NIH Public Access
Author Manuscript
IEEE Trans Nucl Sci. Author manuscript; available in PMC 2008 January 9.
NIH-PA Author Manuscript
Published in final edited form as:
IEEE Trans Nucl Sci. 2006 November 1; 2: 729–734.
Lens-Coupled CCD Detector for X-ray Crystallography
Timothy J. Madden [Member, IEEE],
Argonne National Laboratory, Argonne, IL 60439 USA (telephone: 630−252−7914, e-mail:
tmadden@anl.gov).
William McGuigan,
OPTICS One, West Lake Village, CA 91362 USA (telephone: 805−373−9340, e-mail:
wmcguigan@optisone.com).
Michael J. Molitsky,
Argonne National Laboratory, Argonne, IL 60439 USA (telephone: 630−252−1839, e-mail:
mmolitsky@anl.gov).
Istvan Naday,
Aviex, Naperville, IL 60565 USA (telephone: 708−355−3340, e-mail: snaday@core.com).
NIH-PA Author Manuscript
Alan McArthur, and
Aviex, Naperville, IL 60565 USA (telephone: 708−355−3340, e-mail: snaday@core.com).
Edwin M. Westbrook [Member, IEEE]
Molecular Biology Consortium, Chicago, IL 60612 USA (telephone: 630−222−1005, email:westbrook@lbl.gov).
Abstract
An x-ray crystallography detector (Blue-1) has been built based upon a Fairchild 486 backilluminated CCD and a custom lens system designed by Optics One Inc. The advantages of our Blue-1
lens system over more conventional fiber-optic tapers are: lower noise and higher efficiency;
improved point spread function; negligible spatial distortion; and lack of “chicken-wire” patterns.
Also, the engineering is simpler because the CCD is not bonded to the fiber-optic taper. A unique
mechanical design has been employed to accurately focus the image on the CCD. The detector
software is based on MATLAB and takes advantage of its powerful imaging and signal processing
libraries. The CCD timing can be updated on the fly by using a “CCD controller language” to specify
timing.
NIH-PA Author Manuscript
I. Introduction
In the field of molecular biology, crystallized biological materials are studied at synchrotrons
by X-ray diffraction, in which intensities of Bragg peaks incident upon the face of a detector
must be measured efficiently and accurately [1]-[5]. Typically, the X-ray field intensity is
converted to visible light in a phosphor film. This light is transferred to a CCD, usually via a
fiber-optic taper. Fiber optic tapers are normally used for this application to maximize detector
sensitivity and readout speed [6]-[9]. Several fiber-optically coupled detectors can be found in
the literature. A particularly good list of detectors appears in [10].
Although fiber-optic detectors have proven to be successful, there are several drawbacks with
these systems. First, fiber-optic tapers exhibit chicken wire, the projection of a honeycomb
pattern in the images due to the inherent design of the taper [11]. Second, the CCD must be
mechanically bonded to the fiber-optic glass (which is a complex process). Third, the fiberoptic tapers introduce spatial distortion into the images [12]. In large area fiber-optic systems,
several tapers are bonded together into a mosaic. These detectors are insensitive to light (or Xrays) at the seams where the tapers are bonded.
Madden et al.
Page 2
NIH-PA Author Manuscript
Though the light transfer efficiency of a lens system is less than that of a fiber-optic taper, we
have overcome this deficit by using a very large, back-illuminated CCD. Combining this with
a modern lens design, we have successfully developed a fine detector with lens optical coupling
[10].
Lenses do not exhibit several negative properties of fiber-optics, such as chicken wire. Lenses
can be designed to have finer point spread function than tapers, and much less spatial distortion.
Also unlike tapers, lenses can be tiled to have no insensitive areas in the detector face. Some
of us were involved in building several lens-coupled detectors in the past [13]-[15]. One of
these, a very successful lens detector called the “Noir-1,” is currently installed and operational
at Lawrence Berkeley National Laboratory. It has been in essentially continuous use for over
three years in service for a wide crystallographic community, and has taken thousands of high
quality datasets that have been used to solve hundreds of crystal structures.
NIH-PA Author Manuscript
We have recently built a sister to the Noir-1, called the “Blue-1.” Both detectors feature
identical lens systems designed by OPTICS One Inc. [16]. The main difference between the
Blue-1 and the Noir-1 is that, while the Noir-1 utilizes a Fairchild detector head, the Blue-1
features a custom detector head designed at Argonne National Laboratory. The purpose of the
Blue-1 design was to improve noise performance and to allow the detector to be tiled into a
2×2 system, because it is advantageous for a detector to have a large area in crystallography
applications [17], [18].
II. General Detector Features
The Blue-1 detector (Fig. 1) features a custom lens system, designed and built by OPTICS
One, and a Fairchild back-illuminated 486CCD, enclosed in a dewar filled with continuous
flowing dry nitrogen [19]. The nitrogen is supplied by a bottle of gas and is necessary to keep
the CCD dry. At a proper flow rate, the bottle lasts for about a month of continuous use. The
CCD is cooled to −27°C with a thermo-electric cooler (TEC), powered by a low noise
temperature controller. To keep the TEC cool, the TEC is attached to a metal plate through
which 10°C chilled water flows (temperature controlled by a Neslab RTE7).
The phosphor is Gd2S2O:Eu, deposited on a light-tight aluminized mylar sheet. The phosphor
sheet is fastened to the lens system by drawing a vacuum with a continuously running cheap,
small vacuum pump (the Air Admiral).
NIH-PA Author Manuscript
The CCD dewar features a front non-reflective window, which is bolted directly to the lens.
Thus, the CCD is fixed to the lens to preserve focal length. The electronics are connected by
short cables and are mechanically separate from the dewar. The camera is focused by moving
the CCD via a flex plate design, described below.
The detector head contains Argonne-designed analog and digital electronic circuit boards
attached to water-cooled plates, which remove heat and reduce thermal drift effects. The
electronics contain four analog-to-digital converters to simultaneously digitize each of the four
analog outputs from the Fairchild CCD. Digital circuitry is included that generates CCD clock
signals, and interfaces to the computer. The detector is connected to a Windows computer via
a National Instruments 6534 parallel interface card [20]. The computer runs a custom software
system created at Argonne for the detector. This software can run either standalone with a
MATLAB GUI, or with an EPICS interface [21], [22].
III. Optical Design
The lens system was designed and built by OPTICS One Corporation in West Lake Village,
CA. The design utilizes nine lenses, two of which are aspherically ground (Fig 2). The main
IEEE Trans Nucl Sci. Author manuscript; available in PMC 2008 January 9.
Madden et al.
Page 3
NIH-PA Author Manuscript
objectives of the lens design were: to provide a fine point spread function, to maximize
sensitivity, to allow tiling of lenses into a 2×2 array, and to properly focus the entire color
spectrum emitted by the phosphor. Table I summarizes the lens design parameters.
The lens system was designed to provide a sufficiently small point spread function that 99%
of the light from a point source on the phosphor falls within a diameter of 150μm in the image.
Figure 3 illustrates the point response diagrams simulated by OPTICS One, from a point source
on axis, and 10mm, 20mm, etc. radially away from the axis, on the image plane (the edge of
the image is 112.48mm from the axis).
In the 12 diagrams of Fig. 3, the differently colored rings represent the shapes of the foci of
different light wavelengths. The black circle in each diagram represents a 100μm diameter.
The lens is designed to be tiled as a 2×2 system, to create a detector of 317mm × 317mm active
area. Because the imaging fields of each lens overlap (by ∼0.5mm), the resulting detector has
no dead areas at module boundaries. To implement the 2×2 large array (Blue-4), a single large
phosphor would be installed on the detector to be imaged by the four lenses. At the time of this
publication, the tiled system has not yet been built. A model of the tiled lens system is shown
in Fig. 4.
IV. Detector Sensitivity
NIH-PA Author Manuscript
It has been argued in the literature that lens systems are too inefficient for X-ray detector
applications [10]. The efficiency of our lens, calculated in (1) from the numerical aperture (NA)
of 0.16 [15]:
E = τ × ( NA)2
(1)
(with τ = 0.8, the approximate total transmission coefficient of the lens system), is about 2%.
We kept the NA high by reducing the demagnification factor (2.589×), since the Fairchild
CCD486 is very big (61.44mm). About 500 optical photons are generated from a 12keV Xray hitting the phosphor [23]. This means that, for each X-ray photon, about 10 optical photons
make it to the CCD from the phosphor. The back-illuminated 486CCD has a very high quantum
efficiency: about 90%. Thus, of the 10 photons reaching the CCD, 9 of them generate an
electron in the CCD. This sensitivity has been verified by our measurements and is comparable
to commercial fiber-optic taper detectors, such as that of Area Detector Systems Corp (ADSC),
which gives 8.5e- for a 12keV X-ray photon [24]. With a read noise of 12e-, the Blue-1 signalto-noise ratio is about 0.75×. In comparison, the ADSC Q315 read noise is 18e-, so its signalto-noise ratio is about 0.47.
NIH-PA Author Manuscript
V. Focusing System
The detector is focused by moving the CCD relative to the lens. Because the depth of the focal
plane is less than 100μm, the CCD must be moved very finely. Also, because the CCD must
be kept cold and dry, it must be moved while keeping it sealed in a dewar. This is accomplished
by mounting the CCD on a “flex plate,” a thin aluminum plate that moves like a drum head.
Three micrometers push on the flex plate to allow CCD movement with micron precision. This
arrangement allows the CCD to be moved along the lens axis, and it can also be tilted in any
direction. A photograph of the detector head showing the dewar and micrometers is shown in
Fig. 5.
VI. System Specifications
General specifications of the detector system are shown in Table II.
IEEE Trans Nucl Sci. Author manuscript; available in PMC 2008 January 9.
Madden et al.
Page 4
NIH-PA Author Manuscript
In the Blue-1 detector, the Fairchild CCD486 is operated in 2-by-2 binning mode, so the total
number of pixels is 2048×2048. All four outputs of the CCD are read simultaneously at 500kHz
to give a total pixel rate of 2MHz.
The point spread function was measured by irradiating a 10μm-diameter hole in a tantalum
film with a 12keV X-ray beam, after the detector had been focused. Images were darksubtracted and corrected for nonuniform response. The response function fitted very well
(correlation coeff: 0.99) to a Gaussian function with σ=0.10mm, down to a noise floor of
±2ADU, where the peak response at the center was 25,000ADU [25].
Read noise was measured to be 12e- rms by reading only the serial register with no vertical
shifts. The electronic gain was measured by the mean/variance method [26], applied to a series
of dark images.
VII. Electronics Design
The detector electronics feature three major circuit boards and several minor circuit boards.
Major boards include the analog board, drive board, and digital board. Other boards include
the Interface Board and CCD socket assembly. A photograph of the detector head showing the
major circuit boards appears in Fig. 6.
NIH-PA Author Manuscript
The drive board has a field programmable gate array (FPGA) generating CCD signals as TTL
pulses. These pulses are converted to proper voltages acceptable by the CCD. To reduce noise,
the CCD clocks are filtered. The analog board contains four analog signal paths, one for each
CCD output. The digital board has enough RAM to store two image frames such that, as one
frame is downloaded to the computer, a new frame is read from the CCD creating a pipeline.
At the time of this publication, the pipeline operation has not yet been implemented. The digital
interface board has buffers to drive a parallel cable to the National Instruments card installed
on the PC. The CCD plugs into a socket assembly including a low noise preamplifier. The
CCD socket assembly resides in the gas-filled dewar. The analog and drive boards are fastened
to water-cooled aluminum plates for cooling and thermal regulation.
VIII. Analog Signal Path
The analog signal path was designed so that the CCD is the largest noise source. This was
accomplished by selecting a very high quality analog-to-digital converter (ADC), the AD7677
from Analog Devices, and by using a very low noise operational amplifier, the Texas
Instruments THS4032, in the first stage of the signal chain [27], [28]. The signal path is
diagrammed in Fig. 7.
NIH-PA Author Manuscript
The analog signal path has eight gain settings to set the gain from 5e-/ADU to 10e-/ADU in
eight steps. This is accomplished by switching the feedback resistance in an operational
amplifier circuit. To optimize read noise performance, the signal path has two selectable
bandwidths to allow reading the CCD at two different pixel rates. At the time of this publication,
only the 500kHz pixel rate has been used.
The analog signal path utilizes correlated double sampling (CDS) to reduce the noise added
by the CCD reset gate [29]. This is accomplished by shunting the capacitively coupled signal
to a programmed DC voltage after the reset. This DC voltage sets the background bias in the
captured image from the detector. The ADG201 from Analog Devices is used for the CDS
circuit. Because the ADC is differential, the CCD signal is converted from single ended to
differential before feeding to the ADC.
IEEE Trans Nucl Sci. Author manuscript; available in PMC 2008 January 9.
Madden et al.
Page 5
IX. CCD Control Language
NIH-PA Author Manuscript
An FPGA was designed for generating CCD timing as logic pulses. Because the CCD timing
must be altered for different readout speeds, binning, and CCD timing optimization, it was
desirable to alter timing without recompiling the FPGA. For this reason, the FPGA was
designed to function like a very simple microprocessor, complete with its own assembly
language. We call this language “CCD Control Language.” Using CCD Control Language, it
is possible to quickly reconfigure the CCD timing on the fly. In practice, writing Assemblylike language is timing-consuming. Therefore, a higher layer of software was created allowing
the use of Tcl scripts to create CCD timing sequences [30]. The steps to create CCD timing
are as follows. First, create Tcl arrays of 1s and 0s representing the CCD timing. Second,
convert the Tcl arrays into CCD timing language using custom Argonne-designed Tcl scripts.
Third, use Argonne-designed scripts to assemble the CCD Control Language into a binary file
to download to the FPGA. In this way, CCD timing is updated on the fly, with automatically–
generated CCD Control Language.
X. Calibration Process
The Blue-1 detector was shipped from Argonne National Laboratory to Lawrence Berkeley
National Laboratory for testing on beam line 4.2.2 at the Advanced Light Source synchrotron.
NIH-PA Author Manuscript
We calibrated the Blue-1 detector by standard procedures [31], [32]. Flood images were
recorded by tuning the beamline X-ray energy to 12keV and irradiating a thin (10μm) gold
foil. Gold absorbs X-rays with energy above 11,919keV, its absorption edge. The subsequent
gold fluorescence is primarily at 9,713keV, and it is truly isotropic. However gold foil, being
essentially a polycrystalline powder, also exhibits Bragg powder diffraction rings, the
narrowest of which is the (111) reflection (gold is a cubic crystal, Fm3m, a = 4.0786Å),
scattering from the gold foil at a Bragg θ angle of 12.68°. To avoid this ring and also avoid the
direct X-ray beam, the detector was placed 40cm from the gold foil and offset vertically 18cm.
Strong (∼20,000ADU/pixel) fluorescence signals were obtained with a 1,000s exposure.
Sixteen flood images were recorded, averaged, subjected to outlier (“zinger”) suppression, and
re-averaged with zingers removed to give a consensus flood image. Sixteen dark images were
also recorded (identical geometry used for flood images, but without X-ray exposure). These
also were averaged, zinger-suppressed, and reaveraged. The consensus dark image was
subtracted from the consensus flood image, and this X-ray field intensity image was compared
algebraically to the theoretical reference field, i.e., the expected field strength in this geometry,
which follows a cos2(θ) functional form. In this calibration procedure we calculate an image
file that is, for each pixel, the multiplicative factor that corrects for nonuniform sensitivity/
response over the entire detector face.
NIH-PA Author Manuscript
Sixteen dark images were recorded at recording times of 0.1s, 1s, 10s, 100s, 200s, . . ., 1,000s.
The 0.1s dark image was taken to be the “DC Offset” image. For each recording time, the
sixteen images were averaged, zinger-suppressed, and reaveraged. To determine electronic
gain, the variance (square of the standard deviation) of these dark images was compared to
their averages, pixel by pixel. The average ratio of (σ2/<Dark>), taken to be the electronic gain
of the system [26], was 4.89 e-/ADU.
To study spatial distortion of the detector image, we fabricated a 1mm-thick brass plate
covering the entire detector face. We machined a square raster of holes (each 1/32” diameter),
10 holes/inch, in the brass plate, put this plate over the detector face, and exposed the assembly
to the same fluorescence flood field used above. Good image contrast of the spots due to holes
was obtained in 100s images. A computer program was written to find the centroids of these
spots, fit a polynomial function to each vertical and horizontal line of spot centroids, and
IEEE Trans Nucl Sci. Author manuscript; available in PMC 2008 January 9.
Madden et al.
Page 6
NIH-PA Author Manuscript
calculate the conformal mapping between the observed spot positions and their expected
positions in a square matrix. The essential bottom-line of these tests was that there was minimal
distortion of the image.
XI. Crystallographic Testing
The first complete dataset of a protein crystal recorded by the Blue-1 detector was from a
routinely-used benchmark protein crystal, hen egg-white lysozyme. One hundred eighty data
frames, each recording diffraction from the crystal as it rotated 1°, were recorded. Each data
frame was recorded for 1 second. One such diffraction image is shown in Fig. 8.
At the time we recorded this dataset, the Blue-1 detector had not yet been perfectly focused,
as can be appreciated here, since the Bragg spots are slightly blurred in this image.
NIH-PA Author Manuscript
As a measure of data quality, crystallographers use a quantity called the “R-merge” to measure
the quality of the diffraction data [33]. The overall R-merge for this first dataset was 3%, for
209,211 Bragg spot measurements out to the edge of the detector (1.5Å resolution). This is
quite good. R-merge values are never better than 2%. Typically for delivery of a new
commercial detector system, an acceptance requirement would be that a lysozyme dataset be
recorded with an R-merge better than 5%, so this detector is already good enough to be used
routinely at a synchrotron beamline. We anticipate that the Blue-1 detector will perform even
better once it is focused properly, and once we have had the opportunity to truly tune it up.
The data frames recorded by our lens system are exceptionally clean; they have no chicken
wire patterns because they lack fiber-optics. Also, we verified that no spatial correction is
needed for solving crystal structures because the lens essentially does not distort the image. In
fact the lysozyme dataset discussed above was processed without any distortion corrections at
all.
XII. Conclusion
NIH-PA Author Manuscript
By building the Blue-1 detector, we have demonstrated that an excellent crystallographic
detector system can be developed in which the light transfer medium is a lens, rather than a
fiber-optic taper. A drawback to the lens system is that, in order to achieve the required
sensitivity, a large back-illuminated CCD is necessary, and the lens must be long. Indeed, the
lens system on the Blue-1 detector is 750mm long, taking up room in the hutch. The
performance of the lens system is comparable or superior to conventional fiber-optic systems
in terms of point spread function, noise, and sensitivity. The lens data are more uniform, and
they are cleaner than those from a fiber-optic system. Furthermore, no software spatial
correction is needed.
We continue testing and evaluating the Blue-1 detector. It is already excellent. Over the next
several months we expect to be able to demonstrate that it will perform even better.
Acknowledgment
The authors would like to thank the following people: Patricia Fernandez, Ira Goldberg, John Quintana, Rich Voogd,
Michael Anthony, Stephen Ross, Antonino Miceli, John Weizeorick, John Lee, John Morse, Todd Hayden, Patrick
DeLurgio and Paul Vu. This work has been supported by the National Institutes of Health through grant R01 RR16334,
and by the National Science Foundation through grant DBI01-16615. The work at Argonne is supported by the U. S.
Department of Energy, Office of Science, Office of Basic Energy Sciences, under contract no. DE-AC02-06CH11357.
This work was supported by the National Institutes of Health under grant R01 RR16334 and the National Science
Foundation under grant DBI01-16615.
IEEE Trans Nucl Sci. Author manuscript; available in PMC 2008 January 9.
Madden et al.
Page 7
References
NIH-PA Author Manuscript
NIH-PA Author Manuscript
NIH-PA Author Manuscript
1. Hendrickson WA. Determination of macromolecular structures from anomalous diffraction of
synchrotron radiation. Science 1991;254:51–58. [PubMed: 1925561]
2. Hendrickson WA, Ogata CM. Phase determination from multi-wavelength anomalous diffraction
measurements. Meth. in Enzymol 1997;276:494–523.
3. Holmes KC. From Rare to Routine. Nature Struct. Biol 1998;5:618–619. [PubMed: 9699606]
4. Helliwell, JR. Macromolecular Crystallography with Synchroton Radiation. Cambridge University
Press; UK: 1992. p. 181-210.
5. Sweet RM. The technology that enables synchrotron structural biology. Nature Str. Biol 1998;5:654–
656.
6. Westbrook EM, Naday I. Charge-coupled device-based area detectors. Meth. in Enzymol
1997;276:244–268. [PubMed: 9048377]
7. Naday I, Westbrook EM, Westbrook ML, Travis DJ, Stanton M, Phillips WC, Xie J. Characterization
and data collection on a direct-coupled CCD X-ray detector. Nucl. Instrum. and Meth. A
1994;348:635–640.
8. Naday I, Ross S, Westbrook EM, Zentai G. Charge-coupled device/fiber optic taper array x-ray detector
for protein crystallography. Optical Engin 1998;37:1235–1244.
9. Phillips WC, Stanton M, O’Mara D, Naday I, Westbrook EM. Modular CCD detector for x-ray
crystallography. Proc. S. P. I. E 1993;1900:53–58.
10. Gruner SM, Tate MW, Eikenberry EF. Charge-Coupled Device Area X-ray Detectors. Review of
Scientific Instruments 2002;73(8):2815–2842.
11. Incom Inc.. http://www.incomusa.com/http://www.incomusa.com/Corporate website
12. Coleman CI. Imaging characteristics of rigid coherent fibre optic tapers. Adv. in Elec. and Electron
Phys 1985;64B:649–661.
13. Strauss MG, Naday I, Sherman IS, Kraimer MR, Wesbrook EM. CCD-based synchrotron X-ray
detector for protein crystallography-performance projected from an experiment. I. E. E. E. Trans.
Nucl. Sci 1987;NS-34:389–395.
14. Strauss MG, Westbrook EM, Naday I, Coleman TA, Westbrook ML, Travis DJ, Sweet RM, Pflugrath
JW, Stanton M. CCD-based detector for protein crystallography with synchrotron X-rays. Nucl.
Instrum. and Meas. A 1990;297:275–295.
15. Westbrook EM, Morse J, Fisher RE, McGuigan W, Onishi S, Vu P. Design and characterization of
a large-area, lens-coupled CCD detector for macromolecular crystallography. Proc. of S. P. I. E
2002;4784:375–385.
16. OPTICS. http://www.optics1.com/http://www.optics1.com/One corporate website
17. Stanton M. Area detector design Part I. Formulation of design criteria and application to X-ray
crystallography. Nucl. Instrum. and Meth. A 1993;325:550–557.
18. Stanton M, Phillips WC, O’Mara D, Naday I, Westbrook EM. Area detector design Part II. Application
to a modular CCD-based detector for X-ray crystallography. Nucl. Instrum. and Meth. A
1993;325:558–567.
19. Fairchild Imaging Inc.. CCD 486 datasheet. http://www.fairchildimaging.com/main/documents/
CCD486DataSheetRevB.pdfhttp://www.fairchildimaging.com/main/documents/
CCD486DataSheetRevB.pdf
20. National Instruments Inc.. NI PCI 6534 high speed pattern I/O handshaking. http://sine.ni.com/nips/
cds/view/p/lang/en/nid/201909http://sine.ni.com/nips/cds/view/p/lang/en/nid/201909
21. The Mathworks. MATLAB, the language of technical computing. http://www.mathworks.com/
products/matlab/http://www.mathworks.com/products/matlab/
22. Argonne National Laboratory. Experimental physics and industrial control system. http://
www.aps.anl.gov/epics/http://www.aps.anl.gov/epics/
23. Stanton M. The detective quantum efficiency of CCD and vidicon-based detectors for X-ray
crystallographic applications. J. Appl. Cryst 1992;25:638–645.
24. Area Detector Systems Corporation. Technical comparison. http://www.adsc-xray.com/
prod_compare.htmlhttp://www.adsc-xray.com/prod_compare.html
IEEE Trans Nucl Sci. Author manuscript; available in PMC 2008 January 9.
Madden et al.
Page 8
NIH-PA Author Manuscript
25. Harris, J.; Stocker, H. Handbook of Mathematics and Computational Science. Springer; New York:
1998. p. 796-797.
26. Mortara L, Fowler A. Evaluations of charge-coupled device (CCD) performance for astronomical
use. Proc. S.P.I.E 1981;290:28–33.
27. Analog Devices Inc.. AD7677, 16-Bit, 1 LSB INL, 1 MSPS differential PulSAR ADC. http://
www.analog.comhttp://www.analog.com
28. Texas Instruments Inc.. THS4032, 100-MHz low noise voltage-feedback amplifier. http://
focus.ti.com/docs/prod/folders/print/ths4032.htmlhttp://focus.ti.com/docs/prod/folders/print/
ths4032.html
29. Barbe D. Imaging devices using the charge-coupled concept. Proc. I.E.E.E 1975;63:38–67.
30. Flynt, C. Tcl/Tk for Real Programmers. Morgan Kaufmann; New York: 1999. p. 33-72.
31. Stanton M, et al. Correcting spatial distortion and non-uniform response in area detectors. J. Appl.
Cryst 1992;25:549–558.
32. Thomas DJ. Calibrating an area detector diffractometer: integral response. Proc. Royal Soc. Lond
1990;428:181–214.
33. Stout, GH.; Jensen, LH. X-ray Structure Determination. 2nd edition. John Wiley & Sons; N.Y.: 1989.
NIH-PA Author Manuscript
NIH-PA Author Manuscript
IEEE Trans Nucl Sci. Author manuscript; available in PMC 2008 January 9.
Madden et al.
Page 9
NIH-PA Author Manuscript
Fig. 1.
Argonne Blue-1 detector. The detector head was designed and built at Argonne National
Laboratory. Lens was designed and built at OPTICS One.
NIH-PA Author Manuscript
NIH-PA Author Manuscript
IEEE Trans Nucl Sci. Author manuscript; available in PMC 2008 January 9.
Madden et al.
Page 10
NIH-PA Author Manuscript
Fig. 2.
Diagram of lens design. Nine lenses are included with seven spherical and two aspherical. The
total length from object (phosphor on left) to CCD at far right is 750mm.
NIH-PA Author Manuscript
NIH-PA Author Manuscript
IEEE Trans Nucl Sci. Author manuscript; available in PMC 2008 January 9.
Madden et al.
Page 11
NIH-PA Author Manuscript
NIH-PA Author Manuscript
Fig. 3.
Encircled energy plots of lens design. Simulated image on CCD of a point source.
NIH-PA Author Manuscript
IEEE Trans Nucl Sci. Author manuscript; available in PMC 2008 January 9.
Madden et al.
Page 12
NIH-PA Author Manuscript
NIH-PA Author Manuscript
Fig. 4.
Four lenses stacked as a tiled array.
NIH-PA Author Manuscript
IEEE Trans Nucl Sci. Author manuscript; available in PMC 2008 January 9.
Madden et al.
Page 13
NIH-PA Author Manuscript
NIH-PA Author Manuscript
Fig. 5.
Detector head showing dewar and micrometers.
NIH-PA Author Manuscript
IEEE Trans Nucl Sci. Author manuscript; available in PMC 2008 January 9.
Madden et al.
Page 14
NIH-PA Author Manuscript
NIH-PA Author Manuscript
Fig. 6.
Photograph of detector head showing major circuit boards.
NIH-PA Author Manuscript
IEEE Trans Nucl Sci. Author manuscript; available in PMC 2008 January 9.
Madden et al.
Page 15
NIH-PA Author Manuscript
Fig. 7.
NIH-PA Author Manuscript
Diagram of analog signal path. The CCD signal is capacitively coupled to the preamplifier, in
the dewar. The preamplifier feeds to the selectable gain stage, filtering stage, and DCS circuit.
The signal is converted to differential before sending to the ADC.
NIH-PA Author Manuscript
IEEE Trans Nucl Sci. Author manuscript; available in PMC 2008 January 9.
Madden et al.
Page 16
NIH-PA Author Manuscript
NIH-PA Author Manuscript
Fig. 8.
Sample diffraction image from Argonne Blue-1 detector.
NIH-PA Author Manuscript
IEEE Trans Nucl Sci. Author manuscript; available in PMC 2008 January 9.
Madden et al.
Page 17
Table I
General Lens Specifications
NIH-PA Author Manuscript
Specification
Value
Active Imaging Area
Demagnification
Factor
Image on CCD
Numerical Aperture
Efficiency
Sensitivity at Edge
Spot Size/Resolution
Spatial Distortion
160mm × 160mm
2.589×
61.44mm × 61.44mm
0.16 (at phosphor)
>2%
57% of sensitivity at center
99% encircled energy within a 150μm diameter in the image plane
<0.5%
NIH-PA Author Manuscript
NIH-PA Author Manuscript
IEEE Trans Nucl Sci. Author manuscript; available in PMC 2008 January 9.
Madden et al.
Page 18
Table II
Detector Specifications
NIH-PA Author Manuscript
Specification
Value
Frame Rate
Pixel Rate
Read Noise
Full Well
Dark Current
Dark Current Noise
Pixel Size(at phosphor)
Number of Pixels
Conversion Gain
Electronics Gain
Dynamic Range
Point Spread Function
FWHM
Phosphor
0.44 frames/s
500,000 pixels/s
12e- rms
∼400,000 e32 e-/sec-pixel, @ −27°C
6 e- rms for 1 second image
78 μm × 78 μm
2048×2048 pixels
9e-/12keV X-ray photon, measured
5 e-/ADU, measured
∼33,000
Gaussian, σ=0.1 mm
180μm (2.3 pixels), measured
Gd2O2S: Eu
NIH-PA Author Manuscript
NIH-PA Author Manuscript
IEEE Trans Nucl Sci. Author manuscript; available in PMC 2008 January 9.