Translational Development Strategy for Magnetic Seizure
Therapy
Stefan Rowny, MD1, Karla Benzl1,2, and Sarah H. Lisanby, MD1
Stimulation and Therapeutic Modulation Division, Department of Psychiatry, Columbia
University College of Physicians and Surgeons/New York State Psychiatric Institute, New York,
NY, USA
1Brain
2New
York University Medical School, New York, NY, USA
Abstract
NIH-PA Author Manuscript
Electroconvulsive therapy (ECT) has unparalleled antidepressant efficacy, but its cognitive side
effects may be persistent. Research suggests that the side effects may be at least partially
dissociable from the therapeutic effects of ECT, suggesting that distinct cortical networks may
underlie them and introducing a role for focal seizure induction as a means of minimizing side
effects. In magnetic seizure therapy (MST), magnetic fields avoid tissue impedance and induce
electrical currents confined to superficial cortex, facilitating focal seizure induction. The
translational development strategy for MST has included: (1) device development, (2) feasibility
in animals and initial human trials, (3) testing in nonhuman primates on safety and mechanisms of
action (with neuroanatomical, neurophysiological and cognitive endpoints), (4) safety testing in
patients, (5) initial efficacy testing in patients, (6) dosage optimization, and (7) randomized
comparison with ECT. These stages have been iterative, with results of early clinical testing
prompting device enhancements that were, in turn, tested in nonhuman primates prior to human
trials. Safety testing was aided by development of a nonhuman primate model of human ECT, and
the validation of a cognitive battery for the monkey that is sensitive to the range of effects of ECT
on human memory. Human testing has been facilitated by the development of an international
consortium of centers addressing various aspects of technique and dose/response relationships.
Challenges facing MST are common to other device based therapies: characterizing dose/response
relationships, optimizing efficacy, and developing efficient and reliable methods to induce lasting
therapeutic change in the circuitry underlying depression.
NIH-PA Author Manuscript
Keywords
Magnetic Seizure Therapy (MST); depression; electroconvulsive therapy (ECT); repetitive
transcranial magnetic stimulation (rTMS); seizure
Corresponding Author: Stefan Rowny, MD, Brain Stimulation and Therapeutic Modulation Division, Columbia University College of
Physicians and Surgeons, New York State Psychiatric Institute, 1051 Riverside Drive, Unit 21, New York, NY 10032, Office:
212-543-6055, Fax: 212-543-4340, srowny@columbia.edu, srowny@gmail.com.
Publisher's Disclaimer: This is a PDF file of an unedited manuscript that has been accepted for publication. As a service to our
customers we are providing this early version of the manuscript. The manuscript will undergo copyediting, typesetting, and review of
the resulting proof before it is published in its final citable form. Please note that during the production process errors may be
discovered which could affect the content, and all legal disclaimers that apply to the journal pertain.
Rowny et al.
Page 2
I. Introduction
NIH-PA Author Manuscript
This manuscript reviews a new investigational therapeutic intervention, Magnetic Seizure
Therapy (MST) and the translational strategy that has been applied to its development. This
novel brain stimulation modality has its roots in electroconvulsive therapy (ECT), which
was introduced to clinical practice in 1938 by Drs. Cerletti and Bini in Italy, and transcranial
magnetic stimulation (TMS), which became widely used as a research tool nearly half a
century after ECT. TMS is a non-invasive, focal means to stimulate the superficial cerebral
cortex, but at higher dosages, it can induce seizures. MST refers to the intentional induction
of a seizure for therapeutic purposes using repetitive transcranial magnetic stimulation
(rTMS). Thus, the devices that are used for MST are more powerful than the devices used in
rTMS, in that they are capable of sustaining long stimulation trains at high frequencies for
the purpose of controlled seizure induction in patients under anesthesia. The aim of this
experimental convulsive treatment is to achieve more focal seizure induction that would
retain the proven efficacy of ECT in the treatment of depression, while reducing the
cognitive side effects typically associated with ECT. The approach to realize this goal has
been to take advantage of the focality of magnetic fields to more precisely target the brain
circuitry believed to be involved in the pathogenesis of depression, while sparing the neural
circuitry implicated in the cognitive side effects.
NIH-PA Author Manuscript
While ECT remains the most effective and rapidly acting antidepressant available, with an
unparalleled spectrum of applications spanning depression, psychosis, mania, catatonia, and
movement disorders, the mechanisms of action of ECT and the reason for its unmatched
efficacy remain a topic under study. In part, this may be because the discovery that seizures
were therapeutic was serendipitous: literally decades of its clinical use have been needed to
define its therapeutic spectrum and start unraveling its mechanisms. In contrast,
pharmaceutical agents today are subjected to systematic animal testing and staged human
investigation to determine dosimetry, efficacy, and side effects prior to routine clinical use.
This rational approach to intervention development is only now beginning to be applied to
device-based therapies in psychiatry, such as TMS and MST. Here we summarize the
translational development strategy as applied to MST, and discuss implications for future
studies with this and other device-based therapeutic interventions in psychiatry.
II. Physiologically-Informed Device Development
NIH-PA Author Manuscript
The introduction of ECT transpired at least three decades prior to the initiation by various
countries of mandated, step-wise staged testing of new pharmacological treatments. This is
now widely accepted as the standard procedure for the development of new therapeutic
drugs, with Stage I- through Stage IV clinical trials typically being conducted before any
new treatment reaches the wide market. Because ECT was introduced before the US Food
and Drug Administration (FDA) implemented requirements for testing the safety and
efficacy of medical devices, it was not subjected during its early development to the types of
systematic testing that are the norm today. Decades of clinical use eventually revealed that
ECT was more effective for depression than for schizophrenia (its original therapeutic
target). Clinical studies began to optimize the procedure to reduce its adverse side effects,
such as the introduction of anesthesia and engineering changes in electrical pulse shape (sine
wave, to brief pulse, to ultrabrief pulse), modifications in electrode placement (bilateral,
right unilateral, bifrontal), and dosage relative to seizure threshold. Eventually these
discoveries informed the re-design of the devices, but these improvements took decades to
achieve and some are only now entering routine clinical practice outside of academic
centers.
Exp Neurol. Author manuscript; available in PMC 2010 December 4.
Rowny et al.
Page 3
NIH-PA Author Manuscript
The recent approach to rational drug discovery, on the other hand, is taking a more targeted
approach. Increasingly, pharmacological medications are being created to interact with
specific targets based upon knowledge of pathophysiology and molecular pathways. This
rational approach to intervention development could likewise be applied to device
development. The first step would be the identification of a clinically salient physiological
target. Next would be the development of a purpose-built device designed to reach that
physiological target. The ability of the device to achieve the physiological target would then
be tested in pre-clinical and clinical studies. These results would be used to refine the design
of the device, and eventually the refined device would be subjected to step-wise clinical
testing of its safety and efficacy.
NIH-PA Author Manuscript
This staged, rational developmental approach has been applied to MST (as illustrated in
Figure 1). The MST device was designed to reach a specific therapeutic target, guided by the
aspects of ECT seizure induction that are postulated to be therapeutic (focal frontal seizure
induction) and those postulated to be related to cognitive side effects (medial temporal
structures). A purpose-built device was developed to achieve reliable seizure induction
under anesthesia; that prototype device was tested in its ability to achieve this physiological
target in a nonhuman primate model and in initial trials in humans, its safety and efficacy
were piloted, and results informed the redesign of the device to achieve higher output. As
reviewed below, work with MST originated with animal studies and has progressed through
stages of testing in animals and humans that have resulted in iterative improvements in
device design.
III. Rationale for MST
ECT has unparalleled efficacy in the treatment of severe major depression, but its cognitive
side effects limit its use (Lisanby, et al., 2000). Extensive research has been conducted in an
attempt to improve the risk to benefit ratio of ECT by altering the process of seizure
induction, such as electrode placement and electrical parameters of stimulation (Lisanby,
2007). Manipulating these parameters has lessened its side effects, but amnesia is still a
major concern for some patients, and may be longer lasting than previously thought
(Sackeim, et al., 2007).
NIH-PA Author Manuscript
There is evidence that the efficacy and side effects of ECT can be at least partly dissociable,
as demonstrated by the discovery that high dose right unilateral (RUL) ECT has a
comparable efficacy to bilateral (BL) ECT, with a significant decrease in amnesia (Sackeim,
et al., 2000). Furthermore, the topographic EEG changes observed during ECT differ from
the EEG changes associated with clinical response (Sackeim, et al., 2000). These findings
suggest that distinct networks may be associated with ECT efficacy and side effects, and that
there may be a role for focal seizure induction to exploit this separation of networks.
Research suggests that retrograde amnesia may be related to seizure activity in the medial
temporal lobe (Luber, et al., 2000) and consequently, it was proposed that techniques
inducing seizure with reduced current spread to this area may cause less retrograde amnesia
(Lisanby, 2002, White, et al., 2006). The medial temporal structure specifically associated
with amnesia is the hippocampus, which has been shown to be uniquely affected by induced
seizures, and to have a low seizure threshold (Lisanby, et al., 2000).
Magnetic stimulation is more focal than direct electrical stimulation because it avoids the
impedance of the scalp and skull and results in an induced electric field confined to
superficial cortex. This gives more control over the induced electricity and the resultant
seizure, and presents the possibility to induce focal seizures in targeted regions while
sparing others. Instead of electrical stimulation via metal electrodes in conductive contact
with the external surface of the head as used in ECT, the MST device uses a magnetic field
Exp Neurol. Author manuscript; available in PMC 2010 December 4.
Rowny et al.
Page 4
NIH-PA Author Manuscript
with changing flux, which passes unimpeded through the skull and scalp to induce electric
current in targeted, specific regions of the cortex. In ECT-induced seizures, control over
current density and distribution is limited by impedance of the skull and the resulting
shunting of electrical stimulus. Thus, magnetic stimulation has allowed more control over
current paths and current density within cortical tissue (Lisanby, et al., 2003).
IV. Intervention Development Platform for MST
The intervention development platform for MST has been supported by an interdisciplinary
collaboration, requiring expertise in three core domains: (1) technology development, (2)
preclinical testing, and (3) clinical trials. (1) Technology development includes the
engineering aspects of device design, modeling (of the circuit topology, coil design, and
magnetic and electric field distribution), and in vitro testing (including bench testing and
phantom measurements). (2) Preclinical testing includes establishing an animal model,
physiological studies (on the electric field distribution and seizure topography), behavioral
effects (including cognitive side effects), and neuroanatomical effects (for demonstration of
safety and also for mechanisms of action). (3) Clinical testing starts with proof of concept
studies, followed by dose-finding studies, randomized controlled trials, and replication
studies at other centers.
NIH-PA Author Manuscript
Feedback from these 3 core domains has influenced iterative MST device improvements.
With the results from each domain, we went back to the manufacturer with a new set of
requested specifications for device performance and output. The preclinical trials used
animal models to characterize the physiology of MST-induced seizures, the cognitive side
effects of MST versus ECT, and the safety of MST. The clinical trials have demonstrated the
safety and feasibility of MST, shown evidence of superior side effect profile, and promising
preliminary evidence for efficacy. These early trials, however, also suggested that the MST
device had limited output, and that its efficacy as a therapeutic treatment may be doserelated. As a result, the early-stage devices that produced seizures with 50Hz stimulations
applied at 100% power output for 8 seconds, had to be modified through interdisciplinary
collaboration and in collaboration with the device manufacturer, to yield a higher-dose
stimulation consisting of 10 second10-second duration of 100 Hz at 100% power output
(See Figure 2).
V. Technology Development
NIH-PA Author Manuscript
MST is administered to anesthetized subjects via a custom modified rTMS device that
delivers stimuli at convulsive levels. The greatest challenge to early prototypes of an MST
device design has been overcoming the anticonvulsant properties of anesthesia. Using
anesthetic agents that have a lesser effect on seizure threshold (e.g., etomidate) has reduced
this problem somewhat (White, et al., 2006). Further development of the MST device has
involved careful consideration of coil design, stimulus parameters, power requirements,
heating, and pulse characterization.
Different coil designs have different magnetic field distributions, resulting in variation in
seizure threshold, seizure onset, and patterns of seizure spread. We have found the non-focal
round coil and the more focal double-cone coil to be reliable in seizure induction, whereas
the most focal figure-8 coil has been inefficient in inducing seizure (See Figure 3) (Lisanby,
et al., 2003,Lisanby, et al., 2001). Studies are currently exploring how technological changes
to coil design can increase the depth of penetration and produce seizure induction from
structures deep to the outer surface of the cortex (Deng, et al., 2008,Zangen, et al., 2005).
Another potential coil modification is the addition of a metal core that would intensify field
strength and allow lower power requirements for the device (Epstein and Davey, 2002).
Exp Neurol. Author manuscript; available in PMC 2010 December 4.
Rowny et al.
Page 5
NIH-PA Author Manuscript
Heating of the coil and other device components occurs during MST, and must be properly
handled. These constraints have been lessened through the selection of appropriate heattolerant materials and pre-cooling the coil (Lisanby, 2007). The EEG electrodes can also
heat up during MST treatment, and nonmetallic EEG electrodes are now available for use.
Furthermore, it has also been noted that briefer pulses produce less coil heating, since less
energy is needed to cause neuronal depolarization (Jalinous, 2002, Lisanby, 2007,
Peterchev, et al., 2007, Peterchev, et al., 2008).
MST devices have larger electrical supply demands than the standard rTMS device, but
improvements in design have made the current model more compact and practical in a
clinical setting (Lisanby, 2007). Presently, MST devices generate biphasic sinusoidal
magnetic pulses because they have higher power efficiency and less coil heating than do
monophasic devices (Jalinous, 2002). Studies are currently investigating the relative
effectiveness of different TMS pulse characteristics, such as pulse shape and width.
Adjusting pulse parameters may improve the efficacy of rTMS, and consequently, MST
(Peterchev, et al., 2007, Sommer, et al., 2006, Tings, et al., 2005). Indeed, we recently
reported that unidirectional pulses are more efficient in inducing seizure with ECT
(Spellman, et al., 2009).
NIH-PA Author Manuscript
New device designs have increased stimulus output and enabled longer trains of stimulation.
Based on experience with ECT, it was suggested that devices which sustain long trains of
stimulation may be more efficient when compared to those that generate excessively high
frequency for short periods of time (Lisanby, 2007). This lead to the current MST device,
which sustains peak output at 10 seconds as opposed to 8 seconds. The newer model has
also increased output from 50 Hz to 100 Hz at 100% stimulus intensity. The output was
increased to ensure that the MST device could overcome the anticonvulsant effects of
anesthesia at sufficient dosages to retain antidepressant efficacy.
The development of the first generation MST device was supported by an NIH grant R01
MH60884, which funded the purchase of the first MST device from the company, according
to our requested specifications for device output. The development of the second generation
MST device was supported by the competitive renewal of this same NIH grant, which
funded the upgrade to the higher output device, according to our specifications.
VI. Translational Development Strategy for MST
NIH-PA Author Manuscript
The translational development strategy for MST started with demonstrating (a) feasibility in
the nonhuman primate and human, (b) testing safety in the nonhuman primate and human,
and (c) evaluating efficacy in the human. The developmental process has been iterative,
toggling back and forth between the nonhuman primate and human studies at each stage, as
reviewed below (and as illustrated in Figure 4).
VI. A. Feasibility of MST
VI. A. 1. Feasibility of MST in the Nonhuman Primate Model—The first phase in
the development of MST was to demonstrate feasibility in an animal model, and then show
feasibility in initial proof-of-concept human trials. We selected the rhesus monkey as the
animal model for MST feasibility testing after failing to produce seizures in rodents.
Monkeys presented a brain-to-coil size ratio that was closer to humans, and the use of
pediatric sized coils as opposed to the much smaller rodent coils reduced problems
stemming from excessive coil heating. They also have the advantage of being able to
perform cognitive tasks with which to access the relative cognitive effects of MST and ECT.
In 1998, we performed the first MST-induced seizures in anesthetized rhesus monkeys by
applying a custom modified Magstim device, which was capable of an output of 40 Hz at
Exp Neurol. Author manuscript; available in PMC 2010 December 4.
Rowny et al.
Page 6
NIH-PA Author Manuscript
100% stimulus intensity (Lisanby, et al., 2001). We found the procedure to be reproducible
and procedurally quite similar to conventional electroconvulsive shock (ECS) with the
exception that post-ictal recovery was exceptionally quick, and post-ictal changes in heart
rate and blood pressure did not seem as marked.
NIH-PA Author Manuscript
VI. A. 2. Feasibility of MST in Depressed Patients—We went on to perform the first
feasibility testing of MST in a depressed patient in 2000, in Bern, Switzerland (Lisanby, et
al., 2001). The patient was a 20-year-old woman with medication-resistant depression who
was referred for inpatient convulsive therapy. She received a course of four MST sessions,
followed by ECT. The MST sessions were conducted in the ECT suite using general
anesthesia, and occurred at 3 times per week for 4 sessions total. With the exception of
earplugs, modified electroencephalographic electrodes, and the use of an MST coil, the
procedure was identical to ECT. MST was administered with a pre-cooled coil positioned at
the vertex, or a figure-8 coil held to the right prefrontal cortex. The treatments were
delivered at 40 Hz and 100% of the maximum stimulator output of the device available at
that time. Each MST session successfully induced a tonic-clonic seizure, as was indicated by
motor contraction and EEG. Although the seizures resembled those of ECT, there was a
remarkable difference favoring MST in post-ictal recovery time, and in the lack of
significant side effects. The patient tolerated the treatments well and only reported a mild
headache after one treatment. There were no subjective complaints of impaired memory, and
the Mini Mental Status Exam score remained at 30/30 throughout the course of MST
treatment, although further neuropsychological testing was not performed. Of note the
Hamilton Rating Scale for Depression decreased from a baseline of 20 to 13 after the fourth
MST treatment.
The same program in Bern, Switzerland later reported the successful treatment of a second
patient suffering from medication-resistant major depression (Kosel, et al., 2003). The
treatment parameters were similar to those in the first case, and the patient received twelve
MST sessions in an inpatient setting. Tonic-clonic seizures were induced for a total of
eleven out of the twelve trials. The time taken to regain orientation after treatment was lower
than the average reorientation time for all conventional ECT methods. MST also showed a
lower cognitive side effect profile when compared to ECT. The patient tolerated the MST
well and had no complaints of any cognitive impairment, headache, or muscle pain.
VI. B. Safety of MST
NIH-PA Author Manuscript
VI. B. 1. Safety of MST in the Nonhuman Primate Model—After demonstrating
feasibility, the next step was to systematically examine the safety of the procedure in the
monkey model and in the human. In the monkey, we characterized the (1)
neurophysiological, (2) cognitive, and (3) anatomical effects of MST.
Neurophysiological studies of MST have shown the electric field and resultant seizure to be
more focal than with ECS. By implanting multi-contact intracerebral electrodes, we
recorded the electric fields induced in the brain with MST and ECS, and also characterized
the resultant seizures. We reported that the electric fields induced by MST within the cortex
are less intense and more confined to the superficial cortex than those induced by ECS
(Lisanby, et al., 2003, Peterchev, et al., 2007). Specifically, we found MST induced seizures
to be less robust, less generalized to deeper brain structures (i.e. hippocampus), and to have
less post-ictal suppression. Seizures induced by MST were also found to have a reduced
prolactin surge (Morales, et al., 2003) and to have a less immediate post-treatment
bradycardia, consistent with less seizure spread to the diencephalons and insula,
respectively.
Exp Neurol. Author manuscript; available in PMC 2010 December 4.
Rowny et al.
Page 7
NIH-PA Author Manuscript
To characterize the neurophysiological effects of chronic exposure to MST, we performed
quantitative analyses of ictal expression and post-ictal suppression following a 4-week
course of daily ECS, MST, or anesthesia-alone sham (Cycowicz, et al., 2008). Both ECS
and MST were given bilaterally at 2.5 × seizure threshold. These scalp EEG results
supported our earlier findings from intracerebral electrode recordings demonstrating that
MST- and ECS-induced seizures elicit differential patterns of EEG activation. ECS showed
significantly more marked ictal expression, and more intense post-ictal suppression than
MST in the theta, alpha, and beta frequency bands. However, ECS and MST were
indistinguishable in the delta frequency, demonstrating that MST differs in some respects
but shares some physiological effects seen with ECS.
NIH-PA Author Manuscript
To examine the cognitive effects of chronic exposure to MST, it was first necessary to
develop and validate a primate model for the amnestic effects of ECT in the monkey. To this
end, we developed and tested the Columbia University Primate Cognitive Profile (CUPCP)
(Moscrip, et al., 2004). The CUPCP is a three-task cognitive battery presented to rhesus
monkeys via a touch sensitive computer monitor. Task 1 measures post-intervention
orientation time and long-term memory. Task 2 measures working memory, immediate
learning, and retention over a delay. Task 3 measures serial memory and involves 3-item
lists to assess learning and memory skills. Anterograde amnesia was indicated by poor
retention of a 3-item list learned on the same day after intervention. Retrograde amnesia was
indicated by poor performance on lists learned previously to the intervention. Using this
animal model, we found that chronic exposure to daily MST results in more benign
cognitive side effects than ECS (Moscrip, et al., 2006). Monkeys were slower to complete
tasks following ECS versus MST, and were overall less accurate in long and short-term
memory and recall tasks post ECS when compared to sham and MST.
Next, we studied the anatomical effects of chronic exposure to MST using
neuropathological, stereological, and immunohistochemical methods. We reported that
neither MST nor ECS induced signs of neuropathological lesion (Dwork, et al., 2004).
Furthermore, stereological cells counts of prefrontal cortex (3 regions) and the dentate gyrus
(2 regions) showed no reductions in neuron or glial cell counts following treatment (Dwork,
et al., Submitted, Lisanby, et al., 2005). Neuroanatomical studies have also demonstrated
differences in the effects of MST and ECS on neuroplasticity. ECS has been shown to
increase mossy fiber sprouting in the dentate gyrus of rhesus monkeys after chronic
exposure. Such an effect was not found after chronic exposure to MST (Scalia, et al., 2004)
using the first generation MST device.
NIH-PA Author Manuscript
VI. B. 2. Safety of MST in Depressed Patients—The safety of MST in major
depression was investigated in a randomized, double-blinded, within-subject comparison of
MST to ECT (Lisanby, et al., 2003). Ten inpatients with major depression referred for
convulsive therapy were administered MST in two of the first four sessions (random order),
and were administered ECT for the remaining sessions. ECT and MST seizure thresholds
were titrated in the first two sessions. In the last two sessions, the participants were given
suprathreshold doses of ECT and MST, on separate days. Before and after every treatment, a
blind rater administered a neuropsychological battery and side-effect rating scale. In all ten
participants, seizures were successfully induced with both threshold and suprathreshold level
MST. The tonic-clonic seizures resembled those of ECT, except they had a shorter EEG and
motor duration than ECT-induced seizures. The ictal power of the MST seizure was also less
than that of the ECT sessions. The seizures induced by MST also had a lower ictal EEG
amplitude and less post-ictal suppression. After the MST sessions, patients had significantly
shorter times to re-orientation, and demonstrated better results on measures of attention,
retrograde amnesia, and category fluency. They also reported fewer side effects compared to
ECT.
Exp Neurol. Author manuscript; available in PMC 2010 December 4.
Rowny et al.
Page 8
VI. C. Efficacy of MST
NIH-PA Author Manuscript
NIH-PA Author Manuscript
Lacking a validated model for antidepressant efficacy in the rhesus monkey that had shown
responsiveness to ECS, we examined efficacy in depressed patients. The two openly treated
patients in the proof-of-concept feasibility trials both showed clinically significant
antidepressant efficacy, but these were not blinded studies (Kosel, et al., 2003, Lisanby, et
al., 2001). Subsequently, we conducted a 2-center trial, the aims of which were to provide
the first examination of the antidepressant effects of a full course of MST, and to provide the
first data on optimization of MST parameters (Lisanby, et al., 2003, White, et al., 2006).
This randomized, double-masked controlled trial contrasted two forms of MST that differed
in coil type (focal vs. nonfocal) and scalp location (prefrontal cortex vs. vertex). Twenty
adults with a major depressive episode were withdrawn from psychotropic medications and
randomly assigned to receive MST with a focal double cone coil (DCC) on the midline
prefrontal cortex or a nonfocal cap coil on the vertex (coils shown in Figure 3). Seizure
threshold was titrated at the first and penultimate treatments, and the remaining treatments
were given at the maximal device output (50 Hz, 100% intensity, 8 s duration), three
sessions per week. Masked ratings of clinical response were collected at baseline, during the
acute treatment phase, and at 2 and 6 month follow-up. Treatment number was titrated to
achieve maximal improvement. Masked assessments of neuropsychological functioning
were performed at baseline, at each treatment, post treatment, and at 2 months. 20 patients
completed the protocol, receiving an average of 9 MST treatments. The 2 coils differed in
motor threshold (cap < DCC). While all patients had seizures, the efficiency of seizure
induction differed by scalp location (100% at vertex, 12.5% at prefrontal cortex). When the
DCC failed to induce a seizure at prefrontal cortex, the coil was moved to the vertex and
successfully induced seizures. Seizure threshold increased after the course. Depression
scores significantly improved and 53% had at least a 50% improvement. We also compared
the dosage requirements for anesthesia, recovery times and electroencephalographic
bispectral index (BIS) with simultaneously treated patients receiving ECT. MST was again
associated with a decreased orientation time compared to ECT. Interestingly, smaller
succinylcholine doses were required during MST sessions. The motor and EEG seizure
activity did not differ in this trial, but MST was associated with a reduced variability in the
BIS (Lisanby, et al., 2003, White, et al., 2006).
NIH-PA Author Manuscript
These feasibility data support the safety of administering a course of MST in major
depression and show encouraging evidence for efficacy in an open-label setting. However, it
is also likely that the therapeutic effects were limited due to under dosing (patients were
treated an average of 1.3 × seizure threshold) and suboptimal coil placement. With the
substantial increase in seizure threshold across the treatment course, no patient could be
treated at the target of 6 times seizure threshold that increases the efficacy of unilateral ECT.
Device and coil modifications to rectify these output limitations were therefore undertaken,
as reviewed below.
VII. Translational Development Strategy for High Dose MST (HD-MST)
VII. A. Rationale for High Dose MST
The feasibility, safety, and efficacy testing reviewed above were conducted with the first
generation MST device capable of delivering a total of only 400 pulses at peak intensity and
frequency (50 Hz, 100% intensity, 8 second trains). The first efficacy study of this device
reviewed above demonstrated that higher output would be needed to optimize efficacy and
reliability of seizure induction throughout a complete course of treatment. Studies show that
antidepressant response to ECT is dosage sensitive. For example, 6× seizure threshold RUL
ECT can match the efficacy of BL ECT, but with cognitive savings (Sackeim, et al., 2000).
High dose RUL is achieved by increasing the number of pulses in the pulse train (either
Exp Neurol. Author manuscript; available in PMC 2010 December 4.
Rowny et al.
Page 9
NIH-PA Author Manuscript
through increased frequency, or train duration, or both). This suggests that cognitive side
effects might be more related to electric field amplitude than the number of pulses in the
train or the resultant seizure. We sought to test whether the same could be true for HD-MST
by determining whether we could increase efficacy without a commensurate increase in side
effects by increasing the number of pulses in the MST train. We therefore requested the
construction of a device capable of sustaining 100 Hz stimulation at 100% maximal
stimulator output for 10 second trains, and then subjected this second generation device to
the same stages of translational development, putting it back through the stages of
feasibility, safety and efficacy testing.
VII. B. Feasibility of HD-MST
NIH-PA Author Manuscript
We performed the first HD-MST session in a rhesus monkey in 2004, and have gone on to
systematically test the neurophysiological, cognitive, and anatomical effects of HD-MST in
the nonhuman primate model. HD-MST has been procedurally identical to the first
generation MST device, with the exception that the stimulation train is longer. The first
patient with depression to receive HD-MST with this device was treated in 2006 in Cardiff,
Wales, and the second HD-MST patient was treated in Edinburgh, Scotland. We reported
the results of the first 11 cases receiving HD-MST at these 2 centers. The patients reported
less confusion following MST treatment versus ECT, and demonstrated a faster orientation
time post MST (Kirov, et al., 2008). The treatments were well tolerated without significant
adverse events. Scalp EEG recording from one of these initial patients is shown in Figure 5.
VII. C. Safety of HD-MST
Since the amplitude of the pulses in HD-MST is identical to those used in the first
generation MST device, it was not necessary to repeat the recordings of electric field
distribution, as they would be identical on a per-pulse basis.
The first neurophysiological characterization of seizures induced by HD-MST was presented
by Cycowicz et al. (Cycowicz, et al., In press). In this study, four rhesus monkeys received 4
weeks of daily treatment with ECS (2.5× seizure threshold), HD-MST (6× seizure
threshold), and sham, in a counter-balanced order, and EEGs were recorded during and
following the seizure. We found ECS showed significantly more ictal power in all frequency
bands than HD-MST.
NIH-PA Author Manuscript
Cognition was also less affected by HD-MST than ECS. Using an expanded version of the
CUPCP, we reported that HD-MST retains cognitive advantages relative to ECS (Spellman,
et al., 2008). In this within-subject study, the rhesus monkeys were trained on five cognitive
tasks evaluating automatic memory, anterograde learning and memory, spatial and serial
working memory, and anterograde and retrograde simultaneous chaining. They were
administered daily HD-MST (6× seizure threshold), ECS (2.5 × seizure threshold), and
sham for 4 weeks and were tested on the cognitive battery twice daily following treatment.
We found HD-MST had a less severe cognitive side-effect profile when compared to ECS.
Specifically, the time to complete tasks after HD-MST was not different from sham
treatment. Furthermore, subjects took longer to complete tasks following ECS when
compared to HD-MST and sham. Treatment effects were also found in four out of six
measures of accuracy, in which ECS was worse than sham, but HD-MST was not
(Spellman, et al., 2008). It is interesting to note that the degree of cognitive impairment was
correlated with higher ictal expression in the scalp EEG (Cycowicz, et al., 2008).
VII. D. Efficacy of HD-MST
Clinical trials are now being conducted to compare the efficacy of HD-MST with ECT for
the treatment of depression. A two-center randomized trial is currently underway at
Exp Neurol. Author manuscript; available in PMC 2010 December 4.
Rowny et al.
Page 10
NIH-PA Author Manuscript
Columbia University Medical Center / New York State Psychiatric Institute and at the
University of Texas Southwestern Medical Center, contrasting HD-MST with ultrabrief
pulse right unilateral ECT at 6× seizure threshold. There are similar trials underway at
Cardiff University in Wales and the Royal Edinburgh Hospital in Scotland. The goal of
these trials is to investigate alternate stimulation sites for seizure induction, alternate coil
configurations, and to compare MST with conventional ECT. Other work is planned or
underway in Germany, Australia, and the US on a different MST device capable of even
higher stimulation frequencies.
VIII. Mechanism of Actions
NIH-PA Author Manuscript
To date there are no studies using functional imaging to examine the mechanisms of action
of MST. Since both MST and ECT are convulsive technologies, it is assumed there may be
partial overlap in their mechanisms. ECT exerts a range of neurobiological effects on
functional brain activation, neurotransmitter systems including those of seratonin,
hypothalamic–pituitary–adrenal axis (HPA-axis) normalizing the dexamethasone
suppression test, and neuroplasticity. ECT is also a powerful anticonvulsant (Rowny and
Lisanby, 2008). The enhanced focality of MST leads to the expectation that MST will exert
less of an effect on deeper brain structures. The ability to induce a seizure with MST without
exposing deeper brain regions to the direct effects of the induced electric field provides an
opportunity to study the relative contribution of the electricity and the seizure in producing
the clinical effects seen with ECT. However, the relations between differential impact of
seizures on functional brain activation and clinical outcome are not known. Such
information will be necessary to inform the development of versions of convulsive therapy
with an improved risk/benefit ratio.
IX. Conclusions and Future Directions
The strategy that has been applied to the development of MST illustrates the power of the
translational approach to rational device design. The process started with engineering
development, followed by animal testing, and finally clinical testing. However, these stages
were iterative, with new device designs being re-evaluated in the animal model prior to
human testing.
NIH-PA Author Manuscript
There is much left to be learned about MST, as it is still in the early phases of clinical
testing. Ongoing preclinical and clinical trials are currently seeking to answer questions
about dosimetry, optimal stimulation site, mechanism of action, and patient selection. The
results from these and future studies will inform how MST is delivered at will influence its
ultimate clinical role. Future generations of MST technology should be fine-tuned in regards
to optimal dosing strategies, coil type and coil position. Developing new coil designs will
likely allow enhanced focality of seizure induction and should minimize the spread of
seizure, as these are the overall goals of MST. Additionally, the optimal dosage of MST has
yet to be established. Although MST is currently performed 3 times per week, as is ECT,
there may be a more efficacious frequency of treatment. Furthermore, more work must be
done to understand how patients should be selected for treatment. Currently, MST is tested
in depressed patient who are referred for convulsive therapy. The patient subgroups that will
benefit the most from MST have yet to be identified.
Convulsive therapy is still the most effective and rapidly acting antidepressant available, and
is useful for the treatment and management of other conditions where less invasive
pharmacological treatments have failed. Work to date suggests that MST has a more benign
cognitive side effect profile than ECT, and current studies are comparing its efficacy as an
antidepressant treatment to ECT. MST also provides a new tool to study the mechanism of
Exp Neurol. Author manuscript; available in PMC 2010 December 4.
Rowny et al.
Page 11
NIH-PA Author Manuscript
action of ECT. Because it can deliver focal stimulation to specific regions of the brain, it can
be used to study the relationship between stimulation site and clinical outcomes. Ultimately,
understanding what makes ECT so much more effective and rapidly acting than other
available therapies may inform the development of more targeted and safer treatments in the
future.
Acknowledgments
The authors appreciate the help of Ms. Yael Cycowicz who assisted with the revision of this paper.
Disclosures: Support for the development of Magnetic Seizure Therapy has come from NIMH R01 MH60884, the
Stanley Medical Research Foundation, American Federation for Aging Research / Beeson Scholars Program, the
National Alliance for Research on Schizophrenia and Depression (NARSAD), and equipment support from
Magstim Company and MagVenture A/S. The authors hold no patents or stocks related to MST.
For work unrelated to MST, Dr. Lisanby has received research support from Neuronetics, Cyberonics, ANS,
DARPA, and NYSTAR. Columbia University has filed a patent application for novel TMS technology developed
in Dr. Lisanby's laboratory, not related to the topic presented here. Dr. Lisanby Chairs a Data Safety and
Monitoring Board for a trial of a cortical stimulator sponsored by Northstar Neuroscience.
References
NIH-PA Author Manuscript
NIH-PA Author Manuscript
Cycowicz YM, Luber B, Spellman T, Lisanby SH. Differential neurophysiological effects of magnetic
seizure therapy (MST) and electroconvulsive shock (ECS) in non-human primates. Clinical EEG
and Neuroscience 2008;39
Cycowicz YM, Luber B, Spellman T, Lisanby SH. Neurophysiological Characterization of High Dose
- Magnetic Seizure Therapy: Comparisons with Electroconvulsive Shock (ECS) and Cognitive
Outcomes. Journal of ECT. In press.
Deng ZD, Peterchev AV, Lisanby SH. Coil design considerations for deep-brain transcranial magnetic
stimulation (dTMS). Conf Proc IEEE Eng Med Biol Soc 2008;2008:5675–5679. [PubMed:
19164005]
Dwork AJ, Arango V, Underwood M, Ilievski B, Rosoklija G, Sackeim HA, Lisanby SH. Absence of
histological lesions in primate models of ECT and magnetic seizure therapy. Am J Psychiatry
2004;161:576–578. [PubMed: 14992989]
Dwork AJ, Romme-Christensen J, Bonde-Larsen K, Scalia J, Underwood MD, Arango V, Pakkenberg
B, Lisanby SH. Unaltered Neuronal and Glial Counts in Frontal Cortex and Hippocampus
Following Magnetic Seizure Therapy or Electroconvulsive Therapy in a Rhesus Monkey Model.
Submitted.
Epstein CM, Davey KR. Iron-core coils for transcranial magnetic stimulation. J Clin Neurophysiol
2002;19:376–381. [PubMed: 12436092]
Jalinous, R. Principles of magnetic stimulator design. In: Pascual-Leone, A.; D, N.; Rothwell, J.;
Wassermann, EM.; Puri, BK., editors. Handbook of Transcranial Magnetic Stimulation. Arnold;
London: 2002. p. 30-38.
Kirov G, Klaus EP, Scott AF, Atkins M, Khalid N, Carrick L, Stanfield A, O'Carrol RE, Mustafa HM,
Lisanby SH. Quick recovery of orientation after magnetic seizure therapy for major depressive
disorder. The British Journal of Psychiatry. 2008
Kosel M, Frick C, Lisanby SH, Fisch HU, Schlaepfer TE. Magnetic seizure therapy improves mood in
refractory major depression. Neuropsychopharmacology 2003;28:2045–2048. [PubMed: 12942146]
Lisanby SH. Update on magnetic seizure therapy: a novel form of convulsive therapy. J ECT
2002;18:182–188. [PubMed: 12468992]
Lisanby SH. Electroconvulsive therapy for depression. N Engl J Med 2007;357:1939–1945. [PubMed:
17989386]
Lisanby SH, Husain MM, Morales OG, Thornton WL, White PF, Payne N, Rush AJ, Sackeim HA.
Controlled clinical trial of the antidepressant efficacy of magnetic seizure therapy in the treatment
of major depression. 2003;166
Exp Neurol. Author manuscript; available in PMC 2010 December 4.
Rowny et al.
Page 12
NIH-PA Author Manuscript
NIH-PA Author Manuscript
NIH-PA Author Manuscript
Lisanby SH, Luber B, Finck AD, Schroeder C, Sackeim HA. Deliberate seizure induction with
repetitive transcranial magnetic stimulation in nonhuman primates. Arch Gen Psychiatry
2001;58:199–200. [PubMed: 11177122]
Lisanby SH, Luber B, Schlaepfer TE, Sackeim HA. Safety and feasibility of magnetic seizure therapy
(MST) in major depression: randomized within-subject comparison with electroconvulsive
therapy. Neuropsychopharmacology 2003;28:1852–1865. [PubMed: 12865903]
Lisanby SH, Maddox JH, Prudic J, Devanand DP, Sackeim HA. The effects of electroconvulsive
therapy on memory of autobiographical and public events. Arch Gen Psychiatry 2000;57:581–590.
[PubMed: 10839336]
Lisanby SH, Moscrip T, Morales O, Luber B, Schroeder C, Sackeim HA. Neurophysiological
characterization of magnetic seizure therapy (MST) in nonhuman primates. Suppl Clin
Neurophysiol 2003;56:81–99. [PubMed: 14677385]
Lisanby SH, Pakkenberg B, Dwork AJ, Christensen JR, Larsen KB, Scalia J, Underwood M, Sackeim
HA, Arango V. Stereological and neuropathological examination of the safety of electroconvulsive
shock (ECS) and magnetic seizure therapy (MST) in nonhuman primates. Biological Psychiatry
2005;57:147S.
Lisanby SH, Schlaepfer TE, Fisch HU, Sackeim HA. Magnetic seizure therapy of major depression.
Arch Gen Psychiatry 2001;58:303–305. [PubMed: 11231838]
Luber B, Nobler MS, Moeller JR, Katzman GP, Prudic J, Devanand DP, Dichter GS, Sackeim HA.
Quantitative EEG during seizures induced by electroconvulsive therapy: relations to treatment
modality and clinical features. II. Topographic analyses. J ECT 2000;16:229–243. [PubMed:
11005044]
Morales O, Luber B, Kwon E, Ellsasser R, Sackeim HA, Lisanby SH. Prolactin response to convulsive
therapy: magnetic seizure therapy (MST) versus electroconvulsive shock (ECS) in nonhuman
primates. J ECT 2003:19–58A.
Moscrip TD, Terrace HS, Sackeim HA, Lisanby SH. A primate model of anterograde and retrograde
amnesia produced by convulsive treatment. J ECT 2004;20:26–36. [PubMed: 15087994]
Moscrip TD, Terrace HS, Sackeim HA, Lisanby SH. Randomized controlled trial of the cognitive sideeffects of magnetic seizure therapy (MST) and electroconvulsive shock (ECS). Int J
Neuropsychopharmacol 2006;9:1–11. [PubMed: 16045810]
Peterchev A, Kirov G, Ebmeier K, Scott A, Husain M, Lisanby SH. Frontiers in TMS technology
development: controllable pulse shape TMS (cTMS) and magnetic seizure therapy (MST) at 100
Hz. Biological Psychiatry 2007;61:107S.
Peterchev AV, Jalinous R, Lisanby SH. A transcranial magnetic stimulator inducing near-rectangular
pulses with controllable pulse width (cTMS). IEEE Trans Biomed Eng 2008;55:257–266.
[PubMed: 18232369]
Rowny, S.; Lisanby, SH. Brain Stimulation in Psychiatry. In: Tasman, A.; Maj, M.; First, MB.; Kay,
J.; JA, L., editors. Psychiatry. 3rd. Wiley; 2008.
Sackeim HA, Luber B, Moeller JR, Prudic J, Devanand DP, Nobler MS. Electrophysiological
correlates of the adverse cognitive effects of electroconvulsive therapy. J ECT 2000;16:110–120.
[PubMed: 10868321]
Sackeim HA, Prudic J, Devanand DP, Nobler MS, Lisanby SH, Peyser S, Fitzsimons L, Moody BJ,
Clark J. A prospective, randomized, double-blind comparison of bilateral and right unilateral
electroconvulsive therapy at different stimulus intensities. Arch Gen Psychiatry 2000;57:425–434.
[PubMed: 10807482]
Sackeim HA, Prudic J, Fuller R, Keilp J, Lavori PW, Olfson M. The cognitive effects of
electroconvulsive therapy in community settings. Neuropsychopharmacology 2007;32:244–254.
[PubMed: 16936712]
Scalia J, Lisanby SH, Underwood M, Sackeim H, Dwork AJ, Morales O, Fung E, Arango V. The
spatial distribution of mossy fiber sprouting in non-human primate model for electroconvulsive
therapy and magnetic seizure therapy. Biological Psychiatry 2004;55:207S.
Sommer M, Alfaro A, Rummel M, Speck S, Lang N, Tings T, Paulus W. Half sine, monophasic and
biphasic transcranial magnetic stimulation of the human motor cortex. Clin Neurophysiol
2006;117:838–844. [PubMed: 16495145]
Exp Neurol. Author manuscript; available in PMC 2010 December 4.
Rowny et al.
Page 13
NIH-PA Author Manuscript
Spellman T, McClintock SM, Terrace H, Luber B, Husain MM, Lisanby SH. Differential effects of
high-dose magnetic seizure therapy and electroconvulsive shock on cognitive function. Biol
Psychiatry 2008;63:1163–1170. [PubMed: 18262171]
Spellman T, Peterchev AV, Lisanby SH. Focal Electrically Administered Seizure Therapy: A Novel
form of ECT Illustrates the Roles of Current Directionality, Polarity, and Electrode Configuration
in Seizure Induction. Neuropsychopharmacology. 2009
Tings T, Lang N, Tergau F, Paulus W, Sommer M. Orientation-specific fast rTMS maximizes
corticospinal inhibition and facilitation. Exp Brain Res 2005;164:323–333. [PubMed: 15868175]
White PF, Amos Q, Zhang Y, Stool L, Husain MM, Thornton L, Downing M, McClintock S, Lisanby
SH. Anesthetic considerations for magnetic seizure therapy: a novel therapy for severe depression.
Anesthesia & Analgesia 2006;103:76–80. [PubMed: 16790630]
Zangen A, Roth Y, Voller B, Hallett M. Transcranial magnetic stimulation of deep brain regions:
evidence for efficacy of the H-coil. Clin Neurophysiol 2005;116:775–779. [PubMed: 15792886]
NIH-PA Author Manuscript
NIH-PA Author Manuscript
Exp Neurol. Author manuscript; available in PMC 2010 December 4.
Rowny et al.
Page 14
NIH-PA Author Manuscript
NIH-PA Author Manuscript
Figure 1. Stages of physiologically-informed intervention development as applied to Magnetic
Seizure Therapy (MST)
NIH-PA Author Manuscript
The processes starts with a clinically salient physiological target, which in this case was a
pattern of brain activation associated with therapeutic response to ECT, with frontal
inhibitory effects as illustrated by frontal delta-band EEG activity (Luber, et al., 2000). A
prototype device was designed to achieve this physiological target, which had maximal
stimulator output of 50 Hz for 8 seconds, at maximal stimulator output (100% intensity).
The ability of this prototype device to achieve the physiological target was tested in
nonhuman primates, using intracerebral electrodes to record the strength of the induced
electric field of MST, contrasting it with electroconvulsive shock, and to characterize the
topography of the induced seizures. The safety and efficacy of this prototype was piloted in
depressed patients, yielding supporting evidence of safety relative to ECT and encouraging
signs of efficacy in open trials, however there were suggestions that output restrictions may
be limiting efficacy. This led to a refined design for a high dose MST device, capable of 100
Hz for 10 seconds at 100% intensity.
Exp Neurol. Author manuscript; available in PMC 2010 December 4.
Rowny et al.
Page 15
NIH-PA Author Manuscript
NIH-PA Author Manuscript
Figure 2. Evolution of MST Devices for Non-Human Primate and Clinical Trials
NIH-PA Author Manuscript
A) First generation nonhuman primate MST device, used in the first MST feasibility trials in
rhesus monkeys (Lisanby, et al., 2001) and the first depressed patient (Lisanby, et al., 2001).
A conventional rTMS device was expanded by adding 4 charging units, yielding a total of 8
booster modules and a maximal stimulator output of 40 Hz.
B) First generation human MST device used in the first randomized trial of MST in the
treatment of depression conducted at Columbia and University of Texas Southwestern
Medical Center (Lisanby, et al., 2003, White, et al., 2006). Sixteen charging units were
employed, resulting in a maximal stimulator output of 50 Hz, 8 seconds, 100% intensity.
C) Second generation nonhuman primate high dose MST (HD-MST) device used in the first
animal studies of HD-MST (Spellman, et al., 2008). A three-phase power supply and other
revisions to circuit topology were used to achieve 100 Hz, 10 second trains at 100%
intensity.
D) Second generation human HD-MST device used in the first human studies of HD-MST
(Kirov, et al., 2008). The Magstim Theta is capable of 1000 pulses per train at 100 Hz and
100% intensity, as well at theta burst stimulation. Three coils can be connected
simultaneously to facilitate seizure threshold titration. A touch screen monitor provides an
interactive user interface. The self-contained console has a small footprint to fit easily in a
typical clinical ECT suite.
Exp Neurol. Author manuscript; available in PMC 2010 December 4.
Rowny et al.
Page 16
NIH-PA Author Manuscript
Figure 3. MST coil design
Figure 8 coil (left), double cone coil (middle), cap coil (right). The figure 8 coil was least
efficient in seizure induction. The Double cone coil conforms well to vertex and midline
prefrontal placements and was more efficient in seizure induction. The cap coil, which is a
round coil with concave windings to conform closely to the scalp, was highly efficient in
inducing seizure with a bilateral placement.
NIH-PA Author Manuscript
NIH-PA Author Manuscript
Exp Neurol. Author manuscript; available in PMC 2010 December 4.
Rowny et al.
Page 17
NIH-PA Author Manuscript
NIH-PA Author Manuscript
Figure 4. Translational development strategy applied to MST
The development of MST progressed through the stages of feasibility testing, safety
evaluation, and efficacy trials. At each stage, work toggled back and forth between animal
testing in nonhuman primates and clinical trials. Device design was iteratively improved,
incorporating results from each stage of development. Animal testing provided the unique
opportunity to test the safety of MST using postmortem anatomical studies, while clinical
trials provide the unique opportunity to assess efficacy in depressed patients.
NIH-PA Author Manuscript
Exp Neurol. Author manuscript; available in PMC 2010 December 4.
Rowny et al.
Page 18
NIH-PA Author Manuscript
NIH-PA Author Manuscript
Figure 5. Scalp EEG of a patient undergoing high dose MST for treatment of depression
This is a representative 2-channel, bifrontotemporal scalp EEG recording during one of the
initial human HD-MST sessions from the case series reported in Kirov et al. (Kirov, et al.,
2008). The period of the MST train is marked. Following the train, high frequency spiking
can be seen in both channels that evolves into spike-slow wave configuration, and then slow
wave activity. This was accompanied by a generalized tonic-clonic seizure as documented
by motor convulsion in an unanesthetized limb, employing the cuff-technique.