JO URNAL O F
C HRO MATO G RAPHY A
ELSEVIER
Journal
of Chromatography
zyxwvu
A, 688 (1994) 97-105
Analysis of colloids
VII. *Wide-bore
hydrodynamic chromatography,
a simple
method for the determination of particle size in the nanometer
size regime
Ch.-H. Fischer*, M. Giersig zyxwvutsrqponmlkjihgfedcbaZYXWV
Hahn-M eitner-Institut
Berlin. Department
CK, Glienicker Strasse 100, D- 14109
Berlin,
Germany
First zyxwvutsrqponmlkjihgfedcbaZYXWVUTSRQPONMLKJIHGFEDCBA
received 6 July 1994; revised manuscript
received 6 October
1994 zyxwvutsrqponmlkjihgfedcbaZY
Abstract
Wide-bore
hydrodynamic
chromatography
in a polyether ether ketone [PEEK) capillary (I.D. 0.7 mm, length 20
m) was used to determine
the weight average diameter d, of colloidal particles.
The method was applied to
cadmium sulphide and gold sols in the diameter range between 3 nm and 27 nm. The method is based on the radial
distribution
of the analyte in the capillary due to the hydrodynamic
flow profile in the capillary and due to the
diffusion coefficient of the particles, which is dependent
on their diameter. The diameter was calculated from the
ratio of the heights of convection
peak and diffusion peak. The size-quantization
effect of small semiconductor
particles made it possible to visualise the separation inside of the capillary. One important advantage of the applied
method is the very much reduced adsorption,
which often causes serious problems
in the HPLC especially
of
inorganic colloids. The results of wide-bore
hydrodynamic
chromatography,
size exclusion chromatography
and
transmission
electron microscopy were compared.
Introduction
The study of colloidal
semiconductor
and
metal particles in the nanometer
size regime is a
steadily growing field in chemistry.
Because ultra
small
particles
exhibit
unusual
physical
and
chemical properties,
e.g. blue-shifted
absorption
and fluorescence
spectra with decreasing
diameter and because of their possible application
in
solar energy technology
and microelectronic
devices they have become the focus of much recent
* Corresponding
author.
c For part VI see Ref. {Xl. for part V Ref. [7/ and for part IV
Ref. 1261.
0021-9h73/94/$07.00
SSl?I
0021-9673(94)0096z-7
(Q
I994 Elsevier
Science
B.V. All rights
physicochemical
research
[ 1,2]. However
the
investigation
of size dependent
properties
requires good and reliable size analysis. The classical technique is transmission
electron microscopy
(TEM),
where the particle diameters
are measured directly. Nevertheless,
problems
can arise
from radiation
damage due to the high energy
applied to the material [3]. If no expensive image
processing system is available,
the measurement
of the sizes on the micrographs
is also rather
tedious and often quite subjective.
Size exclusion
chromatography
(SEC) has also recently
been
shown to be a very convenient
method for the
size determination
for inorganic colloids [4,5]. In
the case of HPLC-SEC,
the system measures
reserved
C‘hromatogr
A 68X (lY94)
97- 10- 5
size distributions
within a few minutes. once the
raphy. it is the peak shape,
rather
than the
initial
calibration
has been carried
out [6,7].
retention
time, which was used for the calculaSpeed is important
when kinetics of the growth
tion of the molecular
weight. The method
is
based on the dependence
of the diffusion coeffiof very unstable
colloids
is investigated
[6].
Further
advantages
of the chromatographic
cient on the molecular weight. Conversely,
diffumethod
are the good statistics
of the result,
sion coefficients of species with known molecular
on-line coupling with diode array detectors zyxwvutsrqponmlkjihgfedcbaZYXWVUTSRQPONMLKJIHGFED
for
weight can be calculated
from the peak width
studies of size depending
WVIVis spectra [6] and
[20]. Vanderslice
et. al. investigated
the conthe possibility
of scaling-up
the system for precentration
profiles
in such flow systems
and
parative
separations
[8]. However,
with some
calculated
them under various conditions
[21].
kinds of colloids, e.g. PbS, adsorption
or particle
In the past the method in question
has been
growth during the passage through the column
contributed
to the group of flow injection
analycauses difficulties
even in the presence
of the
sis (FIA).
However,
most often in FIA constabilisers in the eluent. The hydrodynamic
chrocentrations
are measured
by peak size either
matography
in packed columns showed the same
after mixing with special reagents
or without
limitations
]9]. In the classical
hydrodynamic
mixing by the use of specific detectors.
Though
chromatography
(HDC) in narrow capillaries or
the described
method has doubtless
similarities
with other name capillary
hydrodynamic
fracwith FIA, it is based on the hydrodynamic
flow
tionation
(CHDF)
organic
particles
could be
profile in the capillary, which is generated
by the
separated
according
to the diameter
with good
flow resistance
of the walls. The species to be
resolution
[lo-131.
Silebi
[ 111 described
the
analysed
interact
dynamically
with the flow
analysis of latex particles with diameters down to
medium
and indirectly
with the walls. This is
88 nm (diameter
range of the inorganic particles
typical for chromatography,
though not a specunder investigation:
3 nm-27
nm). Adsorption
trum of various (particular)
retention
times is
was less pronounced
than in packed columns,
measured.
Therefore
we prefer the name widebut still a problem. Therefore
a chromatographic
bore hydrodynamic
chromatography
(HDC)
in
technique
could be helpful,
where there is a
order to stress similarities
and differences
to the
smaller surface area than in a packed column or
classical HDC.
in a narrow capillary.
In 1978 Mullins and Orr [14] and in 1979 Noel
et al. [15] reported
the fractionation
of miExperimental
crometer-sized
particles
in a capillary
with an
internal
diameter
of 250 pm. Submicrometer
Chromatography
particles could not be distinguished.
In 1984 Kelleher and Trumbore
[16] described
W ide- bore
HDC
an easy method
for the determination
of the
The experimental
set-up consisted of a Merckmolar weight of biopolymers
just by pumping a
Hitachi L6000 HPLC pump, a Merck Autosamsample plug through a rather thick capillary. The
pler A2000 (sample volume 100 ~1) and a Waters
internal
diameter
was some tenths of a mil990 diode array detector,
Autosampler
and delimetre
and therefore
much thicker
than in
tector were connected
via a 20 m long, 0.7 mm
classical
hydrodynamic
chromatography
(capilI.D. polyether
ether ketone (PEEK)
capillary.
lary hydrodynamic
fractionation)
where the diThe flow rate was 0.8 mllmin.
The eluent for the
ameter
is typically some microns.
The expericadmium
sulphide
sols was lop3 M cadmium
ment was carried out initially
with a normal
perchlorate
(Ventron)/
+10e3 M sodium polyUV/Vis detector, but later with special RI detecphosphate
(based on the phosphate units, Riedel
tors which measured
the radial concentration
de Haen),
and for gold sols lo-’ M sodium
gradient
[17-191.
Unlike
normal
chromatogcitrate was used.
Ch.-H.
Fischer.
M. Giersig
:
J. Chromntogr.
SEC
Two 125 x 4 mm columns (Knauer Stiulentechnik, Berlin, Germany)
in series were used: For
cadmium
sulphide Nucleosil 5OOC4 (7 pm) and
Nucleosil lOOOC4 (7 pm) and for gold Nucleosil
500 (15-25 pm) and Nucleosil
lOOOC4 (15-25
pm). Eluents, pump and detector were the same
as for HDC.
A 68X (1994)
size distributions
micrographs.
99
97- 105
were
measured
from
electron
Electron microscopy
.4 small drop of sample was adsorbed
onto a
40%mesh copper grid coated with a 50 A thick
carbon support film. After 10 seconds of contact
time the fluid was blotted off. The grids were
dried under argon and examined in a Philips CM
Preparation zyxwvutsrqponmlkjihgfedcbaZYXWVUTSRQPONMLKJIHGFEDCBA
ofthe colloids
12 transmission
electron
microscope
with an
acceleration
voltage of 120 kV. The microscope
Cadmium sulphide sols
was equipped
with a supertwin
lens and an
Hydrogen
sulphide
gas or aqueous
sodium
EDAX detector.
For imaging, axial illumination
hydrogen
sulphide solution was injected through
was used as well as the “nanoprobe
mode” with
a septum into an aqueous
solution of 10-j M
a beam spot size of 1.5 nm, to enable diffraction
Cd( ClO,),
and 6 1 10ml M sodium
polyphospatterns of the individual clusters to be obtained.
phate, through which nitrogen had been bubbled
All images were made under conditions
of minifor ten minutes.
The solution was shaken prior
mum phase contrast and low electron dose with a
to use. The particle size was controlled
by the
magnification
of 120 000 and 430 000 X .
initial pH value of the solution before sulphidc
addition.
A lower starting pH leads to smaller
particles
[22). Aged samples containing
larger
Results and discussion
colloids were also used in some experiments.
M ethod
Gold sols
Gold sols were prepared by using a mixture of
trisodium
citrate and tannic acid (Mallinckrodt
product
no. 8835)
as reducing
agent
[23].
KAuCI,
(85 ml, 0.1%) was heated to 60°C and
stirred rapidly. A second reducing solution was
prepared
by mixing trisodium citrate (4 ml, 1% )
tannic
acid (O-S ml, 1%) and an equivalent
amount of K,CO, (O-5 ml. 10 2 M and making
up to 25 ml.- This solution
was also heated to
60°C and then added rapidly to the chloroaurate
solution.
The colour of these sols developed
almost instantly.
The solution
was then boiled
for several minutes and allowed to cool. Tannic
acid increases
the rate of nucleation,
thereby
generating
smaller
particles.
The higher
the
tannic acidzcitrate
ratio, the smaller the particle
size. The lowest size limit achievable
was found
to be about 2-3 nm. The sols so prepared were
stable for months,
although
sometimes
a slow
sedimentation
was observed
over time, which
could be removed by centrifugation.
The particle
When a liquid passes through a capillary under
laminar flow conditions,
a parabolic
flow profile
is formed,
i.e, fast flow in the centre and decreasing
velocity
towards
the walls. Dissolved
species are- transported
forward with the liquid
flow, but they can also move in other directions
by diffusion.
Here the motion perpendicular
to
the flow direction
is of particular
importance.
It
brings the solute from the centre to the walls and
vice versa, i.e. from faster streams
to slower
flowing parts of the cross-section.
However,
this
motion is dependent
on the diffusion coefficient
of the sample.
When
a sample
of very big
colloidal particles or very large macromolecules
is injected into such a flow system with appropriate flow rate, their diffusion is low compared
to the speed of the forward stream. Therefore
the radial movement
is negligible.
In Fig. la the
axial and radial concentration
distribution
of
sample
species
in the
capillary
is shown
schematically.
based on the calculations
of Van-
loo
Ch.-H.
dif f usion
al
Fischer, M. Giersig
I J. Chromatogr.
A 688 (1994)
97-105 zyxwvutsrqponmlkjihgfedcbaZYXWVU
relatively
symmetric
late-eluting
diffusion
zyxwvutsrqponmlkjihgfedcbaZYXWVUTSRQPONMLKJIHGFEDCBA
roe f f icient
ioi
peak
diffusion
coefficient.
Materials
with a medium
diffusion
coefficient
show an elution profile with the elements of both
kinds of peaks, the ratio depending
on the size of
the diffusion coefficient.
Fig. lb shows the elution profiles corresponding
to the upper situations. The diffusion coefficient
of dissolved
organic polymers depends on the size of their coils
which is proportional
to the molecular weight. In
the case of colloidal
particles their diameter
is
the important
parameter.
The experimental
setup is very simple (Fig. 2). A pump delivers the
eluent, and the sample is introduced
by a sample
valve and pumped through the capillary to the
detector. zyxwvutsrqponmlkjihgfedcbaZYXWVUTSRQ
low
for material
with a high
medium
high
- zyxwvutsrqponmlkjihgfedcbaZYXWVUTSRQPONMLKJIHGFEDCBA
-a
2
0
0
b1 iGj
0
axial distance
I
La-u.1 -
.
;:%odof l
-a
0
-
time 1a.u.l
Fig. 1. Behaviour
of material with small, medium and high
diffusion
coefficient
during laminar flow through
a “widebore” capillary.
(a) Axial and radial concentration
distribution expressed
in equiconcentration
lines, based on the
theoretical
calculations
of Vanderslice
et al. 1211. (b) Corresponding
elution profiles.
derslice
et al. [21].
The left part of Fig. 1
represents
the situation
just described
for low
diffusion
coefficient.
The analyte
follows the
laminar
flow profile,
apparently
without
any
additional
motion.
On the other hand species
with a very high diffusion
coefficient
exchange
efficiently
between slow and fast areas (Fig. 1,
right part), and therefore
the concentration
is
more uniform over the diameter and the average
speed is slower than in the first case. Species
with medium diffusion coefficient (Fig. 1, centre)
show a behaviour
between
the extreme
cases.
These distributions
cannot be easily visualised.
But with a simple experimental
set-up consisting
of a pump, an injection
valve with sample loop,
the capillary and a detector with a through-flow
cell, an elution profile can be obtained which is
similar
in principle
to a chromatogram.
The
response
gives the integral radial concentration
at a certain
axial distance
from the injection
point. Due to the flow, the whole distribution
is
pushed through the detector cell with time. The
result is an early eluting, steeply increasing,
but
strongly
tailing and therefore
asymmetric
convection peak for slowly diffusing species and a
Application
Cadmium
sulphide sols
For the first experiments
cadmium
sulphide
sols were used. Much experience
exists in SEC
of these semiconductor
colloids,
so that the
results could easily be compared.
Stabilisers such
as polyphosphates
have to be added to these
aqueous
colloids
in order
to reduce
particle
growth. These molecules
form complexes
with
the surface of the particles and protect it against
direct contact with others and therefore
against
coagulation.
On the other hand it also reduces
adsorption
on the surface of the column and the
stationary
phase.
A series of CdS sols of different particle size,
lo-’ M each, were prepared,
whereby the size
was controlled
via the pH value before
the
sulphide
addition.
These
sols were
injected
simultaneously
in the HDC capillary
and onto
the SEC column.
The eluent compositions
of
both methods were the same: lop3 M cadmium
Fig. 2. Scheme
of experimental
set-up
in wide-bore
HDC.
Ch.-H.
Fischer,
M. Ciersig
1. Chromatogr.
A 688 (1994)
101
97- 105
shorter retention
volumes,
the HDC peak
became less symmetric. In addition to the late
diffusion peak, the early convection peak grew.
Finally only the latter remained with a pronounced tailing.
For proof of the separation inside a wide
capillary the size quantization effect (Q effect) of
the nm-sized semiconductor particles could be
used. For these particles, the onset of absorption
shifts to shorter wavelengths with decreasing
particle size [24,25]. When a CdS sol with a
broader size distribution was analysed, different
1 &,Yj$,
27.l)nm
1
chromatograms were obtained depending on the
observation wavelength. At shorter wavelengths
3
4
6
14
22 zyxwvutsrqponmlkjihgfedcbaZYXWVUTSRQPONMLKJIHGFEDCBA
(250 nm and below) all sizes are detected and
time / min
time / min
have the same molar extinction coefficient,
Fig. 3. Cadmium
sulphide
sols of different
particle
sizes
whereas at longer wavelengths smaller particles
analysed
by SEC (left) and wide-bore
HDC (right).
The
absorb less than larger ones. Consequently we
samples are sorted, the smallest particles on top, largest at
found
in Fig. 4a at 250 nm a large diffusion peak
the bottom.
The weight average diameters
d, are given on
next
to
a small shoulder due to a convection
the right hand side.
peak. With increasing wavelength, the shoulder
perchlorate/
* lop3 A4 sodium polyphosphate.
grew and at 500 nm the convection peak was
Fig. 3 shows the results. On the left, the size
pronounced,
because smalIer particles do not
exclusion chromatograms and on the right the
absorb any more in this wavelength range. The
corresponding
hydrodynamic
chromatograms.
separation is also evident from the spectra taken
The samples are sorted with respect to increasing
with the diode array spectrometer during the run
particle diameter. As the SEC peak shifted to
(Fig. 4b). The spectrum at 8.7 min had an onset
SEC
I
WIDE BORE HOC
5UOnm
\
--..-~
~
I
I
I
I
I
6
10
14
18
22
time
/min
I
h
250
.
300
.
.
400
wavelength
Fig. 4. Wide-bore
HDC with diode array detection
and the size quantization
effect of a cadmium
broad size distribution.
Left: Chromatograms
obtained
at different wavelengths.
Right: UV-Vis
(dotted line), at 11.9 min (solid line) and the difference A between both spectra (dashed line).
.
.
500
/ nm
sulphide sol with a relatively
spectra measured
at 8.7 min
102 zyxwvutsrqponmlkjihgfedcbaZYXWVUTSRQPONMLKJIHGFEDCBA
Ch.-H. Fischer. M. Giersig i 1. Chromarogr. A 688 (1994) 97- 105
of absorption
near 500 nm and no further fine
weight from the obtained
elution profiles. When
structure,
typical for rather large CdS particles.
a gradient detector was used, the molecular mass
The spectrum
obtained
at 11.9 min showed a
M, was determined
by means of the asymmetry
similar onset, but also a small maximum
at 330
ratio of the derivative
signal [27]. Trumbore
et
nm and a shoulder at 350 nm. This spectrum is a
al. [28] suggested for the normal elution profiles,”
superposition
of smaller
and bigger particles,
the ratio R of the height of the convection
peak
since the bigger particles of the sample are still
h, to that of the diffusion peak h, (Eq. 1) for the
eluting. As is evident from the chromatogram
at
determination
of the molar weight M, of a
500 nm they are only of medium
size with
polymer
by empirical
correlation.
When one of
convection
and diffusion
peaks of about the
the both peaks is not sufficiently
pronounced,
same height. Therefore
a spectrum with a prothe height of the chromatogram
at the position in
nounced
maximum
typical for particles below 3
question is taken for the calculation.
nm is obtained,
when the first spectrum
is
subtracted
from the second (spectrum A in Fig.
R = h,ih,
(1)
4b). The maxima in this region are due to socalled magic agglomeration
numbers,
i.e. energetically very stable agglomerates
[ 1,2]. zyxwvutsrqponmlkjihgfedcbaZYXWVUTSRQPONMLKJIHGFED
We also used the height ratio R, but as a
function of the weight average diameter d,, since
colloid chemists
prefer
the use of diameter,
Gold sols
Gold sols were prepared by the reduction of
considering
that the solid particles are rigid and
tetrachloroaurate
with citrate in the presence of
non-swelling
as opposed to the organic polymer
tannic acid [23]. Increasing
concentration
of the
coils. For the calibration,
a number
of colloids
latter yields smaller particles.
A series of these
were prepared and their weight average diameter
sols was also investigated
by transmission
elecwas determined
by computer
evaluation
of the
tron microscopy (TEM). In the micrographs,
the
SEC chromatograms
or in the case of gold sols
diameters
of a sufficiently
high number
of pardirectly by TEM. In Fig. 6 the ratio R as a
ticles, i.e. more than 1.50, was measured and the
function
of the diameter
d, is given for both
size distribution
constructed
(Fig. 5, left). Then
colloids CdS and Au and there is a clear clepenSEC was carried out (Fig. 5, centre) on Nucleosil
dence. From these calibration
plots the average
500 and Nucleosil
1000 (15-25 Frn) [26]. When
particle size of unknown
samples of the same
smaller silica material was utilised, the gold sol
material can easily be determined.
The slopes of
was irreversibly
adsorbed on the column. There
both curves are quite different_ Two reasons can
are some samples with bimodal
distribution in
be given to explain
this. Firstly,
the eluent
the SEC and only a single size population
in
composition
was different and secondly, the true
TEM. It shows that sometimes
in the TEM less
size of the solid particles is different
from the
frequent populations
can be overlooked.
For the
effective particle size due to the electrical double
wide-bore
HDC of gold SOIS. a 1 mM sodium
layer, which is formed at the solid-liquid
intercitrate solution was used as the eluent, the same
face by electrolytes
of the solution.
The thickas that used with SEC. Again, the same trend in
ness of this layer is dependent
on the solid
the chromatogram
shape from pure diffusion to
material
but also on the particular
electrolyte.
pure convection
peak was observed
(sample
Under
the conditions
of Fig. 6 the electrical
a--, e in Fig. 5, right). These parallel
experidouble layer of the gold particles
seems to be
ments
allow a good comparison
of all three
thicker and therefore
diffusion
plays a less immethods.
portant role than in the case of cadmium
sulphide.
Calculation
Finally it should be mentioned
that the concentration
of the colloid itself also has an effect
In the past different
approaches
have been
on the diffusion rate and therefore
on the result
used in the determination
of the molecular
of size determination.
The higher the concen-
Ch.-H.
Fischer,
M. Giersig
I .I.
Chromatogr.
A 688 (1994)
---
_-
.-
i
----
-_-
----
--_--
-mm-em-
j.
-P!lJed
%
103
97-105
1
4-
0
zyxwvutsrqponmlkjihgf
104
Ch.-H.
Fischer.
M. Giersig
i J Chromalogr.
A 688 (1994) 97-105
2
.
z
II
ET
0
1
I
I
I
I
2
3
4
5
6
[CdS] / mM
Fig. 7. Effect of concentration
on the result of wide-bore
HDC. A 5 mM cadmium sulphide sol was diluted stepwise
down to 0.1 mM. Height ratio R = h,ihz of convection
peak
to diffusion peak (0) and the particle diameter
dual, calculated from R as a function of CdS concentration
(0).
0
5
10
diameter
15
20
25
30
/ nm
Fig. 6. Calibration
plot for the wide-bore
HDC of gold (a)
and cadmium sulphide (0) soIs. It shows the ratio R = h, /h,
of the height of convection
peak over diffusion peak as a
function of weight average particle diameter d,. determined
by SEC or electron microscopy.
respectively.
Experimental
conditions
are given in the text.
Comparison zyxwvutsrqponmlkjihgfedcbaZYXWVUTSRQ
of the methods and conclusion
TEM shows directly
the size and shape of
particles, but radiation damage can occur [3] and
alter the original particle size. Furthermore,
less
frequent
populations
could be overlooked
[26]
and sometimes during sample preparation,
smaller and larger particles separate to some extent by
tration
of the colloidal
particles,
the faster is
diffusion,
so that the statistics
in an observed
their diffusion
between
slowly and fast flowing
part of the w-hole sample is not perfect any more.
parts of the cross-section
of the capillary and the
Without
a digital analyser
the analysis is time
more pronounced
is the diffusion peak. This is
consuming
and tedious. SEC gives a statistically
demonstrated
with a sol containing
5. zyxwvutsrqponmlkjihgfedcbaZYXWVUTSRQPONMLKJIHGFE
lop3 M
more accurate view of the true size distribution
CdS, which was diluted
with water stepwise
as long as all colloidal
material
elutes.
The
down to 1 - lOPa A4 (Fig. 7). From the particular
analysis is very fast. However,
the chromatogheight ratio R of the chromatogram
and the
raphy columns have a limited lifespan.
A new
calibration
curve (Fig. 6) diameters
between 89
column must then be calibrated
again by TEM,
nm and 104 nm were calculated.
Although
the
because the colloids in question
are not stable
influence
of the concentration
is not dramatic
over long periods. Further
drawbacks
are that
one should try to work always with identical
adsorption
can occur with very active colloids
colloid
concentrations
as those used for the
and the equipment
is relatively expensive. Widecalibration.
For higher concentrations
the effect
bore HDC cannot give direct information
about
is smaller
than for more dilute ones. On the
the size distribution. However,
in the case of
other hand highly concentrated
colloidal
solusmall semiconductor
or metal particles where the
tions might become unstable,
or particle growth
size-quantization
effects allow an independent
may occur. A 10-j M solution is recommended
measure
of the particle
size, it can be done
as a good compromise.
Then a fivefold higher or
indirectly
via the chromatograms
at various dea tenfold
lower concentration
would cause an
tection wavelengths
(see Fig. 3). But the wideerror of only 7 or 8 percent.
respectively.
bore HDC has many advantages.
Whereas in the
Ch.-H.
Fischer,
M. Gierslg
/ J.
classic HDC equipment
and handling
are quite
sophisticated
and detection is difficult due to the
small
dimensions,
the wide-bore
HDC
uses
cheap, empty,
standardised
capillaries.
Therefore one calibration
can be directly transferred
to
any other capillary of the same type. Although
this work was done with HPLC pumps, these
sophisticated
instruments
are not necessary
for
this technique.
A precise flow rate is not required because only the ratio of the peak heights
is measured,
and not the retention
time. The
method is fast and less calculation
is necessary
than in SEC. But most important
is the lack of
any packing material, the large surface of which
often causes problems of reversible and irreversible adsorption.
In a relatively wide capillary of
0.8 mm I.D. the surface does not play a significant role. Therefore
the method
is especially
recommended
for colloids
with high surface
activity.
Chramatogr.
A 688 (lY94)
tion Assessment
[lo]
ton,
C.A.
Sci.,
[ll]
C.A.
Sci..
(I21 J.G.
[l] A. Henglein,
Top. Curr. Chrm.,
143 (1988) 115.
[2] H. Weller, Advanced M aterials. 5 (1993) 88.
[3] E. Zeitler
(Editor),
Cryoscqy
and radiation damages.
North-Holland
Publ. Comp.. Amsterdam.
1982.
[4] J.J. Kirkland,
J. Chromatogr., 185 (1979) 273.
[S] Ch.-H. Fischer, J. Lilie, H. Weller, L. Katsikas and A.
Henglein,
Ber. Bunsenges.
Phys. Chem., 93 (1989) 61.
(AC’S Symposium
Sci..
American
Chemical Society, WashingDC, 1987, p. 256.
Silebi and J.G. DosRamos,
J. Colloid Interface
130 (1989) 14.
Silebi and J.G. DosRamos,
J. CoIloid Interface
133 (1989) 302.
DosRamos
and C.A. Silebi. J. Colloid Znterface
135 (1989) 165.
Silebi and J.G. DosRamos,
AZChE J., 35 (1989)
[13] C.A.
1351.
[14] M.E. Mullins and C. Orr, ht. J. M ultiphase Flow, 5
(1979) 79.
[15] R.J. Noel, K.M. Gooding,
F.E. Regnier, D.M. Ball, C.
Orr and M.E. Mullins, J. Chromafogr., 166 (1978) 373.
[16] F.M. Kelleher
and C.N. Trumbore,
Anal. Biochem.,
137 (1984) 20.
Anal.
Hancock
(1988)
Chem.,
and
R.E.
58 (1986) 3207.
Synovec,
Anal. Chem.,
60
2812.
[ 191 D.O.
Hancock,
C.N. Renn and R.E. Synovec,
Anal.
Chem., 62 (1990) 2441.
[ZO] G. Gerhardt
and R.N. Adams, Anal. Chem., 54 (1982)
2618.
[21] J.T. Vanderslice,
A.G. Rosenfeld
and G.R. Beecher,
Anal. Chim. Acru,
[22] L. Spanhel,
Am.
[23] J.W.
[24]
[ZS]
References
and Characterization
Series, Vol. 332),
[18] D.O.
The authors wish to express their thanks to
Dr. Paul Mulvaney,
Melbourne,
for the preparation of the gold colloids as well as for fruitful
discussions.
The help with the laboratory
work
of Ms. U. Michalczik
and the assistance of Ms.
U. Bloek with the electron microscopy
is gratefully acknowledged.
105
[6] Ch.-H.
Fischer, H. Weller, L. Katsikas and A. Henglein, Langrnuir,
5 (1989) 429.
[71 Ch.-H. Fischer, M. Giersig and T. Siebrands,
/. Chromarogr., 670 (1994) 89.
[8] Ch.-H. Fischer, J. Liq. Chromarogr., 17 (1994) 3593.
[9: A.W. Thornton,
J.P. Olivier,
C.G. Smart and L.B.
Gilman,
in T. Provder (Editor),
Particle Size Distribu-
[I71 J. Pawliszyn,
Acknowledgement
97- 105
[26]
[27]
[28]
Chem.
179 (1986)
M. Haase.
Sot.,
119.
H. Weller
109 (1987)
and A. Henglein,
J.
5649.
Slot and H. J. Geuze,
Eur. J. Cell. Chem., 38
(1985) 87-93.
C.R. Berry, Phys. Rev., 161 (1967) 611.
A. Fojtik, H. Weller, LJ. Koch and A. Henglein,
Ber.
Bunsenges. Phys. Chem., 88 (1984) 969.
T. Siebrands,
M. Giersig,
P. Mulvaney
and Ch.-H.
Fischer, Langmuir, 9 (1993) 2297.
V. Murugaiah
and R.E. Synovec, Anal. Chim. Acta, 246
(1991) 241.
C.N. Trumbore,
M. Grehlinger.
M. Stowe and F.M.
Kelleher,
J. Chromatogr., 322 (1985) 443.