Not including fundamental parameters algorithms, an element
can only be quantified, that is concentrations calculated from
spectral counts or pulses, if there is a reference database of counts/
concentrations for that element. Although an XRF spectrum records
all the fluorescence energy emitted by a sample, without a reference calibration the analytical software cannot associate that energy with the characteristic emission of a particular element and/or
calculate the elemental concentration from the measured fluorescence. Therefore, a calibration must contain all the potential elements of interest for a given material. For archaeological ceramics
and sediments, elements of interest ‘visible’ by pXRF include
MgeSi, KeCa, TieZn, RbeNb, and Th. Other potential elements of
interest, such as P, S and Cl, are oftentimes not visible due to their
low concentrations in archaeological ceramics and sediments,
spectral overlap, and/or because of peaks resulting from the anode
material in the X-ray tube.
A.M.W. Hunt, R.J. Speakman / Journal of Archaeological Science 53 (2015) 1e13
7
Fig. 7. Biplots of measured vs expected concentrations of Ba for 20 CRMs analyzed by WD-XRF (blue triangle), ED-XRF (green circle), pXRF clay/sediment calibration (red square),
and pXRF ‘mudrock’ calibration (black diamond). (For interpretation of the references to color in this figure legend, the reader is referred to the web version of this article.)
One of the factors affecting the accuracy of reported elemental
concentrations is the dynamic or analytical range of the calibration for that element. Each CRM/standard in an elemental calibration provides the analytical software with a reference point
consisting of the number of measured or normalized counts and
the elemental concentration associated with these counts. Most
analytical software assumes a directly proportional relationship
between counts and concentration, i.e., as the number of counts
increases so does concentration. This relationship is used by the
software to create a model, usually linear, of measured counts and
concentration. When the software is presented with measured
counts for an element of unknown concentration, it uses this
model to predict where those counts are located within the parameters of the model, between the high and low reference
concentrations, and calculates the elemental concentration associated with that number of counts. Although calibrations can
calculate concentrations outside of the high-low reference range,
calculated concentrations are less accurate the farther away the
model moves from a reference standard. Therefore, it is important
when building a calibration that the dynamic range of each
elemental simulates the expected compositional range of the
sample material.
Matrix matching the standards and sample material is, perhaps,
the most important consideration when building a calibration. Xrays attenuate, refract, and are absorbed differently by different
materials in accordance with their density, chemical composition,
and crystal structure. The interaction between a material and
excitation/emission X-ray radiation affects its fluorescence
response and the detection of that response as much as, if not more
than, the accelerating voltage and current of the excitation energy
and/or detector resolution. Matrix effect(s) is the blanket term used
to explain these interactions. The three most important interactions for pXRF analysis of archaeological ceramics and sediments are: grain size effects, heterogeneity, and mineralogical
effects. Grain size effects result in the differential penetration of the
X-rays into the sample: X-rays are able to pass through smaller
grains, exciting deeper into the sample and generating a more
representative fluorescence response; larger grains may exceed the
penetration depth of the X-ray so that only a single phase in the
sample is excited. Heterogeneity of the material is one of the most
commonly discussed issues in compositional analysis and results
from the inhomogeneous mixing of phases in a materials so that
compositional analysis of the material is not ‘representative’ of its
total chemical content and/or composition of the sample varies
8
A.M.W. Hunt, R.J. Speakman / Journal of Archaeological Science 53 (2015) 1e13
Fig. 8. Biplots of measured vs expected concentrations of the low-Z elements for 20 CRMs analyzed by WD-XRF (blue triangle), ED-XRF (green circle), pXRF clay/sediment calibration (red square), and pXRF ‘mudrock’ calibration (black diamond). (For interpretation of the references to color in this figure legend, the reader is referred to the web version of
this article.)
across the sample area (see Fig. 3 in Speakman et al., 2011).
Mineralogical effects result from differences in crystal structure,
composition, and density of different mineral phases which lead to
the differential attenuation of excitation and emission X-rays and
they pass through these phases; a highly attenuating phase may
prevent excitation X-rays from exciting other phases due to loss of
energy and/or prevent lower energy emission X-rays from exiting
the sample.
Grain size and mineralogical effects and heterogeneity are
significantly reduced when the analyte is prepared as a pressed
powder pellet or fused bead. Therefore, a calibration built using
CRMs prepared as pressed pellets is only appropriate for sample
material prepared the same way: ‘matrix matched’ to reduce matrix effects. For this reason, we believe that fully quantitative
analysis of archaeological ceramics and sediments by ED-XRF can
only be conducted using pressed pellets and/or fused beads; data
resulting from the analysis of whole sherds and loose sediments
cannot be considered fully quantitative.
The procedure for making pressed pellets is simple and inexpensive: at CAIS 10 g of powdered sample (ground to approximately 10 mm or the consistency of talcum powder) is homogenized
in an agate mortar with a binding agent (we use 2 mL of Elvacite
dissolved in acetone) and pressed into a 40 mm aluminum sample
cup at 23e25 PSI for 3 min. Another benefit of pressed pellets is
that the die press ensures a uniformly flat analytical surface eliminating matrix effects related to surface topography (which are
admittedly minor in ED-XRF). All of the CRMs used in this study
were prepared as pressed pellets according to this protocol.
An important consideration for matrix matching a calibration to
the sample material is the matrix or material itself. Reliable analysis
of copper alloys requires a calibration built using copper alloy
CRMs, reliable analysis of obsidian requires a calibration built using
A.M.W. Hunt, R.J. Speakman / Journal of Archaeological Science 53 (2015) 1e13
9
Fig. 8. (continued).
obsidian CRMs, and reliable analysis of ceramic and sediments requires a calibration built using clay and sediment CRMs. Archaeological ceramics and sediments are often considered comparable to
rocks: both are formed by geologic processes, composed of discrete
mineral phases, and have similar geochemistry. However, sediments and rocks differ in several significant ways which prevent
rocks, even sedimentary rocks such as mudstones, from being a
‘matrix match’ for archaeological ceramics and sediments.
Rock is substantially denser and contains significantly less
structural water than sediments and clays. Preparing sample material as pressed pellets eliminates matrix issues related to density;
however, there remains the issue of structural water and hydrous
mineral phases. During the lithification process, sediments are
compacted, pore space is reduced, and some of the structural water
contained in the mineral phases is released. This water may carry
mineral components and/or soluble phases in solution, often
precipitating as new mineral phases in the remaining pore space
cementing the sediments into rock. Sediments and clays, on the
other hand, have not been through the lithification process and
contain their original water content both as structural water and
hydrous phases. Drying sediment samples prior to pressing them
into pellets removes water absorbed by the sediments/clay minerals but is typically at temperatures too low to drive off the
structural water. Fired clay or ceramic has typically reached temperatures high enough for long enough to cause dehydration and
dehydroxylation. However, during the subsequent cooling of the
vessel, its use and/or deposition, ceramics absorb water into their
void space and minerals rehydroxylate. Thus, archaeological ceramics and sediments have a higher structural water content than
sedimentary rocks.
Twenty CRMs (Table 2), 6 sediments and 14 clays, were prepared
as pressed pellets and analyzed using a Bruker Tracer IIIV pXRF
spectrometer and elemental concentrations were calculated using a
clay/sediment calibration developed by CAIS and a manufacturer's
10
A.M.W. Hunt, R.J. Speakman / Journal of Archaeological Science 53 (2015) 1e13
Fig. 9. Biplots of measured vs expected concentrations of the mid-Z elements for 20 CRMs analyzed by WD-XRF (blue triangle), ED-XRF (green circle), pXRF clay/sediment calibration (red square), and pXRF ‘mudrock’ calibration (black diamond). (For interpretation of the references to color in this figure legend, the reader is referred to the web version of
this article.)
recommended factory calibration for ceramics and soils, hereafter
referred to as the ‘mudrock’ calibration (Table 3). The ‘mudrock’
calibration appears to be the equivalent of what is oftentimes
referred to as the ‘soils’ mode by other manufacturers and is
Bruker's recommended method for quantitative compositional
analysis of archaeological ceramics and sediments. Therefore, all 20
CRMs were analyzed according to the manufacturer's protocol for
the ‘mudrock’ calibration and the protocol developed by CAIS for
the clay/sediment calibration (Table 4).
As illustrated in Fig. 8, the ‘mudrock’ calibration is unable to
measure magnesium (Mg), manganese (Mn), and Fe as well as the
clay/sediment calibration. We believe that for Mg this is a result of
analytical protocol: Mg is best detected in the presence of helium
(He) or full vacuum (Section 2.3). Fig. 6 demonstrates that, although
neither pXRF calibration analyses V, Cr, Co or Ni adequately, the
clay/sediment calibration generates a linear response to changes in
concentration whereas the ‘mudrock’ calibration appears to
‘collapse’. The non-linear response of the ‘mudrock’ calibration for
V, Cr, Co and Ni may result from an inadequate dynamic range for
these elements in the calibration or be caused by matrix effects. The
latter appears to be the cause of the poor performance of the
‘mudrock’ calibration for the mid-Z trace elements Rb, Sr, Y, Zr, Nb
and Th: elements typically quantified easily by pXRF.
Linearity of instrument response is an indication of how accurately the analytical software is able to calculate elemental concentrations using a particular calibration. Table 5 presents the
results of the compositional analysis of CRM SARM 69 using the
‘mudrock’ and clay/sediment calibrations. SARM 69 was selected
A.M.W. Hunt, R.J. Speakman / Journal of Archaeological Science 53 (2015) 1e13
11
Fig. 9. (continued).
because it is composed of fired potsherds and was not used to build
the clay/sediment calibration. Analyses were run in triplicate and
measurement precision for each element is reported as percent
relative standard deviation (% RSD). The ‘mudrock’ and clay/sediment calibrations have similar precision for the low-Z elements,
however, the ‘mudrock’ calibration calculates inaccurate concentrations for most elements except Mn. Performance of the two
calibrations for V, Cr, Co, and Ni is variable. The ‘mudrock’ calibration grossly underreports the Cr content of SARM 69; the clay/
sediment calibration also underreports Cr but it is an entire order of
magnitude closer to the certified value. Both calibrations underreport Ni and under/overreport V by the similar amounts. The clay/
sediment calibration calculates highly accurate concentrations for
the mid-Z trace elements performing almost identical to the benchtop ED-XRF. The ‘mudrock’ calibration also calculates accurate
concentrations for these elements, however, they are not as close to
the certified or ED-XRF values as those generated by the clay/
sediment calibration.
Table 2
The 20 CRMs used to evaluate performance of clay/sediment and ‘mudrock’ pXRF calibrations.
Clay/ceramic
Sediment
NIST 679
NIST 97b
NIST 98b
NCS DC 60102
NCS DC 60103
NCS DC 60104
NCS DC 60105
NCS DC 61101
C-137
C-138
C-139
NIST 8704
GBW 07310
GBW 07311
GBW 07312
GBW 07302
GBW 07405
NCS HC 14807
NCS HC 14808
NCS HC 14809
2.3. Helium vs. vacuum
X-rays, particularly low energy X-rays, attenuate and are
absorbed by air. Bench-top XRF spectrometers operate under full
vacuum to eliminate the interference of air with the detection of
low-Z elements. Some pXRF spectrometers have the ability to
generate partial vacuum: the space between the window and the
detector is evacuated. However, during this process, the window is
pulled into the chamber creating a slight concavity between the
window and sample surface into which air is (re)introduced to the
system.
Air is composed primarily of nitrogen (78%) and oxygen (20%)
but also contains water vapor (1%) and argon, carbon dioxide, dust
and pollen (<1%). Most of these components are ‘invisible’ by pXRF;
argon is not. In Fig. 10, the argon peak is clearly visible in the
Table 3
Clay and sediments CRMs used to build the clay/sediment pXRF
calibration at CAIS.
Clay/ceramic
Sediment
NIST 679
NIST 97b
NIST 98b
NCS DC 60102
NCS DC 60103
NCS DC 60104
NCS DC 60105
NCS DC 61101
C-137
C-138
C-139
NIST 2710
NIST 8604
GBW 07310
GBW 07311
GBW 07312
GBW 07302
GBW 07405
LKSD-1
LKSD-2
LKSD-3
LKSD-4
STSD-1
STSD-2
STSD-3
STSD-4
MESS-2
PACS-2
SARM 46
SARM 52
NCS HC 14807
NCS HC 14808
NCS HC 14809
MURR New Ohio Red
MURR Ohio Gold
MURR Talc-free
MURR Terra Cotta
12
A.M.W. Hunt, R.J. Speakman / Journal of Archaeological Science 53 (2015) 1e13
Table 4
Analytical protocols used to evaluated performance of clay/sediment and ‘mudrock’
PXRF calibrations.
CAIS Clay/Sediment calibration
Low Z elements 200 s count time
15 kV/20 mA
He flow
No filter
Mid Z elements 200 s count time
40 kV/30 mA
‘green’ filter:
300 mm Al/20 mm Ti/150 mm Cu
Bruker mudrock calibration
180 s count time
15 kV/20 mA
Vacuum
No filter
60 s count time
40 kV/30 mA
‘yellow’ filter:
300 mm Al/20 mm Ti
spectrum collected in air/no vacuum. This spectrum also illustrates
that in air, with the exception of the two low-Z elements present in
high concentration (Al and Si), no counts are recorded for elements
below the Rh Compton peak, Z < 19. Under partial vacuum, the
argon peak is no longer visible, however X-rays for elements Z < 13
are not recorded (Fig. 10). The spectrum collected with helium (He)
flowing through the chamber/window, displacing all the air
Table 5
Expected and measured concentrations for SARM 69 using INAA, WD-XRF, ED-XRF,
pXRF clay/sediment calibration and pXRF ‘mudrock’ calibration. Major elements are
reported as wt%; minor and trace elements reported as ppm unless otherwise stated.
Uncertified concentrations reported in brackets. Concentrations reported in shaded
cells are informational only. Expected and measured concentrations for major elements reported as elements were converted to oxides using the conversion factors
in Glascock (2006). These values are marked with an asterisk (*).
Element Cert
MgO
1.85
Al2O3
14.4
SiO2
66.6
K2O
1.96
Ba
518
INAA
(n ¼ 5)
14.26
1.80 %RSD
1.89
10.40 %RSD
CaO
2.37 2.40
7.80 %RSD
TiO2
0.777 0.77
8.30 %RSD
MnO
0.13 0.14
4.40 %RSD
Fe2O3 (T) 7.18 7.01
0.90 %RSD
Co
Cr
Nb
Ni
Rb
Sr
Th
V
Y
Zr
504
8.60 %RSD
26
0.90 %RSD
213
1.60 %RSD
WD-XRF
(n ¼ 10)
ED-XRF
(n ¼ 10)
pXRF
Clay/Sed
(n ¼ 3)
pXRF
mudrock
(n ¼ 3)
1.88
0.27 %RSD
14.39
0.14 %RSD
65.88
0.09 %RSD
1.97
0.21 %RSD
2.41
0.28 %RSD
0.776
0.36 %RSD
0.129
0.63 %RSD
7.33
0.14 %RSD
1.46
6.43 %RSD
13.89
0.21 %RSD
65.18
0.05 %RSD
2.04
0.80 %RSD
2.29
0.45 %RSD
0.77
0.59 %RSD
0.13
3.62 %RSD
6.78
0.26 %RSD
1.16
3.32 %RSD
14.22
0.29 %RSD
67.50
0.19 %RSD
2.07
0.40 %RSD
2.22
0.39 %RSD
0.71
0.73 %RSD
0.12
1.36 %RSD
6.48
0.47 %RSD
11.84*
2.60 %RSD
14.94*
0.38 %RSD
55.04*
0.36 %RSD
3.69*
0.33 %RSD
4.04*
0.19 %RSD
1.10*
0.77 %RSD
0.13*
1.66 %RSD
8.69*
0.20 %RSD
521
505
2.04 %RSD 2.24 %RSD
28
n. m.
9
111.65 %RSD
223
199
164
0.68 %RSD 1.81 %RSD
(9)
9
8
2.94 %RSD 6.90 %RSD
53
66
50
39
143.50 %RSD 1.20 %RSD 7.32 %RSD
(66) 68
72
62
3.90 %RSD
0.53 %RSD 3.30 %RSD
(109) 85
109
102
137.00 %RSD 0.33 %RSD 2.30 %RSD
(9)
9
n. m.
7
3.60 %RSD
0.01 %RSD
(157) 160
154
137
4.20 %RSD
2.08 %RSD 12.83 %RSD
(29)
29
29
1.50 %RSD 2.77 %RSD
254
(271) 208
260
12.00 %RSD 0.27 %RSD 3.26 %RSD
426
3.11 %RSD
<16
2216
27.58 %RSD
22
8.98 %RSD
167
81
1.61 %RSD 1.85 %RSD
9
7
4.28 %RSD 3.28 %RSD
36
41
12.47 %RSD 1.85 %RSD
61
57
3.06 %RSD 4.07 %RSD
102
107
1.22 %RSD 2.81 %RSD
8
6
8.74 %RSD 2.09 %RSD
120
190
0.98 %RSD 2.44 %RSD
30
31
1.53 %RSD 3.60 %RSD
248
268
0.16 %RSD 0.82 %RSD
between the detector and sample surface, enables the detector to
record counts for Mg (Z ¼ 12) (Fig. 10). Notice that Na is not visible
by pXRF using any of these analytical conditions.
From this, we conclude that the presence of air between the
sample surface and window under partial vacuum, while not
enough to generate a visible argon peak, is enough to interfere with
the detection of Mg Therefore, Mg concentrations calculated by
pXRF under partial vacuum should be considered suspect.
2.4. pXRF optimization for archaeological sediments and ceramics
The ability to generate accurate and reliable compositional data
for archaeological ceramics and sediments by pXRF requires a
matrix matched calibration and a material specific analytical protocol. A matrix matched calibration requires, not only that the
calibration standards and sample material be of the same type, i.e.,
both clay/sediments, but that the calibration standards and sample
material be prepared the same way, i.e., as pressed pellets. Analysis
of unprepared samples is not fully quantitative using a pXRF
spectrometer because of the heterogenous nature of ceramic and
sediment samples and matrix effects which prevent X-rays from
interacting with the unprepared sample in the same way as the
prepared calibration standards. This differential interaction or
response to the X-rays causes the analytical software to over or
undercalculate elemental concentrations in the sample material.
Archaeologists sometimes abrade the surface of ceramic sherds to
remove surface corrosion and treatments, such as slips, paints and
glazes. These ‘prepared’ surfaces are, perhaps, more representative
of the body/matrix material of the ceramic, however, they do not
necessarily result in improved analytical accuracy because they do
not address the matrix effects preventing accurate quantification of
unprepared sherds by pXRF.
Optimal performance of pXRF spectrometers for archaeological
ceramics and sediments also requires using appropriate analytical
protocol. Given the low concentrations and energies of many of the
elements of interest for archaeological ceramics and sediments, we
optimized our performance using a count time of 200 s at 15 kV/
20 mA in He to quantify low-Z elements and 200 s at 40 kV/30 mA
using an Al:Ti:Cu filter to quantify mid-Z elements. Even using a
matrix matched calibration and the above analytical protocol it is
important to remember that Na, P and Ba L lines do not provide
reliable numbers by pXRF and the analysis of V, Cr, Co and Ni appears to be semi-quantitative at best.
To summarize, optimization of pXRF spectrometers for the
quantitative analysis of archaeological ceramics and sediments
requires:
a matrix matched calibration
samples prepared as pressed pellets
He flow for low-Z element detection
an appropriate filter for mid-Z element quantification
A pXRF cannot accurately quantify Na, P, V, Cr, Co, Ni and Ba
(using L lines) in archaeological ceramics and sediments at typical
concentrations.
3. Conclusions
Commercially pXRF spectrometers, which have been around
since the 1960's, were never really designed to be replacements for
fully quantitative laboratory-based systems. Instead these instruments were designed to aid both scientists and non-scientists
in the identification of hazardous materials and heavy metals in
ground water and sediments, for quality control applications in
industrial metals, mining, and recycling settings, and/or to
A.M.W. Hunt, R.J. Speakman / Journal of Archaeological Science 53 (2015) 1e13
13
Fig. 10. pXRF spectrum comparing the detection of low-Z elements in air (green), under partial vacuum (blue) and with a helium flow (red). (For interpretation of the references to
color in this figure legend, the reader is referred to the web version of this article.)
conducted specialized studies where sampling and/or export of
samples was prohibitive. In contrast to earlier instruments which
required specialized scientific knowledge, modern pXRF instruments have become little more than point-and-shoot devices
requiring little-to-no specialized knowledge. The desire to use
these instruments as an inexpensive short cut to quantitative bulk
chemical analysis of archaeological materials, such as ceramics and
sediments, is understandable and has resulted in numerous studies
that would not have been possible otherwise. However, as discussed elsewhere (Speakman and Shackley, 2013), these analyses
must be undertaken with caution and some degree of understanding of the physics involved. Unlike obsidian, for which it is
relatively easy to determine a geologic/geographic source, provenance studies of ceramics are inherently challenging under the best
of circumstances, using some of the more powerful analytical
techniques (Hunt, 2012). It is primarily for this reason that archaeologists historically have not used XRF for provenance studies
of ceramics; despite the large number of laboratory based ED- and
WD-XRF spectrometers available at virtually every major institution, in most cases, XRF is insufficient for provenance determination of archaeological ceramics.
As demonstrated here, pXRF spectrometers, under the right
conditions, such as a matrix matched calibration, He-flow and
prepared samples, can perform similar, if not identical, to bench-
top ED-XRF spectrometers for certain elements. However, a pXRF
spectrometer cannot accurately quantify Na, P, V, Cr, Co, Ni and the L
lines of Ba in an archaeological ceramic or sediment matrix at the
concentrations in which they are typically present. As a result of
these limitations, compositional analysis of archaeological ceramics
and sediments by pXRF cannot and should not be considered a
substitute for fully quantitative analysis by WD-XRF, INAA and/or
ICP-MS.
References
Frahm, E., Schmidt, B., Gasparyan, B., Yeritsyan, B., Karapetian, S., Meliksetian, Kh,
Adler, D.S., 2014. Ten seconds in the field: rapid Armenian obsidian sourcing
with portable XRF to inform excavations and surveys. J. Archaeol. Sci. 41,
333e348.
Glascock, M., 2006. Tables for Neutron Activation Analysis, sixth ed. University of
Missouri, Columbia, MO.
Hunt, A.M.W., 2012. On the origin of ceramics: moving toward a common understanding of ‘provenance’. Archaeol. Rev. Camb. 27, 85e97.
Hunt, A.M.W., Speakman, R.J., Comparison of five portable X-ray Fluorescence
(pXRF) spectrometers for archaeological and culture history applications:
ceramics and sediments (in preparation)
~ an
~ ez, J.G., 2011. Sourcing ceSpeakman, R.J., Little, N.C., Creel, D., Miller, M.R., In
ramics with portable XRF spectrometers? a comparison with INAA using
Mimbres Pottery from the American Southwest. J. Archaeol. Sci. 38, 3483e3496.
Speakman, R.J., Shackley, M.S., 2013. Silo science and portable XRF in archaeology: a
response to Frahm. J. Archaeol. Sci. 40, 1435e1443.